Swan-Ganz Catheter, Models: 131F7/131F7P, 131F7J, 141F7… recallZ-2202-2026 · Becton, Dickinson and Company · initiated Apr 8, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Becton, Dickinson and Company or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.
FDA root cause category: Process change control
What the company is doing
On 4/8/2026, mailed correction notices were mailed to customers and distributors who were informed of the following: Consider the following to reduce the potential for hub breakage issues. Minimize manipulation of proximal injectate hub and lumen. Pay frequent attention to the (blue) proximal injectate lumen for leakage or breakage, and exchange with another catheter if the issue is encountered. Modify existing clinical workflows to minimize flexing and stress of (blue) proximal injectate lumen. For example: o Securing the catheter lumens to an IV arm board to mitigate bending/flexing of the lumen(s) o Consider using the stopcock on white hub to allow for intermittent transducer pressure tubing changes without twisting white hub on Swan Ganz (blue) proximal injectate lumen. o Prevent CO-Set syringe and proximal injectate lumen from dangling/hanging without support. o Change to automated cardiac output models to mitigate device stress. Providers and institutions may choose to utilize transthoracic or transesophageal echocardiography in combination with techniques to monitor central venous pressure via a central venous catheter in patients in need of right ventricle monitoring and management. Converting to Acumen IQ / FloTrac sensor can provide continuous minimally invasive hemodynamic monitoring from the arterial line. Consider using ClearSight cuff for continuous non-invasive hemodynamic monitoring. Distributors should forward the correction notice to customers who have purchased the impacted product. Complete and return the acknowledgement form via email to [email protected] Questions can be directed to the firm's Customer Service at (800) 424-3278
Recall details
- Recall number
- Z-2202-2026
- Status
- Open, Classified
- Initiated by the company
- Apr 8, 2026
- Posted by the FDA
- May 15, 2026
- Recalling firm
- Becton, Dickinson and Company
- Quantity
- 589,223
- Product code
- DYB
- FDA submissions
- K231248, K233824
- Distribution
- Worldwide - US Nationwide distribution including in the states of TX, SC, CT, GA, VA, MO, NM, UT, NJ, CA, FL, PA, IN, TN, WA, KY, MD, WV, MI, MA, NC, MN, NY, AZ, AL, LA, CO, OK, HI, MS, IL, OH, NE, DC, KS, NV, WI, SD, IA, AR, ID, OR, WY, MT, RI, VT, DE, AK, NH, ND, ME, PR and the countries of Japan, Brazil, Colombia, Australia, New Zealand, Poland, Czech Republic, South Africa, Mauritius, Germany, Israel, Iran, Slovakia, Switzerland, Slovenia, Bulgaria, Cyprus, Saudi Arabia, United Arab Emirates (UAE), Kuwait, Malta, Bahrain, Qatar, Hungary, Morocco, Romania, Lebanon, Tunisia, Oman, Turkey, Austria, Belgium, Luxembourg, France, Netherlands, Italy, Spain, Canary Islands, Portugal, United Kingdom, Northern Ireland, Ireland, Denmark, Sweden, Iceland, Norway, Estonia, Finland, Greece, Chile, Puerto Rico, Canada.
Product description
Swan-Ganz Catheter, Models: 131F7/131F7P, 131F7J, 141F7, 151F7, 834F75, 096F6/096F6P, C144F7, 782F75M, 132F5, C146F7, 831F75/831F75P, TS105F5, 774F75, 777F8, AIQSGF8;
Code information (models, lots, serial numbers)
Show the full code information (14k characters)
