DEFENDO Air/Water/Suction and Biopsy Valve 100316Recalls & reported problems

Medivators Inc. · 3-piece kit Single-Use Air Water, Suction and Biopsy Valves - Pentax · Class 2

All models of DEFENDO Air/Water/Suction and Biopsy Valve (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
16
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 00677964055928), received between Jun 10, 2022 and Jan 11, 2024 — classified as 16 malfunctions.

The most frequently reported specific device problem is “Difficult to Open or Close” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 3 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem7
  • Difficult to Open or Close4
  • Insufficient Device Problem Information3
  • Physical Resistance/Sticking3

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • Insufficient Information4
  • Hematoma1

Recent adverse event reports

The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 11, 2024Difficult to Open or Close; Difficult to Open or Close

    THE DISTRIBUTOR REPORTED THAT A SUCTION VALVE FROM THEIR DEFENDO VALVES & KITS "CAN NO LONGER BE RELEASED/CLOSED AFTER OPENING" DURING A PATIENT PROCEDURE RESULTING IN SUCTIONING OF THE MUCOUS MEMBRANE IN THE STOMACH AND BLEEDING. THE USER SWITCHED TO A DIFFERENT VALVE TO COMPLETE THE PROCEDURE. NO ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED.

    Patient problems: Insufficient Information

    MAUDE report 18499267 ↗
  • MalfunctionJan 11, 2024Difficult to Open or Close; Difficult to Open or Close

    THE DISTRIBUTOR REPORTED THAT A SUCTION VALVE FROM THEIR DEFENDO VALVES & KITS "CAN NO LONGER BE RELEASED/CLOSED AFTER OPENING" DURING A PATIENT PROCEDURE RESULTING IN SUCTIONING OF THE MUCOUS MEMBRANE IN THE STOMACH AND BLEEDING. THE USER SWITCHED TO A DIFFERENT VALVE TO COMPLETE THE PROCEDURE. NO ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED.

    Patient problems: Insufficient Information

    MAUDE report 18499076 ↗
  • MalfunctionJan 11, 2024Difficult to Open or Close; Difficult to Open or Close

    THE DISTRIBUTOR REPORTED THAT A SUCTION VALVE FROM THEIR DEFENDO VALVES & KITS "CAN NO LONGER BE RELEASED/CLOSED AFTER OPENING" DURING A PATIENT PROCEDURE RESULTING IN SUCTIONING OF THE MUCOUS MEMBRANE IN THE STOMACH AND BLEEDING. THE USER SWITCHED TO A DIFFERENT VALVE TO COMPLETE THE PROCEDURE. NO ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED.

    Patient problems: Insufficient Information

    MAUDE report 18499064 ↗
  • MalfunctionDec 5, 2023Difficult to Open or Close; Insufficient Device Problem Information; Difficult to Open or Close; Insufficient Device Problem Information

    THE DISTRIBUTOR REPORTED THAT SUCTION VALVES FROM THEIR DEFENDO VALVES & KITS "CAN NO LONGER BE RELEASED/CLOSED AFTER OPENING" DURING PATIENT PROCEDURES RESULTING IN SUCTIONING OF THE MUCOUS MEMBRANE IN THE STOMACH AND BLEEDING. THE USERS SWITCHED TO A DIFFERENT VALVE TO COMPLETE THE PROCEDURE. NO ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED. THE DISTRIBUTOR STATED THE VALVES WERE USED BY FOUR DIFFERENT DOCTORS; HOWEVER, THEY WERE UNABLE TO PROVIDE THE EXACT NUMBER OF EVENTS AND DEVICES AFFECTED.

