Skip to content

HIgH-Flo™ Needle Set RMS32414Recalls & reported problems

Repro Med Systems, Inc. · 24 Gauge Needle Sets - 3 needles - 14 mm · Class 2

All models of HIgH-Flo Needle Set (11)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00659443000643), received between Sep 9, 2025 and Sep 9, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Contamination /Decontamination Problem” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K122404, K200176, K211206, K214045, K252015); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Contamination /Decontamination Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 9, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    KORU MEDICAL WAS MADE AWARE THAT DURING INITIAL INCOMING INSPECTION AT A CUSTOMER WAREHOUSE IN JAPAN, UNIDENTIFIED BLACK DEBRIS/CONTAMINANT WAS FOUND INSIDE A SEALED STERILE NEEDLE SET POUCH INVOLVING RMS32409/LOT N.25B05N. MDR 1318360-2025-00020 IS BEING FILED FOR RMS32409/LOT N.25B05N. SUBSEQUENTLY, THE JAPANESE WAREHOUSE PERSONNEL COMPLETED INSPECTION ON THE REMAINING INVENTORY. UPON COMPLETION, AN ADDITIONAL SEALED POUCH WAS FOUND TO CONTAIN UNIDENTIFIED BLACK DEBRIS/CONTAMINANT INVOLVING RMS32414/LOT N.87033. THE CUSTOMER PROVIDED PHOTOS OF THE POUCHES CONFIRMING THE COMPLAINT. THERE WAS NO PATIENT INVOLVEMENT. A REVIEW OF THE DEVICE HISTORY RECORD FOR RMS32414/LOT N.87033 WAS PERFORMED AND NO NONCONFORMANCES OR ABERRANT CONDITIONS WERE NOTED. HIGH-FLO NEEDLE SETS ARE MANUFACTURED AND SEALED INTO THEIR PRIMARY STERILE BARRIER PACKAGING IN A CONTROLLED ENVIRONMENT. IN ADDITION, HIGH-FLO NEEDLE SETS ARE STERILIZED USING GAMMA IRRADIATION WHICH IS A HIGHLY PENETRATING AND EFFECTIVE STERILIZATION MODALITY. FURTHERMORE, VALIDATION OF THE STERILIZATION METHOD WAS PERFORMED USING A CONSERVATIVE METHOD (VDMAX25) WHICH PROVIDES A SAFETY MARGIN OVER ALTERNATE ACCEPTABLE VALIDATION METHO

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23006864 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K122404, K200176, K211206, K214045, K252015. One submission often covers many models, so check the product and lot details in each record.

  • Z-0051-2020 · Aug 20, 2019 · Terminated
    20" 2-Needle 26-Gauge 6mm HlgH-Flo Subcutaneous Safety Needle Sets" Product # RMS22606 - Product Usage: RMS HlgH-Flo Subcutaneous Safety Needle Sets are intended for the delivery
  • Z-0318-2018 · Mar 10, 2016 · Terminated
    HIgH-Flo Subcutaneous Safety Needle Sets, 2-Needle Set/ 24 Gauge 9mm Needles, REF/Part Numbers RMS22606, RMS22609, RMS22612, RMS12606, RMS22409, RMS32609, RMS32609, RMS12409, RMS1

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page