Skip to content

Thin-Flex Single Stage Venous Drainage Cannula TF028LRecalls & reported problems

Edwards Lifesciences LLC · THIN-FLEX SINGLE STAGE VENOUS DRAINAGE CANNULA · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00690103041002), received between Sep 9, 2025 and Sep 9, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).

4 other recalls cover devices cleared under the same FDA submission (K120072); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Device Interaction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 9, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT A TF028L VENOUS CANNULA WAS EXCHANGED WITH A LARGER TF030L CANNULA DUE TO POOR VENOUS DRAINAGE FOLLOWING THE INITIATION OF CARDIOPULMONARY BYPASS. THERE WERE NO PATIENT ADVERSE EVENTS REPORTED. THE PATIENT OUTCOME WAS REPORTED AS UNDER TREATMENT. PER THE REPORTED INFORMATION, THE DEVICE WAS USED DURING MITRAL VALVE REPLACEMENT AND CONCOMITANT CORONARY ARTERY BYPASS GRAFTING. THIS 28F SIZE WAS CHOSEN BASED ON THE PATIENT BODY BUILD AND ANATOMICAL CONSIDERATION WITHOUT ANY CONFUSION REGARDING THE RELATIONSHIP BETWEEN SIDE AND CORRESPONDING FLOW RATE. NO VISUAL ABNORMALITIES WERE OBSERVED IN THE PACKAGING PRIOR TO OPENING AND IN THE DEVICE ITSELF AFTER OPENING. DURING USE, THERE WAS NO KINKING OF THE CANNULA, AND NO EXTERNAL FORCE WAS APPLIED TO THE CANNULA. THE SURGEON AFFIRMED THAT THERE WAS NO MALFUNCTION OF THE DEVICE AND THE DEVICE DID NOT CAUSE OR CONTRIBUTE TO THE EVENT. HE COMMENTED THAT THE SMALL SIZE OF THE CANNULA WAS THE CAUSE OF THE INADEQUATE VENOUS DRAINAGE. THE DEVICE WAS NOT RETURNED FOR EVALUATION AS IT WAS DISCARDED AT THE HOSPITAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23007665 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K120072. One submission often covers many models, so check the product and lot details in each record.

  • Z-3029-2018 · Mar 7, 2018 · Terminated
    Edwards Lifesciences Thin-Flex Single Stage Venous Drainage Cannula, 38FRX40CM, resterilized.
  • Z-3031-2018 · Mar 7, 2018 · Terminated
    Edwards Lifesciences Avid Dual Stage Venous Drainage Cannula, 29/37F (9.6/12.3MM) X 37CM (14.5"), resterilized.
  • Z-3030-2018 · Mar 7, 2018 · Terminated
    Edwards Lifesciences Thin-Flex Single Stage Venous Drainage Cannula, 34F (11.3MM) X 40CM (16"), resterilized.
  • Z-0398-2016 · Nov 4, 2015 · Terminated
    Thin-Flex Venous Cannula, Model Number:TF292902A

Other CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices

All CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS devices →

Sources and updates