Centrella™ Bed P7900BRecalls & reported problems

Hill-Rom, Inc. · Hospital Bed · Class 2

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Hill-Rom Centrella hospital bed has 1,878 adverse event reports on record, including 5 deaths and 21 injuries. Most reports describe the bed's material splitting, cutting or tearing, with some describing damaged power cords with exposed wires, unintended movement of the bed, and some mentioning falls. Two open recalls name this device: one concerns the bed exit alert possibly failing to reach the nurse call system, and the other concerns radio interference from a WatchCare device that may affect other medical equipment.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,937
Deaths reported
5
Injuries reported
23
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1,937 adverse event reports that list this exact device (DI 00887761985162), received between Feb 26, 2021 and Sep 30, 2026 — classified as 5 deaths, 23 injuries, 1,909 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (1,378 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 29 reportsDec 2023: 25 reportsJan 2024: 35 reportsFeb 2024: 5 reportsMar 2024: 42 reportsApr 2024: 50 reportsMay 2024: 39 reportsJun 2024: 49 reportsJul 2024: 53 reportsAug 2024: 44 reportsSep 2024: 41 reportsOct 2024: 26 reportsNov 2024: 53 reportsDec 2024: 38 reportsJan 2025: 65 reportsFeb 2025: 51 reportsMar 2025: 38 reportsApr 2025: 50 reportsMay 2025: 44 reportsJun 2025: 56 reportsJul 2025: 45 reportsAug 2025: 34 reportsSep 2025: 21 reportsOct 2025: 42 reportsNov 2025: 72 reportsDec 2025: 58 reportsJan 2026: 50 reportsFeb 2026: 46 reportsMar 2026: 36 reportsApr 2026: 62 reportsMay 2026: 82 reportsJun 2026: 40 reportsJul 2026: 70 reportsAug 2026: 29 reportsSep 2026: 96 reportsOct 2026: 0 reports20242025202696
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 18, 2024Open, Classified

    The beds have a potential for an interface issue with the Rauland Responder 5 Nurse Call System where the bed exit system will alert at the bed, but may fail to send a remote alert through the nurse call system.

    Z-0228-2025 · Root cause: Software design · FDA record ↗

  2. Sep 30, 2022Open, Classified

    Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).

    Z-0141-2023 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Material Split, Cut or Torn1,378
  • Unintended Movement332
  • No Audible Alarm90
  • Activation Problem60
  • Material Frayed20
  • Defective Alarm15
  • Appropriate Device Problem Term/Code Not Available11
  • False Alarm9

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,900
  • Fall17
  • Bone Fracture(s)6
  • Pressure Sore/Ulcer5
  • Hematoma3
  • No Consequences Or Impact To Patient3
  • Unspecified Tissue Injury3
  • Hip Fracture2

Recent adverse event reports

The 25 most recent of 1,937 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026No Audible Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951554 ↗
  • MalfunctionSep 29, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50927589 ↗
  • MalfunctionSep 29, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50927581 ↗
  • MalfunctionSep 28, 2026Unintended Movement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50915982 ↗
  • MalfunctionSep 24, 2026Unintended Movement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50890505 ↗
  • MalfunctionSep 24, 2026Material Split, Cut or Torn

    Baxter received a report stating that CENTRELLA MED-SURG power cord had cut split damage with exposed copper wires. The process step at which the issue occurred was not specified. There was no patient/user injury, medical intervention, symptom, or adverse event reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50885296 ↗
  • MalfunctionSep 24, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884502 ↗
  • MalfunctionSep 24, 2026Material Split, Cut or Torn

    Baxter received a report stating that CENTRELLA MED-SURG main and auxiliary power cord were stripped with exposed copper wires. The process step at which the issue occurred was not specified. There was no patient/user injury, medical intervention, symptom, or adverse event reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50878519 ↗
  • MalfunctionSep 24, 2026No Audible Alarm

    Baxter received a report stating that CENTRELLA MED-SURG bed communication was not alarming. The process step at which the issue occurred was not specified. There was no patient/user injury, medical intervention, symptom, or adverse event reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50878516 ↗
  • MalfunctionSep 24, 2026Material Split, Cut or Torn

    Baxter received a report stating that CENTRELLA MED-SURG auxiliary power cord had damage with exposed copper wires. This occurred after patient use. There was no patient/user injury, medical intervention, symptom, or adverse event reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50878513 ↗
  • MalfunctionSep 24, 2026Material Split, Cut or Torn

    Baxter received a report that CENTRELLA MED-SURG had auxiliary power cord damaged with copper wire exposed. There was no patient/user injury, medical intervention, symptom, or adverse event reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50878040 ↗
  • MalfunctionSep 23, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50875501 ↗
  • MalfunctionSep 22, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50861059 ↗
  • MalfunctionSep 22, 2026Unintended Movement

    Baxter received a report stating that CENTRELLA MED-SURG brake casters were not holding at the foot end of bed. The process step at which the issue occurred was not specified. There was no patient/user injury, medical intervention, symptom, or adverse event reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50853554 ↗
  • MalfunctionSep 22, 2026Unintended Movement

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50853420 ↗
  • MalfunctionSep 22, 2026Unintended Movement

    Baxter received a report stating that CENTRELLA MED-SURG brake casters were not holding, still rolling when brakes were set . The process step at which the issue occurred was not specified. There was no patient/user injury, medical intervention, symptom, or adverse event reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50853419 ↗
  • MalfunctionSep 18, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50831034 ↗
  • MalfunctionSep 18, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50829904 ↗
  • MalfunctionSep 18, 2026Material Split, Cut or Torn

    Baxter received a report stating that CENTRELLA MED-SURG power cord had frayed damage with exposed copper wires. This occurred during Baxter servicing/testing. There was no patient/user injury, medical intervention, symptom, or adverse event reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812897 ↗
  • MalfunctionSep 18, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812861 ↗
  • MalfunctionSep 17, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50812224 ↗
  • MalfunctionSep 17, 2026No Audible Alarm

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50810751 ↗
  • MalfunctionSep 17, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50810623 ↗
  • MalfunctionSep 17, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50810351 ↗
  • MalfunctionSep 17, 2026Material Split, Cut or Torn

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50810320 ↗