ENDOGATOR hybrid irrigation tubing 100609Recalls & reported problems
Medivators Inc. · Class 2
All models of ENDOGATOR hybrid irrigation tubing (5)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00677964052354), received between Aug 15, 2018 and Feb 23, 2024 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash7
- Adverse Event Without Identified Device or Use Problem1
- Device Handling Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Aspiration/Inhalation1
- Insufficient Information1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 23, 2024Leak/Splash; Leak/Splash
THE USER FACILITY REPORTED THAT WATER LEAKED FROM THE TUBE CONNECTING TO CO2 CONNECTOR WHILE USING THEIR ENDOGATOR HYBRID IRRIGATION TUBING. NO REPORT OF INJURY OR PROCEDURE DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18773840 ↗MalfunctionDec 13, 2023Leak/Splash; Leak/Splash
THE USER FACILITY REPORTED THAT WATER LEAKS INTO THE AIR TUBING WHILE USING THEIR ENDOGATOR HYBRID IRRIGATION TUBING. NO REPORT OF INJURY OR PROCEDURE DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18319586 ↗MalfunctionDec 13, 2023Leak/Splash; Leak/Splash
THE USER FACILITY REPORTED THAT WATER LEAKS FROM THE ENDOSCOPE WHILE USING THEIR ENDOGATOR HYBRID IRRIGATION TUBING DURING THE AIR/WATER TEST. NO REPORT OF INJURY OR PROCEDURE DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18317386 ↗MalfunctionDec 12, 2023Leak/Splash; Leak/Splash
THE USER FACILITY REPORTED THAT WATER LEAKS FROM THE ENDOSCOPE WHILE USING THEIR ENDOGATOR HYBRID IRRIGATION TUBING DURING THE AIR/WATER TEST. NO REPORT OF INJURY OR PROCEDURE DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18315108 ↗MalfunctionDec 8, 2023Leak/Splash; Leak/Splash
THE USER FACILITY REPORTED THAT WATER LEAKS FROM THE ENDOSCOPE WHILE USING THEIR ENDOGATOR HYBRID IRRIGATION TUBING. THIS HAS OCCURRED DURING SET-UP AND PATIENT PROCEDURES. THE PROCEDURES WERE COMPLETED SUCCESFULLY. NO REPORT OF INJURY OR PROCEDURE DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18291681 ↗MalfunctionJul 13, 2023Leak/Splash; Device Handling Problem; Leak/Splash; Device Handling Problem
THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE WHICH INCLUDED USE OF THE ENDOGATOR HYBRID TUBING, STERILE WATER LEAKED FROM THE ENDOSCOPE OBSCURING THE PHYSICIAN'S FIELD OF VISION. NO REPORT OF INJURY OR PROCEDURE DELAY.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 17316600 ↗MalfunctionApr 12, 2023Leak/Splash; Leak/Splash
THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE WHICH INCLUDED USE OF THE ENDOGATOR HYBRID TUBING, STERILE WATER LEAKED FROM THE ENDOSCOPE. THE PATIENT DID NOT REQUIRE ADDITIONAL MEDICAL INTERVENTION. NO REPORT OF INJURY OR PROCEDURE DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16725220 ↗InjuryAug 15, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A PATIENT ASPIRATED DURING AN UPPER GI ENDOSCOPY PROCEDURE USING AN OLYMPUS HQ190 ENDOSCOPE AND WAS HOSPITALIZED FOR ASPIRATION PNEUMONIA. MEDIVATORS ENDOGATOR HYBRID IRRIGATION TUBING WAS BEING USED DURING THE PROCEDURE AND THE FACILITY REPORTED THAT THE TUBING WAS LEAKING AND CAUSED THE PATIENT ADVERSE EVENT.
Patient problems: Aspiration/Inhalation
MAUDE report 7787005 ↗
Other Endoscopic Irrigation/Suction System devices
- OrcaPod™ SUV-628Boston Scientific Corporation · endoscopic irrigation/suction system79 reports
- Suction Valve EN10232GA Health Company Limited · Endoscopic Irrigation/Suction System1 recall0 reports
- andorate® A/W Channel Cleaning Adapter + 4-piece Valve Kit GAR3846GA Health Company Limited · Endoscopic Irrigation/Suction System1 recall0 reports
- BullDog® 1203-71Diversatek Healthcare, Inc. · Endoscopic Irrigation/Suction System1 recall0 reports
- ERBEFLO CleverCap® 20325-201Erbe Medical, LLC · Endoscopic Irrigation/Suction System1 recall0 reports
- 2-piece valve kit (Suction and Air/Water Valve) EN10229.2GA Health Company Limited · Endoscopic Irrigation/Suction System1 recall0 reports
- 4-piece valve kit(Suction, Air/Water, Biopsy Valves & Auxiliary Water Connector) EN10229.1GA Health Company Limited · Endoscopic Irrigation/Suction System1 recall0 reports
- ERBEFLO CleverCap® 20325-203Erbe Medical, LLC · Endoscopic Irrigation/Suction System1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page