Hill-Rom, Inc. · Hospital Bed · Class 2

All models of Centrella Bed (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Hill-Rom Centrella hospital bed has 1,878 adverse event reports on record, including 5 deaths and 21 injuries. Most reports describe the bed's material splitting, cutting or tearing, with some describing damaged power cords with exposed wires, unintended movement of the bed, and some mentioning falls. Two open recalls name this device: one concerns the bed exit alert possibly failing to reach the nurse call system, and the other concerns radio interference from a WatchCare device that may affect other medical equipment.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,831
Deaths reported
5
Injuries reported
23
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1,831 adverse event reports that list this exact device (DI 00887761985162), received between Feb 26, 2021 and Sep 16, 2026 — classified as 5 deaths, 23 injuries, 1,803 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (1,298 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 25 reportsJan 2024: 35 reportsFeb 2024: 5 reportsMar 2024: 42 reportsApr 2024: 50 reportsMay 2024: 39 reportsJun 2024: 49 reportsJul 2024: 53 reportsAug 2024: 44 reportsSep 2024: 41 reportsOct 2024: 26 reportsNov 2024: 53 reportsDec 2024: 38 reportsJan 2025: 65 reportsFeb 2025: 51 reportsMar 2025: 38 reportsApr 2025: 50 reportsMay 2025: 44 reportsJun 2025: 56 reportsJul 2025: 45 reportsAug 2025: 34 reportsSep 2025: 21 reportsOct 2025: 42 reportsNov 2025: 72 reportsDec 2025: 58 reportsJan 2026: 50 reportsFeb 2026: 46 reportsMar 2026: 36 reportsApr 2026: 60 reportsMay 2026: 81 reportsJun 2026: 40 reportsJul 2026: 70 reportsAug 2026: 19 reportsSep 2026: 3 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202681
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 18, 2024Class IIOpen, Classified

    The beds have a potential for an interface issue with the Rauland Responder 5 Nurse Call System where the bed exit system will alert at the bed, but may fail to send a remote alert through the nurse call system.

    Z-0228-2025 · Root cause: Software design · Recall details · FDA record ↗

  2. Sep 30, 2022Class IOpen, Classified

    Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).

    Z-0141-2023 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Split, Cut or Torn1,298
  • Unintended Movement315
  • No Audible Alarm84
  • Activation Problem54
  • Material Frayed18
  • Defective Alarm15
  • Appropriate Device Problem Term/Code Not Available11
  • False Alarm9

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,794
  • Fall17
  • Bone Fracture(s)6
  • Pressure Sore/Ulcer5
  • Hematoma3
  • No Consequences Or Impact To Patient3
  • Unspecified Tissue Injury3
  • Hip Fracture2

Recent adverse event reports

The 25 most recent of 1,831 reports, in the reporter's own words as published by the FDA.

  • InjurySep 16, 2026Defective Alarm

    It was reported that a patient fell from a Centrella Smart+ Bed because the bed exit alarm did not go off. Following the fall, the patient underwent surgery for the reported hip fracture. No additional information regarding the injury, treatment details, or patient outcome was provided.\A Baxter technician inspected the bed and tested the bed exit feature. The bed was found to be functioning as designed. No further information was available at the time of this report.

    Patient problems: Fall

    MAUDE report 50796627 ↗
  • MalfunctionSep 10, 2026Unintended Movement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50716622 ↗
  • InjurySep 9, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that patient developed necrosis and ultimately requiring hip replacement surgery due to the Centrella Smart+ Bed. The patient stated that, due to quadriplegia, he was unable to independently reposition himself and relied on hospital staff to reposition him approximately every two hours. Despite these interventions, the patient alleged that the mattress was excessively firm and ineffective in preventing pressure-related injuries. The patient further alleged that prolonged use of the bed and mattress contributed to impaired blood flow to the left hip, resulting in tissue necrosis and ultimately requiring hip replacement surgery.\The patient reported been admitted to and discharged from the hospital on multiple occasions over an extended period and was unable to identify a specific bed or mattress associated with the reported event.\The facility's Chief Hospital Officer reported that the patient was well known to the facility and had previously raised multiple concerns during prior admissions. The facility indicated that investigations had been conducted regarding the patient's allegations and that those allegations were determined to be unsubstantiated.\The exact caus

    Patient problems: Necrosis

    MAUDE report 50700426 ↗
  • MalfunctionAug 14, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50257790 ↗
  • MalfunctionAug 14, 2026Activation Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50245680 ↗
  • MalfunctionAug 14, 2026Unintended Movement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50245222 ↗
  • MalfunctionAug 13, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50242567 ↗
  • MalfunctionAug 12, 2026Material Split, Cut or Torn

    Baxter received a report stating that CENTRELLA MED-SURG main and auxiliary power cord had damage with exposed copper wires. This occurred before use. There was no patient/user injury, medical intervention, symptom, or adverse event reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50227415 ↗
  • MalfunctionAug 12, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50227059 ↗
  • MalfunctionAug 12, 2026Flare or Flash; Material Split, Cut or Torn

