TSRH® Spinal System 828-083Recalls & reported problems
Medtronic Sofamor Danek, Inc. · ROD 828-083 TSRH 6.35 TI ALLOY L51 · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00673978048680), received between Nov 9, 2017 and Nov 4, 2022 — classified as 3 injuries.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break3
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
- Pain1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryNov 4, 2022Break; Break
INFORMATION WAS RECEIVED FROM A REGULATORY BODY REGARDING A DEVICE USED FOR NON-CERVICAL SPINE POSTERIOR PEDICLE FIXATION IN A SKELETALLY MATURE PATIENT. IT WAS REPORTED THAT AFTER 9 DAYS OF PAIN AND WEAKNESS OF BOTH LOWER EXTREMITIES DUE TO HEAVY OBJECTS HITTING THE LOWER BACK, THE PATIENT WAS ADMITTED TO OUR HOSPITAL FOR TREATMENT ON (B)(6) 2020. ON (B)(6) 2020, THE PATIENT UNDERWENT "OPEN REDUCTION, FUSION AND INTERNAL FIXATION OF LUMBAR 3 VERTEBRAL FRACTURES + POSTERIOR DECOMPRESSION OF SPINAL CANAL". A TOTAL OF 8 SCREWS, 8 NUTS AND 1 TITANIUM ALLOY ROD WERE IMPLANTED DURING THE SURGERY. THE SURGERY WENT WELL AND THE PATIENT WAS DISCHARGED FROM THE HOSPITAL ON (B)(6) 2021. ON (B)(6) 2022, THE PATIENT'S RE-EXAMINATION CT IN OUR HOSPITAL SHOWED THAT THE TITANIUM ROD WAS BROKEN. ON (B)(6) 2022, THE PATIENT WAS ADMITTED TO (B)(6) FOR TREATMENT. ON (B)(6) 2022, THE PATIENT UNDERWENT "POSTERIOR LUMBAR DECOMPRESSION, BONE GRAFTING AND INTERNAL FIXATION", AND THE BROKEN ROD WAS TAKEN OUT SUCCESSFULLY. ON (B)(6) 2022, THE PATIENT WAS DISCHARGED FROM HOSPITAL. THIS PRODUCT NEEDED TO BE DISINFECTED BEFORE USE, AND HAD NO EXPIRATION DATE. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARD
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15733415 ↗InjuryJan 8, 2021Break; Break
INFORMATION WAS RECEIVED FIELD REPRESENTATIVE REGARDING PATIENT WITH POSTERIOR INSTRUMENTED ARTHRODESE T12 S1 FOR A DEGENERATIVE DISCOPATHY INVOLVED IN SYNTHESIS AND REPLACEMENT. LEVELS IMPLANTED - T12 S1. IT WAS REPORTED DURING NORMAL USE, PATIENT REPORTED THAT IN (B)(6) HE PERCEIVED A LOUD DEAF NOISE IN HIS BACK AND SEVERE PAIN ON THE DAY OF HOSPITALIZATION. RX TESTS WERE CARRIED OUT AND THE 2 RODS WERE FOUND TO BE BROKEN. THE 2 BROKEN RODS AND THEIR INTACT SCREWS HAVE BEEN REMOVED AND REPLACED. EVEN IF THE BARS HAD BEEN INTACT THEY WOULD HAVE REPLACED THEM EQUALLY. THERE WAS NO BROKEN FRAGMENTS LEFT IN THE PATIENT. PATIENT IS ALIVE - NO INJURY. ON 2020-JAN-04, RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO MALFUNCTION WITH THE REPORTED BONE SCREWS. PATIENT WAS NOT HOSPITALIZED PROLONG AS A RESULT OF THE EVENT. THERE WAS NO DELAY IN OVERALL PROCEDURE TIME. THERE WERE NO COMPLICATIONS REPORTED AS A RESULT OF THE EVENT. PROCEDURE WAS COMPLETED IN PATIENT WITHOUT PROBLEM. CUSTOMER REFUSED TO RETURN THE PRODUCT AS THEY INTENDED TO KEEP IT.
Patient problems: Pain
MAUDE report 11138902 ↗InjuryNov 9, 2017Break; Break
PROCEDURE: LUMBAR POSTERIOR FUSION IT WAS REPORTED THAT THE PATIENT UNDERWENT OCCIPITO-CERVICAL FUSION SURGERY. POST-OP, LESS THAN 1 MONTH AFTER SURGERY, THE LEFT JOINT ROD RUPTURED AND BROKE. NO FUSION OCCURRED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7015589 ↗
Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices
- EXPEDIUM 179702000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL439 reports
- EXPEDIUM VERSE 199721001Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL227 reports
- XIA 3 48230000Stryker Corporation · APPLIANCE, FIXATION, SPINAL INTERLAMINAL193 reports
- VIPER 186715000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL1 death167 reports
- NA 04.614.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL80 reports
- EXPEDIUM VIPER 196789480Depuy Spine, LLC · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- NA 04.615.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- Arsenal Deformity Spinal Fixation System 47127Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL66 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page