Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00690103000276), received between Aug 23, 2024 and Aug 23, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 23, 2024Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED, AFTER INSERTING THIS SWAN-GANZ CATHETER, WHILE INFLATING, THE DOCTOR FELT A DROP OF AIR AND THE BALLOON WAS TORN. THE CATHETER WAS REMOVED FROM THE PATIENT'S BODY. THE BALLOON HAD BEEN INFLATED WITH THE DEDICATED SYRINGE THAT COMES IN THE KIT AND PRIOR TO INSERTION, THE DOCTOR HAD TESTED THE BALLOON IN A TUB OF WATER AND THE BALLOON WAS NORMAL. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION. AS PER EVALUATION RESULTS, THE BALLOON WAS FOUND TO BE RUPTURED NEXT TO DISTAL END. THE BALLOON EDGES APPEARED NOT TO MATCH UP AT THE RUPTURED REGION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20045655 ↗