DEFENDO Valve and Connector Kit 100310 (UDI 00677964052453)Recalls & reported problems
Medivators Inc. · DEFENDO Valve and Connetor Kit 4-piece kit Single-Use Air Water and Suction and Biopsy Valves and Endogator Connector · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00677964052453), received between Sep 1, 2017 and Feb 26, 2026 — classified as 3 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Device Contamination with Body Fluid1
- Device Handling Problem1
- Fluid/Blood Leak1
- Improper or Incorrect Procedure or Method1
- Insufficient Device Problem Information1
- Leak/Splash1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Airway Obstruction1
- Exposure to Body Fluids1
- Muscle Spasm(s)1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 26, 2026Leak/Splash
THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE WATER WAS LEAKING FROM THEIR DEFENDO DISPOSABLE AIR/WATER VALVES. USER FACILITY PERSONNEL CHANGED THE VALVES, TUBING, AND LOWERED THE WATER LEVEL WITHIN THEIR DEFENDO DISPOSABLE AIR/WATER VALVE RESULTING IN A PROCEDURE DELAY. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24460796 ↗MalfunctionJun 13, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
THE USER FACILITY REPORTED VIA VIGILANCE REPORT TO ANSM THAT THEIR DEFENDO SUCTION VALVE WAS NOT WORKING PRIOR TO USE IN A PROCEDURE. USER FACILITY PERSONNEL OBTAINED A NEW VALVE TO COMPLETE THE PROCEDURE. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19528495 ↗MalfunctionNov 17, 2023Break; Break
THE DISTRIBUTOR REPORTED THAT THE SUCTION VALVE FROM THEIR DEFENDO VALVES & KITS WAS STUCK IN THE DEPRESSED POSITION DURING A PATIENT PROCEDURE. THE VALVE BROKE AS USER FACILITY PERSONNEL ATTEMPTED TO PULL IT OUT CAUSING HALF OF THE VALVE TO BECOME STUCK IN THE ENDOSCOPE. A PROCEDURE DELAY OCCURRED AS USER FACILITY PERSONNEL EXCHANGED THE ENDOSCOPE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18159937 ↗MalfunctionAug 25, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE USER FACILITY REPORTED THAT DURING MULTIPLE PATIENT PROCEDURES WHICH INCLUDED USE OF THE DEFENDO - OLYMPUS VALVE KIT, THE TISSUE SAMPLE GOT STUCK IN THE BIOPSY VALVE WHEN REMOVING THE FORCEPS FROM THE BIOPSY VALVE RESULTING IN A PROCEDURE DELAY. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15297929 ↗InjuryFeb 15, 2019Fluid/Blood Leak; Fluid/Blood Leak
THE FACILITY REPORTED THAT DURING AN EGD ENDOSCOPY PROCEDURE, THE PATIENT EXPERIENCED VOCAL CORD MUSCLE SPASMS. THE DISPOSABLE DEFENDO AIR/WATER VALVE WAS USED DURING THE PROCEDURE AND WAS REPORTED TO HAVE CONTRIBUTED TO THE EVENT.
Patient problems: Airway Obstruction; Muscle Spasm(s)
MAUDE report 8343345 ↗InjuryJun 13, 2018Improper or Incorrect Procedure or Method; Device Handling Problem; Improper or Incorrect Procedure or Method; Device Handling Problem
A MEDIVATORS ACCOUNT MANAGER REPORTED A FACILITY REPROCESSING, THEN REUSING DEFENDO DISPOSABLE AIR/WATER, SUCTION, AND BIOPSY VALVES BETWEEN USES IN PATIENT PROCEDURE. THE DEFENDO AIR/WATER, SUCTION AND BIOPSY VALVE IS INTENDED FOR SINGLE USE, THUS POTENTIAL FOR PATIENT CROSS-CONTAMINATION.
Patient problems: Exposure to Body Fluids
MAUDE report 7598694 ↗InjurySep 1, 2017Device Contamination with Body Fluid; Device Contamination with Body Fluid
A FACILITY REPORTED BLOOD/FLUID BACKFLOW IN THE TUBING TO THE PUMP DURING A PROCEDURE USING MEDIVATORS AUXILIARY WATER JET CONNECTOR. THERE IS POTENTIAL FOR PATIENT CROSS CONTAMINATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6839309 ↗
Other Endoscope Channel Accessory devices
- RX Locking Device and Biopsy Cap M00545260Boston Scientific Corporation · Endoscope channel accessory216 reports
- Orca™ SUV-617Boston Scientific Corporation · Endoscope Channel Accessory1 recall64 reports
- OrcaPod™ SUV-629Boston Scientific Corporation · Endoscope channel accessory112 reports
- andorate® Disposable Endoscope Valves Set GAR004GA Health Company Limited · Endoscope Channel Accessory1 recall1 report
- BullDog 1203-72Diversatek Healthcare, Inc. · Endoscope Channel Accessory1 recall0 reports
- Ruhof Scope Valet Guardian 345SVVO4Ruhof Corporation · Endoscope Channel Accessory1 recall0 reports
- 3-piece valve kit (Suction, Air/Water and Biopsy Valves) EN10229GA Health Company Limited · Endoscope Channel Accessory1 recall0 reports
- AutoCap™ RX M00545100Boston Scientific Corporation · Endoscope channel accessory26 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page