SWAN-GANZ VIP+ 834F75Recalls & reported problems

Edwards Lifesciences LLC · SWAN-GANZ VIP+ TRI-LUMEN INFUSION THERMODILUTION CATHETER · Class 2

All models of SWAN-GANZ VIP (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
12
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00690103151756), received between Dec 12, 2024 and Sep 18, 2026 — classified as 12 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (10 reports).

1 FDA recall record name this device's identifier, 1 still open.

4 other recalls cover devices cleared under the same FDA submission (K160084); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 10 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 8, 2026Class IIOpen, Classified

    Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.

    Z-2202-2026 · Root cause: Process change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings10
  • Detachment of Device or Device Component1
  • Failure to Deflate1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820324 ↗
  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    It was reported that during use of the Swan-Ganz catheter it gave inaccurate values. Displayed values and expected values are unknown. There were microcracks in the blue lumen at the connection with the proximal injectate lumen. There was no allegation of patient injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820322 ↗
  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    It was reported that during use of the Swan-Ganz catheter it gave inaccurate values. Displayed values and expected values are unknown. There were microcracks in the blue lumen at the connection with the proximal injectate lumen. There was no allegation of patient injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820318 ↗
  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    It was reported that during use of the Swan-Ganz catheter it gave inaccurate values. Displayed values and expected values are unknown. There were microcracks in the blue lumen at the connection with the proximal injectate lumen. There was no allegation of patient injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820304 ↗
  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820288 ↗
  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    It was reported that during use of the Swan-Ganz catheter it gave inaccurate values. Displayed values and expected values are unknown. There were microcracks in the blue lumen at the connection with the proximal injectate lumen. There was no allegation of patient injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820282 ↗
  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    It was reported that during use of the Swan-Ganz catheter it gave inaccurate values. Displayed values and expected values are unknown. There were microcracks in the blue lumen at the connection with the proximal injectate lumen. There was no allegation of patient injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820271 ↗
  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    It was reported that during use of the Swan Ganz catheter, it provided high CVP values. The catheter was placed in the OR shortly before the inaccurate values were seen in the unit. After re-zeroing and troubleshooting the CVP transducer, the CVP line was moved to the double lumen Central Venous Catheter (CVC) and the reading there was an expected value. There were microcracks in the blue lumen at the connection with the proximal injectate lumen. The RN was still able to check thermodilution utilizing the blue PA line but monitored CVP on double lumen CVC instead. There was no allegation of patient injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820264 ↗
  • MalfunctionSep 18, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50820263 ↗
  • MalfunctionSep 11, 2026Detachment of Device or Device Component

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 50743210 ↗
  • MalfunctionAug 13, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT, DURING USE, A SWAN GANZ CATHETER HAD INACCURATE READINGS. THE CATHETER WAS REMOVED THE NEXT DAY. NO ALLEGATION OF PATIENT INJURIES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22801273 ↗
  • MalfunctionDec 12, 2024Failure to Deflate; Failure to Deflate

    AS REPORTED, ONCE THIS SWAN-GANZ CATHETER WAS INSERTED INTO PATIENT, IT COULD NOT BE DEFLATED. BEFORE INSERTION, THE CATHETER WAS FLUSHED AND THE BALLOON TESTED SUCCESSFULLY. THE BALLOON INFLATED EASILY AND DEFLATED PASSIVELY. UPON INSERTION TO PATIENT, THE BALLOON INFLATED SMOOTHLY. BEFORE FLOATING, THE PA (PULMONARY ARTERY) PRESSURE WAS REGISTERING A HIGH READING ALREADY. THE CATHETER WAS FLOATED BUT THE USER THOUGHT IT WAS COILING IN THE RIGHT ATRIUM, SO IT WAS DECIDED TO REMOVE THE CATHETER. HOWEVER, THE BALLOON DID NOT DEFLATE PASSIVELY. THE BALLOON WAS TRIED TO BE DEFLATED MANUALLY AS PER CONSULTANT ANESTHETIST'S INSTRUCTIONS. THE TOE PROBE (TRANS-OESOPHAGEAL ECHOCARDIOGRAM) WAS UTILIZED TO MAKE SURE THE BALLOON WAS NOT INFLATED BEFORE WITHDRAWING, BUT IT WAS FINALLY REMOVED WITH THE BALLOON INFLATED WITHOUT ANY ISSUES. ONCE REMOVED, THE BALLOON APPEARED TO BE INFLATED AND HALF-FILLED WITH A CLEAR FLUID OF UNKNOWN ORIGIN. THE PROBLEM WAS SOLVED USING A NEW SWAN-GANZ CATHETER. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20912113 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K160084. One submission often covers many models, so check the product and lot details in each record.

  • Z-0765-2019 · Dec 21, 2018 · Terminated
    Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 151F7 Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly dete
  • Z-0764-2019 · Dec 21, 2018 · Terminated
    Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131VF7P Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly de
  • Z-0762-2019 · Dec 21, 2018 · Terminated
    Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF131F 7J Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly det
  • Z-0763-2019 · Dec 21, 2018 · Terminated
    Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131F7P (5 pack of 131F7) Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians