Skip to content

iCAST 85451Recalls & reported problems

Atrium Medical Corporation · iCAST COVERED STENT, 5MMX22MMX120CM · Class 2 · Implant

All models of iCAST (29)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00650862854510), received between May 10, 2016 and Oct 1, 2021 — classified as 5 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K050814); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Device-Device Incompatibility2
  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Burst Container or Vessel1
  • Component Falling1
  • Detachment Of Device Component1
  • Device Slipped1
  • Leak/Splash1

Patient problems

  • Aneurysm2
  • No Consequences Or Impact To Patient2
  • No Information2
  • Calcium Deposits/Calcification1
  • Hemorrhage/Blood Loss/Bleeding1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 1, 2021Device-Device Incompatibility; Leak/Splash; Device-Device Incompatibility

    REPORT RECEIVED STATED ON 31JUL2017, THE POST PROCEDURE COMPUTERIZED TOMOGRAPHY (CT) SCAN WAS PERFORMED. SITE ANALYSIS NOTED THAT THE DEVICES WERE PATENT WITH NO EVIDENCE OF SEPARATION OF COMPONENTS, OR DEVICE INTEGRITY ISSUES. A TYPE II ENDOLEAK WAS NOTED. CORE LAB ANALYSIS OF THE CT SCAN NOTED PATENT DEVICES WITH NO EVIDENCE OF SEPARATION OF COMPONENTS OR DEVICE INTEGRITY ISSUES. A TYPE III ENDOLEAK WAS NOTED AT THE P-BRANCH AND THE RIGHT RENAL STENT. ON (B)(6) 2018, THE SIX-MONTH FOLLOW-UP CT SCAN AND ULTRASOUND WERE PERFORMED. SITE ANALYSIS NOTED THAT THE DEVICES WERE PATENT WITH NO EVIDENCE OF MIGRATION, SEPARATION OF COMPONENTS, OR DEVICE INTEGRITY ISSUES. A TYPE II ENDOLEAK WAS NOTED. CORE LAB ANALYSIS OF THE CT SCAN NOTED PATENT DEVICES WITH NO EVIDENCE OF MIGRATION, SEPARATION OF COMPONENTS, OR DEVICE INTEGRITY ISSUES. AN UNKNOWN TYPE ENDOLEAK WAS NOTED. CORE LAB NOTED, ¿UNSURE ORIGIN OF ENDOLEAK.¿ CORE LAB EVALUATION OF THE ULTRASOUND NOTED PATENT VESSELS WITH AN UNKNOWN TYPE ENDOLEAK. COMMENTS FROM CORE LAB NOTED ¿ENDOLEAK ON TWO CINE LOOPS: ONE SUGGESTS FROM THE GRAFT ITSELF (TYPE IV) (DISTAL PART OF THE FENESTRATED COMPONENT) WHILE ANOTHER SUGGESTS FROM THE IMA (TYPE I

    Patient problems: Aneurysm; Hemorrhage/Blood Loss/Bleeding

    MAUDE report 12565724 ↗
  • InjuryAug 30, 2021Device-Device Incompatibility; Adverse Event Without Identified Device or Use Problem; Device-Device Incompatibility

    N/A.

    Patient problems: Aneurysm; Calcium Deposits/Calcification

    MAUDE report 12383868 ↗
  • MalfunctionMar 21, 2021Material Rupture; Material Rupture

    REPORT RECEIVED STATED THAT A 4 MM VESSEL THAT WAS POSTERIOR DOWN FACING. SLOWLY INFLATED 5X22X120 ICAST AND DEFLATED ICAST. REINFLATED ICAST AND THE BALLOON POPPED AND INVERTED ON ITSELF.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 11536813 ↗
  • MalfunctionDec 22, 2020Break

    IT WAS REPORTED THAT DURING A PROCEDURE OF RENAL STENT PLACEMENT, THE SURGEON NOTICED THE STENT WAS RIPPED SO HE TOOK IT OUT.

    Patient problems: No Information

    MAUDE report 11061317 ↗
  • MalfunctionJul 2, 2019Burst Container or Vessel

    N/A.

    Patient problems: No Information

    MAUDE report 8752753 ↗
  • InjuryDec 22, 2016Detachment Of Device Component

    AS THE PHYSICIAN WAS PUSHING THE STENT OVER THE WIRE THE STENT CAME LOOSE AND WAS OFF THE BALLOON.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6197804 ↗
  • InjurySep 23, 2016Device Slipped

    THE DEVICE WAS PLACED INTO A 6FR ANSEL SHEATH HUB AND THE STENT SLIPPED BACK ONTO THE BALLOON CATHETER BEFORE IT COULD BE PLACED IN THE PATIENT. THE PATIENT WASN'T HARMED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5973725 ↗
  • InjuryMay 10, 2016Component Falling

    THE PROCEDURE WAS AN INTERVENTION ON A RENAL ARTERY. IT WAS REPORTED THAT THE STENT FELL OFF PRIOR TO INFLATING THE BALLOON.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 5644761 ↗

FDA clearance and classification

Cleared through Special 510(k) K050814 on May 6, 2005 (decision: Substantially Equivalent). Submitted by Atrium Medical Corp. as “Atrium Icast Covered Stent”. The FDA review took 36 days. FDA record ↗

The FDA regulates product code JCT (prosthesis, tracheal, expandable) under 21 CFR 878.3720 as a Class II device. 21 CFR 878.3720 ↗

The FDA lists this type of device as life-sustaining or life-supporting.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K050814. One submission often covers many models, so check the product and lot details in each record.

  • Z-1077-2022 · Mar 3, 2022 · Open, Classified
    iCast Covered Stent System Product Code/Part Number: 85400 85401 85402 85403 85404 85405 85406 85407 85408 85409 85410 85411 85412 85413 85414 85415 85416 85417 85418 85419 85420
  • Z-2761-2017 · May 25, 2017 · Terminated
    Atrium iCAST Balloon Expandable Covered Stent; 7mm x 22mm x 120cm; Model/Part Number 85455 (Marketed in U.S. only) Atrium Advanta V12 Balloon Expandable Covered Stent; 7mm x 22mm

Common questions about the iCAST 85451

Has the iCAST 85451 been recalled?

No FDA recall record in our data names this exact device identifier (DI 00650862854510), based on the FDA recall file dated Nov 27, 2026. However, 2 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the iCAST 85451?

The FDA's MAUDE database lists 8 adverse event reports for this device, received between May 10, 2016 and Oct 1, 2021. In those reports, the most common specific device problems are “Device-Device Incompatibility” (2), “Break” (1) and “Burst Container or Vessel” (1); the most common patient problems are “Aneurysm” (2), “Calcium Deposits/Calcification” (1) and “Hemorrhage/Blood Loss/Bleeding” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the iCAST 85451?

None of the reports are classified as deaths, 5 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the iCAST 85451 FDA approved?

It was cleared through Special 510(k) K050814 on May 6, 2005. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the iCAST 85451 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the iCAST 85451, and is it still sold?

The FDA's device database lists Atrium Medical Corporation as the company responsible for it (brand iCAST, model 85451). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Prosthesis, tracheal, expandable devices

All Prosthesis, tracheal, expandable devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page