Skip to content

ook snow FL36 ook snow (REF FL36)Recalls & reported problems

Umano Medical Inc · The ook snow is an AC-powered medical bed that offers an optional bed status monitoring system. The bed maximum safe working load is of 600lb and a service life of 15 years. · Class 2

All models of ook snow (7)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00670482000500), received between Oct 20, 2022 and Jan 15, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Communication or Transmission Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Communication or Transmission Problem2
  • Electrical Shorting2
  • Defective Alarm1
  • Detachment of Device or Device Component1
  • Improper or Incorrect Procedure or Method1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Bruise/Contusion1
  • Fall1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 15, 2025Defective Alarm; Improper or Incorrect Procedure or Method; Defective Alarm; Improper or Incorrect Procedure or Method

    ON (B)(6) 2024, A CUSTOMER REPORTED TO THE MANUFACTURER'S TECHNICAL SERVICE THAT THE BED EXIT MONITORING SYSTEM WAS TRIGGERING AN ALARM IMMEDIATELY UPON ARMING. THE SITE'S TECHNICIAN INITIATED TROUBLESHOOTING, BUT THE ISSUE REMAINED UNRESOLVED. TROUBLESHOOTING WAS SCHEDULED TO CONTINUE THE FOLLOWING DAY, AND THE MEDICAL STAFF CONTINUED TO USE THE DEVICE. ON (B)(6) 2024, DURING TROUBLESHOOTING, THE SITE'S TECHNICIAN INFORMED THE MANUFACTURER THAT A PATIENT HAD FALLEN FROM THE BED THE PREVIOUS NIGHT (EXACT DATE AND TIME NOT DISCLOSED). THE BED ACTIVITY LOG CONFIRMED THAT THE BED EXIT MONITORING SYSTEM WAS ACTIVATED AT 21:03 AND AN ALARM WAS TRIGGERED AT 8:41.

    Patient problems: Bruise/Contusion; Fall

    MAUDE report 21162327 ↗
  • MalfunctionMar 5, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    CUSTOMER CALLED THE TECHNICAL SERVICE OF MANUFACTURER TO REPORT ABOUT A BED WITH TWO BROKEN BRAKE BARS. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18837556 ↗
  • MalfunctionJun 27, 2023Communication or Transmission Problem; Communication or Transmission Problem

    THE MANUFACTURER WAS CONTACTED BY CUSTOMER ALLEGING THAT THE NURSE CALL SIGNAL WAS NOT BEING RECEIVED THROUGH THE NURSE CALL SYSTEM. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17209070 ↗
  • MalfunctionDec 1, 2022Communication or Transmission Problem; Communication or Transmission Problem

    TECHNICAL SERVICE OF THE MANUFACTURER WAS INFORMED THAT THE BRAKE BED STATUS OF HOSPITAL BED WORKS INTERMITTENTLY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15898175 ↗
  • MalfunctionOct 20, 2022Electrical Shorting; Electrical Shorting

    CUSTOMER REPORTED TO UMANO REPRESENTATIVE THAT THE BED IS OVERLOADING THE NURSE CALL CIRCUIT BOARD MAKING IT NON FUNCTIONAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15638143 ↗
  • MalfunctionOct 20, 2022Electrical Shorting; Electrical Shorting

    CUSTOMER REPORTED TO UMANO REPRESENTATIVE THAT THE BED IS OVERLOADING THE NURSE CALL CIRCUIT BOARD MAKING IT NON FUNCTIONAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15638017 ↗

Other Bed, Ac-Powered Adjustable Hospital devices

All Bed, Ac-Powered Adjustable Hospital devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page