    Patient problems: Insufficient Information

    MAUDE report 18266336 ↗
  • MalfunctionMay 26, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED VIA USER FACILITY MEDWATCH REPORT # (B)(4) THAT DURING PATIENT PROCEDURES WHICH INCLUDED USE OF A DEFENDO DISPOSABLE SUCTION VALVE, THE BUTTON OF THE VALVE WAS GETTING STUCK CAUSING CONSTANT SUCTION. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17011610 ↗
  • MalfunctionMay 18, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE INCLUDING USE OF A DEFENDO DISPOSABLE SUCTION VALVE, THERE WAS TOO MUCH SUCTION CAUSING THE ENDOSCOPE TO SUCTION TO THE PATIENT'S MUCOUS MEMBRANE. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16956577 ↗
  • MalfunctionApr 24, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE INCLUDING USE OF A DEFENDO DISPOSABLE SUCTION VALVE, THERE WAS TOO MUCH SUCTION CAUSING THE ENDOSCOPE TO SUCTION TO THE PATIENT. THE VALVE WAS REMOVED AND THE SUCTION STOPPED. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16797358 ↗
  • MalfunctionFeb 1, 2023Physical Resistance/Sticking; Physical Resistance/Sticking

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE INCLUDING USE OF A DEFENDO DISPOSABLE SUCTION VALVE, THERE WAS TOO MUCH SUCTION CAUSING THE ENDOSCOPE TO SUCTION TO THE PATIENT. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16280376 ↗
  • MalfunctionFeb 1, 2023Physical Resistance/Sticking; Physical Resistance/Sticking

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE INCLUDING USE OF A DEFENDO DISPOSABLE SUCTION VALVE, THERE WAS TOO MUCH SUCTION CAUSING THE ENDOSCOPE TO SUCTION TO THE PATIENT'S MUCOUS MEMBRANE. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16279935 ↗
  • MalfunctionJan 20, 2023Physical Resistance/Sticking; Physical Resistance/Sticking

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE INCLUDING USE OF A DEFENDO DISPOSABLE SUCTION VALVE, THERE WAS TOO MUCH SUCTION CAUSING THE ENDOSCOPE TO SUCTION TO THE PATIENT'S MUCOUS MEMBRANE. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16214530 ↗
  • MalfunctionNov 29, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE INCLUDING USE OF A DEFENDO DISPOSABLE SUCTION VALVE, THERE WAS TOO MUCH SUCTION CAUSING THE ENDOSCOPE TO SUCTION TO THE PATIENT'S MUCOUS MEMBRANE. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15878783 ↗
  • MalfunctionNov 15, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE INCLUDING USE OF A DEFENDO DISPOSABLE SUCTION VALVE, THERE WAS TOO MUCH SUCTION CAUSING THE ENDOSCOPE TO SUCTION TO THE PATIENT'S MUCOUS MEMBRANE. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15799055 ↗
  • MalfunctionOct 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE INCLUDING USE OF A DEFENDO DISPOSABLE SUCTION VALVE, THERE WAS TOO MUCH SUCTION CAUSING THE ENDOSCOPE TO SUCTION TO THE PATIENT'S MUCOUS MEMBRANE. THE VALVE WAS REMOVED TO STOP THE SUCTIONING. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15611273 ↗
  • MalfunctionJul 29, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE INCLUDING USE OF A DEFENDO DISPOSABLE SUCTION VALVE, THERE WAS TOO MUCH SUCTION CAUSING THE ENDOSCOPE TO SUCTION TO THE PATIENT'S MUCOUS MEMBRANE. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15129122 ↗
  • MalfunctionJul 29, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE INCLUDING USE OF A DEFENDO DISPOSABLE SUCTION VALVE, THERE WAS TOO MUCH SUCTION CAUSING THE ENDOSCOPE TO SUCTION TO THE PATIENT'S MUCOUS MEMBRANE. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15128619 ↗
  • MalfunctionJun 10, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE INCLUDING USE OF A DEFENDO DISPOSABLE SUCTION VALVE, THERE WAS TOO MUCH SUCTION RESULTING IN A SMALL HEMATOMA. NO FURTHER MEDICAL TREATMENT WAS REQUIRED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO REPORT OF PROCEDURE DELAY.

    Patient problems: Hematoma

    MAUDE report 14653995 ↗

Other Endoscope Channel Accessory devices

All Endoscope Channel Accessory devices →