    Baxter received a report stating that CENTRELLA MED-SURG main and auxiliary power cord had burnt damage with exposed copper wires. The process step at which the issue occurred was not specified. There was no patient/user injury, medical intervention, symptom, or adverse event reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50226989 ↗
  • InjuryAug 11, 2026No Audible Alarm

    It was reported that a patient fell from a Centrella Smart+ Bed and sustained a fractured hip. It was reported that the patient got out of bed to use the restroom and fell while returning to bed. The customer reported that the bed exit feature was not set correctly at the time of the event. The customer further stated that the patient weighed approximately 65 pounds and may have been below the minimum weight range specified for the bed. The customer confirmed that the event was not due to a device failure and declined to provide additional information regarding the injury or any medical or surgical treatment.\The Centrella Smart+ Bed is intended for use in healthcare environments as a patient support system to prevent and/or treat pressure injuries. The bed is intended for patient populations weighing at least 70 pounds (32 kg) and is capable of supporting patients up to 500 pounds (227 kg). The bed is equipped with a Bed Exit Alert System that utilizes data from the bed scale to detect patient movement based on the selected bed exit mode. The IFU further states that caregivers must perform a New Patient zero procedure before admitting a patient to the bed, as failure to do so may

    Patient problems: Fall

    MAUDE report 50216757 ↗
  • MalfunctionAug 11, 2026Unintended Movement

    Baxter received a report stating that CENTRELLA MED-SURG all four brake casters were not holding . The process step at which the issue occurred was not specified. There was no patient/user injury, medical intervention, symptom, or adverse event reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50202735 ↗
  • MalfunctionAug 11, 2026Unintended Movement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50202476 ↗
  • MalfunctionAug 10, 2026Material Split, Cut or Torn

    Baxter received a report that the CENTRELLA MED-SURG power cord had damage with exposed copper wires. Warehouse repair. This occurred after patient use. There was no patient/user injury, medical intervention, symptom, or adverse event reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50193793 ↗
  • MalfunctionAug 7, 2026Unintended Movement

    Baxter received a report stating that CENTRELLA MED-SURG brake casters were not holding. This occurred during BioMed Testing. There was no patient/user injury, medical intervention, symptom, or adverse event reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50184409 ↗
  • MalfunctionAug 7, 2026No Audible Alarm

    Baxter received a report stating that CENTRELLA MED-SURG bed exit alarm was alarming only at the bed and not at the nurse's station. The issue occurred before use. There was no patient/user injury, medical intervention, symptom, or adverse event reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50184298 ↗
  • MalfunctionAug 7, 2026No Audible Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50175506 ↗
  • MalfunctionAug 6, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50168461 ↗
  • MalfunctionAug 6, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50168309 ↗
  • MalfunctionAug 6, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50168272 ↗
  • MalfunctionAug 6, 2026Material Split, Cut or Torn

    Baxter received a report stating that CENTRELLA MED-SURG power cord had damage with exposed copper wires. This occurred during Baxter servicing/testing. There was no patient/user injury, medical intervention, symptom, or adverse event reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50159082 ↗
  • MalfunctionAug 6, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50155993 ↗
  • MalfunctionJul 31, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50106924 ↗
  • MalfunctionJul 31, 2026No Audible Alarm

    Baxter received a report stating that CENTRELLA MED-SURG bed exit was not alarming. There was no patient/user injury, medical intervention, symptom, or adverse event reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50104834 ↗
  • MalfunctionJul 29, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50077697 ↗

FDA clearance and classification

The FDA regulates product code FNL (bed, ac-powered adjustable hospital) under 21 CFR 880.5100 as a Class II device. 21 CFR 880.5100 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Centrella™ Bed P7900B

Has the Centrella™ Bed P7900B been recalled?

Yes. 2 FDA recall records name this device's identifier: 2 still open. The most recent, Z-0228-2025 (Class II), began on Sep 18, 2024: The beds have a potential for an interface issue with the Rauland Responder 5 Nurse Call System where the bed exit system will alert at the bed, but may fail to send a remote alert through the nurse call system. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Centrella™ Bed P7900B?

The FDA's MAUDE database lists 1,831 adverse event reports for this device, received between Feb 26, 2021 and Sep 16, 2026. In those reports, the most common specific device problems are “Material Split, Cut or Torn” (1,298), “Unintended Movement” (315) and “No Audible Alarm” (84); the most common patient problems are “Fall” (17), “Bone Fracture(s)” (6) and “Pressure Sore/Ulcer” (5). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Centrella™ Bed P7900B?

5 reports are classified as deaths, 23 as injuries and 1,803 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Centrella™ Bed P7900B FDA approved?

Devices of this type (product code FNL) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Hill-Rom, Inc.

Is the Centrella™ Bed P7900B MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the Centrella™ Bed P7900B, and is it still sold?

The FDA's device database lists Hill-Rom, Inc. as the company responsible for it (brand Centrella™ Bed, model P7900B). Its record shows it as in commercial distribution.