Skip to content

N/A 1882900Recalls & reported problems

Medtronic Xomed, Inc. · DRILL 1882900 MINI TREPHINATION 2MM · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00681490047241), received between Jan 11, 2019 and Jan 11, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Fragmentation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Material Fragmentation1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 11, 2019Material Fragmentation; Material Fragmentation

    A HEALTH CARE PROVIDER (HCP) REPORTED VIA A MANUFACTURER REPRESENTATIVE THAT THERE WERE MULTIPLE BURS AFFECTED, ONE REPORTED TO HAVE BROKEN AND OTHERS NOT DRILLING OR "BLUNT" DURING THE PROCEDURE. THE PATIENT WAS ALIVE WITH NO INJURY. ON FOLLOW-UP, IT WAS REPORTED THAT THE PROCEDURE WAS COMPREHENSIVE ENDOSCOPIC SINUS SURGERY WITH FRONTAL TREPHINES. THE HCP HAD TO OPEN MORE THAN ONE DRILL ON A CASE AS COMPARED TO USUALLY JUST USING ONE DRILL PER CASE. THERE WERE FRAGMENTS DETACHED FROM THE DEVICE. THERE WAS A DELAY IN THE PROCEDURE BECAUSE THE DRILL WAS NOT DRILLING INTO THE PATIENT¿S BONE SO THE PROCEDURE TOOK LONGER.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8239305 ↗