HIgH-Flo™ Needle Set RMS22609Recalls & reported problems
Repro Med Systems, Inc. · 26 Gauge Needle Sets - 2 needles - 9 mm · Class 2
All models of HIgH-Flo Needle Set (11)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00659443000308), received between Apr 21, 2022 and Apr 21, 2022 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Contamination /Decontamination Problem” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K122404, K200176, K211206, K214045, K252015); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Contamination /Decontamination Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 21, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem
ON (B)(6) 2022, KORU MEDICAL WAS MADE AWARE THAT DURING INCOMING INSPECTION AT A CUSTOMER WAREHOUSE IN (B)(4) AN INSECT WAS FOUND INSIDE A SEALED STERILE NEEDLE SET POUCH. THE TEAM AT THE (B)(4) WAREHOUSE TESTED THE POUCH WITH THE DEBRIS INSIDE AND VERIFIED VIA HAND PRESSURE THAT THE BAG WAS IN FACT SEALED. SUBSEQUENTLY, THE (B)(4) WAREHOUSE CONDUCTED A 100% INSPECTION ON THE REMAINING INVENTORY FROM MODEL RMS22609/LOT N.80137. UPON COMPLETION, AN ADDITIONAL SEALED POUCH WAS FOUND TO CONTAIN AN UNIDENTIFIED BLACK CONTAMINANT. ADDITIONAL INFORMATION RECEIVED FROM THE INITIAL REPORTER STATED LOT N.80137 CONTAINED 1000 POUCHES. OF THE 1000, 11 POUCHES WERE PROVIDED FOR CLINICAL USE ON STUDY SUBJECTS ON (B)(6) 2022. AT THIS TIME, THE REMAINING 989 BAGS FROM THE LOT ARE QUARANTINED IN THE CUSTOMER'S INVENTORY SYSTEM. A RETURN MATERIAL AUTHORIZATION WAS CREATED FOR RETURN OF THE 2 POUCHES TO KORU MEDICAL FOR FURTHER EVALUATION. THE PRODUCTS ARE PENDING RETURN TO KORU MEDICAL AT THIS TIME. KORU MEDICAL COMPLETED AN INITIAL INVESTIGATION OF THE REPORTED INCIDENT. A SUMMARY OF THE RESULTS IS AS FOLLOWS: THE NEEDLE SETS WERE SHIPPED TO (B)(4) WITH THE INTENTION OF SUPPORTING A CLINICAL STUDY
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14166645 ↗MalfunctionApr 21, 2022Contamination /Decontamination Problem; Contamination /Decontamination Problem
ON 23-MAR-2022, KORU MEDICAL WAS MADE AWARE THAT DURING INCOMING INSPECTION AT A CUSTOMER WAREHOUSE IN (B)(6), JAPAN AN INSECT WAS FOUND INSIDE A SEALED STERILE NEEDLE SET POUCH. THE TEAM AT THE JAPANESE WAREHOUSE TESTED THE POUCH WITH THE DEBRIS INSIDE AND VERIFIED VIA HAND PRESSURE THAT THE BAG WAS IN FACT SEALED. SUBSEQUENTLY, THE JAPANESE WAREHOUSE CONDUCTED A 100% INSPECTION ON THE REMAINING INVENTORY FROM MODEL RMS22609/LOT N.80137. UPON COMPLETION, AN ADDITIONAL SEALED POUCH WAS FOUND TO CONTAIN AN UNIDENTIFIED BLACK CONTAMINANT. ADDITIONAL INFORMATION RECEIVED FROM THE INITIAL REPORTER STATED LOT N.80137 CONTAINED 1000 POUCHES. OF THE 1000, 11 POUCHES WERE PROVIDED FOR CLINICAL USE ON STUDY SUBJECTS ON (B)(6) 2022. AT THIS TIME, THE REMAINING 989 BAGS FROM THE LOT ARE QUARANTINED IN THE CUSTOMER'S INVENTORY SYSTEM. A RETURN MATERIAL AUTHORIZATION WAS CREATED FOR RETURN OF THE 2 POUCHES TO KORU MEDICAL FOR FURTHER EVALUATION. THE PRODUCTS ARE PENDING RETURN TO KORU MEDICAL AT THIS TIME. KORU MEDICAL COMPLETED AN INITIAL INVESTIGATION OF THE REPORTED INCIDENT. A SUMMARY OF THE RESULTS IS AS FOLLOWS: THE NEEDLE SETS WERE SHIPPED TO JAPAN WITH THE INTENTION OF SUPPORTING A CLIN
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14166644 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K122404, K200176, K211206, K214045, K252015. One submission often covers many models, so check the product and lot details in each record.
- Z-0051-2020 · Aug 20, 2019 · Terminated20" 2-Needle 26-Gauge 6mm HlgH-Flo Subcutaneous Safety Needle Sets" Product # RMS22606 - Product Usage: RMS HlgH-Flo Subcutaneous Safety Needle Sets are intended for the delivery
- Z-0318-2018 · Mar 10, 2016 · TerminatedHIgH-Flo Subcutaneous Safety Needle Sets, 2-Needle Set/ 24 Gauge 9mm Needles, REF/Part Numbers RMS22606, RMS22609, RMS22612, RMS12606, RMS22409, RMS32609, RMS32609, RMS12409, RMS1
Other Set, administration, intravascular devices
- Autosoft XC 1001680Unomedical A/S · Set, Administration, Intravascular1 death16,105 reports
- Autosoft90™ 1002817Unomedical A/S · Set, Administration, Intravascular1 recall11,907 reports
- Minimed™ Quick-set™ MMT-399AUnomedical A/S · Set, Administration, Intravascular9 deaths8,669 reports
- Medtronic Extended MMT-431AUnomedical A/S · Set, Administration, Intravascular3 deaths8,574 reports
- MiniMed™ Mio™ Advance MMT-242AUnomedical A/S · Set, Administration, Intravascular15 deaths8,164 reports
- CADD 21-7302-24Icu Medical, Inc. · Set, administration, intravascular4 recalls1 death4,657 reports
- Autosoft XC 1001681Unomedical A/S · Set, Administration, Intravascular1 death4,216 reports
- Minimed™ Quick-set™ MMT-397AUnomedical A/S · Set, Administration, Intravascular19 deaths4,134 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page