Skip to content

VersaCross Access Solution VXSK0024 (REF NA)Recalls & reported problems

Baylis Médicale Cie Inc · 1 x VersaCross Transseptal Sheath (including 1 x J-tip Guidewire); 1 x VersaCross RF Wire (including 1 x Connector Cable (Single Use)) · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00685447011750), received between Mar 20, 2024 and Jun 24, 2024 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Difficult to Advance” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 10, 2025Open, Classified

    Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.

    Z-2103-2025 · Root cause: Packaging change control · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Difficult to Advance1
  • Material Frayed1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Cardiac Perforation1
  • Low Blood Pressure/ Hypotension1
  • Pericardial Effusion1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 24, 2024Material Frayed; Material Frayed

    IT HAS BEEN REPORTED THAT A VERSACROSS ACCESS SOLUTION KIT WAS SELECTED FOR USE FOR AN ATRIAL FIBRILLATION (A FIB) ABLATION PROCEDURE. DURING THE PROCEDURE, AS THE PHYSICIAN WAS ATTEMPTING TO GET ACCESS WITH VERSACROSS RF WIRE, HE WAS EXPERIENCING DIFFICULTY INSERTING THE VERSACROSS SHEATH INTO THE PATIENT'S RIGHT FEMORAL VEIN. THEREFORE, THE PHYSICIAN REMOVED THE VERSACROSS RF WIRE AND NOTICED THAT PART OF THE DEVICE WAS FRAYING AT ITS DISTAL END. THE RF WIRE WAS INSERTED THROUGH A NON-BOSTON SCIENTIFIC INTRODUCER NEEDLE PRIOR TO THE ISSUE BEING NOTED. THE RF WIRE WAS THEN REPLACED, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT COMPLICATIONS WERE REPORTED. PRODUCT IS NOT EXPECTED TO RETURN FOR ANALYSIS (DISPOSED).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19600686 ↗
  • MalfunctionApr 15, 2024Difficult to Advance; Difficult to Advance

    IT WAS REPORTED THAT DURING A PROCEDURE, A VERSACROSS ACCESS SOLUTION KIT WAS SELECTED FOR USE. IT WAS NOTED THAT THE DEVICE WAS NOT ABLE TO GO TRANSSEPTAL. NO FURTHER DETAILS WERE PROVIDED. THE PROCEDURE OUTCOME IS UNKNOWN. NO PATIENT COMPLICATIONS WERE REPORTED. THE DEVICE IS EXPECTED TO BE RETURNED FOR ANALYSIS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19105400 ↗
  • InjuryMar 20, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED A PERICARDIAL EFFUSION OCCURRED. DURING AN ATRIAL FIBRILLATION (A FIB) PROCEDURE, A VERSACROSS ACCESS SOLUTION KIT WAS SELECTED FOR USE. A PERICARDIAL EFFUSION (PE) WAS NOTED WHEN THE TRANSSEPTAL PUNCTURE (TSP) WAS BEING PERFORMED USING VERSACROSS SHEATH AND RADIO FREQUENCY WIRE. THE PE WAS CONFIRMED USING AN ULTRASOUND SCAN WITH A NON-BOSTON SCIENTIFIC CATHETER. HENCE, THE ABLATION WAS NOT COMPLETED. THE PATIENT BLOOD PRESSURE WAS DROPPED. A PERICARDIOCENTESIS WAS PERFORMED AND THE PATIENT WAS STABILIZED. THE DEVICE IS NOT EXPECTED TO BE RETURNED FOR ANALYSIS. EVENT REPORTED VIA MEDWATCH MDR REPORT # MW5151066.

    Patient problems: Cardiac Perforation; Low Blood Pressure/ Hypotension; Pericardial Effusion

    MAUDE report 18941532 ↗

Other Catheter, Septostomy devices

All Catheter, Septostomy devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.