Skip to content

ThinFlex Single Stage Venous Drainage Cannula TF022O90Recalls & reported problems

Edwards Lifesciences LLC · THIN-FLEX SINGLE STAGE VENOUS DRAINAGE CANNULA · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00690103040890), received between Aug 6, 2025 and Aug 7, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Manufacturing, Packaging or Shipping Problem” (2 reports).

4 other recalls cover devices cleared under the same FDA submission (K120072); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Manufacturing, Packaging or Shipping Problem2
  • Particulates2
  • Dent in Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 7, 2025Particulates; Manufacturing, Packaging or Shipping Problem; Particulates; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT A TF022O90 VENOUS CANNULA WAS FOUND TO HAVE AN EXTRA PIECE OF PLASTIC DISTALLY, AT THE END OF THE WIRE BODY, BEFORE PLACING THE DEVICE IN THE PATIENT. THE SURGEON DECIDED TO MOVE FORWARD WITH USING THE DEVICE FOR A PVAC PROCEDURE. THE EXTRA PLASTIC PIECE WAS NOT REMOVED BY THE SURGEON PRIOR TO USE. THERE WAS NO IMPACT TO THE PATIENT OR THE PROCEDURE AS A RESULT OF THIS INCIDENT. PER ADDITIONAL INFO RECEIVED, THE REMAINING CANNULAS IN THE HOSPITAL INVENTORY BELONGING TO THE AFFECTED LOT WERE CHECKED AND THEY APPEARED FINE. THE DEVICES ARE LOOSELY STORED IN A BIN UNDER NORMAL HOSPITAL TEMPERATURE. THE TRAY IN WHICH THE CANNULA IS SEATED (PART OF PACKAGING) WAS STATED TO BE INTACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22734903 ↗
  • MalfunctionAug 6, 2025Particulates; Dent in Material; Manufacturing, Packaging or Shipping Problem; Particulates; Dent in Material; Manufacturing, Packaging or Shipping Problem

    IT WAS REPORTED THAT A TF022O90 VENOUS CANNULA WAS FOUND TO HAVE AN EXTRA PIECE OF PLASTIC DISTALLY, AT THE END OF THE WIRE BODY, BEFORE PLACING THE DEVICE IN THE PATIENT. THE SURGEON DECIDED TO MOVE FORWARD WITH USING THE DEVICE FOR A FONTAN PROCEDURE AND PLUCKED OFF THE PLASTIC AFTER THE CASE WAS OVER. THERE WAS NO IMPACT TO THE PATIENT OR THE PROCEDURE AS A RESULT OF THIS INCIDENT. THE TRAY WAS STATED TO BE INTACT. PER ADDITIONAL INFO RECEIVED, THE REMAINING CANNULAS IN THE HOSPITAL INVENTORY BELONGING TO THE AFFECTED LOT WERE CHECKED AND THEY APPEARED FINE. THE DEVICES ARE LOOSELY STORED IN A BIN UNDER NORMAL HOSPITAL TEMPERATURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22734218 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K120072. One submission often covers many models, so check the product and lot details in each record.

  • Z-3029-2018 · Mar 7, 2018 · Terminated
    Edwards Lifesciences Thin-Flex Single Stage Venous Drainage Cannula, 38FRX40CM, resterilized.
  • Z-3031-2018 · Mar 7, 2018 · Terminated
    Edwards Lifesciences Avid Dual Stage Venous Drainage Cannula, 29/37F (9.6/12.3MM) X 37CM (14.5"), resterilized.
  • Z-3030-2018 · Mar 7, 2018 · Terminated
    Edwards Lifesciences Thin-Flex Single Stage Venous Drainage Cannula, 34F (11.3MM) X 40CM (16"), resterilized.
  • Z-0398-2016 · Nov 4, 2015 · Terminated
    Thin-Flex Venous Cannula, Model Number:TF292902A

Other CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices

All CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page