Freedom 1000 FM1000Recalls & reported problems
Umano Medical Inc · AC-powered medical bed. · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00670482000487), received between Mar 24, 2022 and Jan 17, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Communication or Transmission Problem” (1 report).
2 FDA recall records name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Apr 11, 2022Terminated
There is a potential hazard of uninterrupted motion of the Motorized Wheel System due to the possible failure of a component in the electrical system of the Product.
Z-1135-2022 · Root cause: Nonconforming Material/Component · FDA record ↗
- Apr 11, 2022Terminated
There is a potential hazard of uninterrupted motion of the Motorized Wheel System due to the possible failure of a component in the electrical system of the Product.
Z-1136-2022 · Root cause: Nonconforming Material/Component · FDA record ↗
Most reported problems
Device problems
- Communication or Transmission Problem1
- Failure to Shut Off1
- Flare or Flash1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 17, 2025Flare or Flash; Flare or Flash
CUSTOMER REPORTED THAT A CRACKLING SOUND AND AN ELECTRICAL FLASH WAS NOTICED FROM A REMOVABLE POWER CORD CONNECTED TO THE BED. HE MENTIONED THAT EVERYTHING STOPPED ONCE THE BED WAS UNPLUGGED. NO PATIENT INJURY WAS REPORTED. THE NURSE CUT HER HAND WHILE REACHING THE BED. A MANUFACTURER SERVICE TECHNICIAN VISITED THE SITE TO INVESTIGATE THE DEVICE. THE POWER CORD WAS REPLACED, AND THE BURN MARK ON THE HEAD SECTION OF THE BED WAS REMOVED. THE BED IS NOW FUNCTIONING PROPERLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21183226 ↗MalfunctionSep 6, 2024Communication or Transmission Problem; Communication or Transmission Problem
THE CUSTOMER REPORTED TO UMANO MEDICAL REPRESENTATIVE THAT THE BED IS NOT SENDING SIGNAL TO THE HOSPITAL'S NURSE CALL SYSTEM. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20160448 ↗MalfunctionMar 24, 2022Failure to Shut Off; Unintended Movement; Failure to Shut Off
DISTRIBUTER CONTACTED TECHNICAL SERVICE OF THE MANUFACTURER TO REPORT AN ISSUE WITH THE MOTORIZED WHEEL SYSTEM OF A HOSPITAL BED THAT WAS BEING USED BY ONE OF HIS CUSTOMERS DURING A TRIAL SESSION (THE BED WAS NOT IN USE WITH A PATIENT) HE MENTIONED THAT WHEN THEY RELEASED THE TRIGGER TO STOP THE WHEEL MOVEMENT, THE BED CONTINUED TO MOVE FORWARD AND WAS STOPPED AGAINST THE WALL. THERE WAS NO PATIENT INVOLVEMENT AND NO ADVERSE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13892885 ↗
Other Bed, Ac-Powered Adjustable Hospital devices
- Centrella™ Bed P7900BHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL2 recalls5 deaths1,937 reports
- Progressa® Bed System P7500AHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall1 death1,377 reports
- VersaCare® Med Surg Bed P3200KHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL3 deaths1,193 reports
- Medline MDR107002EMedline Industries, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall1 death374 reports
- Medline MDR107003EMedline Industries, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL1 recall3 deaths362 reports
- InTouch 2141Stryker Corporation · BED, AC-POWERED ADJUSTABLE HOSPITAL2 deaths361 reports
- Citadel Plus FX811B3B4AMABBArjohuntleigh AB · BED, AC-POWERED ADJUSTABLE HOSPITAL2 deaths358 reports
- TotalCare® Bed P1900QHill-Rom, Inc. · BED, AC-POWERED ADJUSTABLE HOSPITAL317 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.