Fem-Flex II Femoral Arterial Cannula FEMII020ARecalls & reported problems
Edwards Lifesciences LLC · FEM-FLEX II Femoral Arterial Cannula · Class 2
All models of Fem-Flex II Femoral Arterial Cannula (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00690103031287), received between Sep 28, 2021 and Sep 28, 2021 — classified as 1 injury.
The most frequently reported specific device problem is “Fracture” (1 report).
7 other recalls cover devices cleared under the same FDA submission (K140208, K200358); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture1
Patient problems
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjurySep 28, 2021Fracture; Fracture
IT WAS REPORTED THAT THE FEMII020A CANNULA SNAPPED/BROKE AFTER THREE DAYS AFTER THE INITIAL PLACEMENT. ADDITIONAL INFORMATION RECEIVED: THE PATIENT WAS ABOUT TO BE PREPPED AND DRAPED HOWEVER WE NOTICED SIGNIFICANT BLEEDING FROM THE RIGHT IJ CANNULA, WE NOTICED THAT THE CANNULA WAS CRACKED AS A RESULT WE QUICKLY STOPPED THE ECMO AND CLAMPED THAT CANNULA AND THIS WAS FOLLOWED BY SWITCHING THE CONFIGURATION OF VV ECMO TO A FEM FEM VV ECMO AND REMOVED THE RIGHT IJ FEMFLEX CANNULA. THE PRESSURE WAS HELD AND NO HEMATOMA DEVELOPED. THE CANNULA WAS REMOVED AND REPLACED WITH A 32 F MC3 CANNULA PLACED IN LEFT IJ. ON POD #4, THE PATIENT EXPIRED. THE PATIENT'S DEATH WAS NOT RELATED TO THE CATHETER.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 12541219 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K140208, K200358. One submission often covers many models, so check the product and lot details in each record.
- Z-2084-2025 · May 14, 2025 · Open, ClassifiedBrand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII018A Software Version: N/A Product Description: The Edwards FEMFLEXII Femoral Perfusion Ca
- Z-2082-2025 · May 14, 2025 · Open, ClassifiedBrand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII016A Software Version: N/A Product Description: The Edwards FEMFLEXII Femoral Perfusion Ca
- Z-2085-2025 · May 14, 2025 · Open, ClassifiedBrand Name: FEM-FLEXII Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII018AS Software Version: N/A Product Description: The Edwards FEMFLEXII Arterial Perfusion C
- Z-2083-2025 · May 14, 2025 · Open, ClassifiedBrand Name: FEM-FLEX II Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII016AS Software Version: N/A Product Description: The Edwards FEMFLEXII Femoral Perfusion C
- Z-3033-2018 · Mar 7, 2018 · TerminatedEdwards Lifesciences Femoral Venous Cannula, 20F X 55CM, resterilized.
- Z-3032-2018 · Mar 7, 2018 · TerminatedEdwards Lifesciences Femoral Venous Cannula, 18FR, resterilized.
- Z-1371-2015 · Mar 23, 2015 · TerminatedEdwards Lifesciences Fem-Flex II Femoral Arterial Cannula 8, 10, 12 French,Sterile EO, Rx Only, Manufacturer Edwards Lifesciences LLC, Irvine, CA.-Model Numbers: FEMII008A, FEMII0
Other CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices
- DLP™ 77010Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass1 recall154 reports
- EOPA™ 77422Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls19 reports
- DLP™ 77008Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls9 reports
- NA HY11W65R1Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls8 reports
- DLP® 12113Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls6 reports
- DLP® Aortic Root Cannula with Flow-Guard™ 11012Medtronic, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS2 recalls5 reports
- EOPA™ 77622Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls4 reports
- NA HY2V10R33Medtronic, Inc. · Catheter, cannula and tubing, vascular, cardiopulmonary bypass2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page