Model/UDI-DI/Lot(Expiration): 131F7P/00690103000412/65902900(8/1/2026), 65902901(8/3/2026), 65910728(8/2/2026), 65910729(7/29/2026),65910730(7/27/2026), 65910731(7/25/2026), 65910732(7/26/2026), 65910735(7/30/2026), 65910736(8/24/2026), 65910737(7/31/2026), 65916927(8/3/2026), 65916928(8/6/2026), 65916929(8/7/2026), 65916930(8/8/2026), 65916931(8/9/2026), 65916932(8/22/2026), 65916933(8/23/2026), 65944791(8/20/2026),65944792(8/21/2026), 65955857(9/9/2026), 65955858(9/17/2026), 65955859(9/26/2026), 65955860(9/27/2026), 65973739(9/5/2026), 65973740(9/6/2026), 65973741(9/7/2026), 65973742(9/10/2026), 65988571(9/28/2026), 65997478(9/16/2026), 65997479(9/17/2026), 65997481(9/19/2026), 65997482(9/20/2026), 66010504(9/10/2026), 66010505(9/11/2026), 66010506(9/12/2026), 66010507(9/13/2026), 66010508(9/14/2026), 66023074(9/30/2026), 66033088(10/5/2026), 66048930(10/9/2026), 66057629(10/6/2026), 66057630(10/8/2026), 66057944(10/15/2026), 66057945(10/17/2026), 66061552(10/24/2026), 66074999(10/4/2026), 66075318(11/19/2026), 66104892(10/21/2026), 66104893(10/22/2026), 66104894(10/23/2026), 66104895(11/13/2026), 66108277(11/15/2026), 66118638(11/22/2026), 66120310(10/29/2026), 66120312(10/30/2026), 66120313(10/31/2026), 66127027(11/15/2026), 66127028(11/20/2026), 66132570(11/21/2026), 66132571(11/25/2026), 66132572(11/27/2026), 66140447(11/2/2026), 66140448(11/5/2026), 66140449(11/6/2026), 66140450(11/7/2026), 66142269(11/28/2026), 66142274(11/29/2026), 66151008(12/20/2026), 66160820(11/30/2026), 66160821(12/3/2026), 66160822(12/4/2026), 66160823(12/5/2026), 66174935(12/6/2026), 66186389(12/9/2026), 66186390(12/10/2026), 66186391(12/11/2026), 66186393(12/19/2026), 66214346(12/12/2026), 66214347(12/17/2026), 66214348(12/13/2026), 66214349(12/16/2026), 66225263(1/24/2027), 66230833(1/27/2027), 66230834(1/28/2027), 66231174(1/29/2027), 66236300(2/3/2027), 66243301(1/30/2027), 66243302(2/1/2027), 66254697(2/4/2027), 66254698(2/10/2027), 66273573(2/11/2027), 66273574(2/12/2027), 66283862(2/8/2027), 66296222(3/5/2027), 66315689(3/6/2027), 66351811(3/8/2027), 66351812(3/11/2027), 66351813(3/12/2027), 66378395(3/13/2027), 66378396(3/18/2027), 66384489(3/18/2027), 66414674(3/22/2027), 66438179(4/3/2027), 66438197(4/12/2027), 66438198(5/24/2027), 66438199(5/27/2027), 66438200(4/10/2027), 66439368(3/26/2027), 66439369(3/31/2027), 66439370(4/2/2027), 66439371(4/7/2027), 66439372(4/9/2027), 66488597(5/28/2027), 66521743(4/16/2027), 66529338(4/15/2027), 66529339(4/21/2027), 66529340(4/22/2027), 66529341(4/23/2027), 66529345(5/23/2027), 66544240(4/24/2027), 66594945(5/14/2027), 66597237(4/29/2027), 66597238(4/30/2027), 66597239(5/10/2027), 66601166(5/2/2027), 66601167(5/8/2027), 66601168(5/12/2027), 66601169(5/13/2027), 66601170(5/14/2027), 66601171(5/29/2027), 66642726(5/9/2027), 66692566(6/6/2027), 66704724(6/4/2027), 66710839(6/2/2027), 66710840(6/3/2027), 66739812(6/13/2027), 66739813(6/18/2027), 66739814(6/17/2027), 66739910(6/24/2027), 66739911(7/17/2027), 66739912(7/9/2027), 66739913(6/16/2027), 66739916(6/23/2027), 66740114(7/17/2027), 66758828(7/10/2027), 66767282(6/28/2027), 66769651(7/18/2027), 66782067(7/21/2027), 66826029(7/23/2027), 66826030(7/24/2027), 66832079(7/25/2027), 66844048(7/31/2027), 66873884(8/2/2027), 66873885(8/5/2027), 66873886(8/18/2027), 66895223(8/20/2027), 66938771(8/22/2027), 66938772(8/26/2027), 66938773(8/27/2027), 66964710(9/1/2027), 66964711(9/4/2027), 66982688(9/26/2027), 67025469(10/20/2027), 67047800(10/17/2027), 67047801(10/21/2027), 67068446(10/28/2027), 67092101(10/28/2027), 67098252(10/30/2027), 67098253(11/3/2027), 67199591(11/12/2027), 67199595(11/13/2027), 67199596(11/11/2027), 67199617(11/21/2027), 67199618(11/22/2027), 67217934(1/7/2028), 67217935(12/4/2027), 67266297(1/12/2028), 67275624(12/20/2027), 67275626(12/3/2027), 67322713(1/13/2028), 67322714(1/16/2028), 67357488(1/29/2028), 67395549(2/5/2028), 67395555(1/28/2028), 66739917(6/12/2027); 131F7/00690103000405/ 65902900(8/1/2026), 65902901(8/3/2026), 65910728(8/2/2026), 65910729(7/29/2026), 65910730(7/27/2026), 65910731(7/25/2026), 65910732(7/26/2026), 65910735(7/30/2026), 65910736(8/24/2026), 65910737(7/31/2026), 65916927(8/3/2026), 65916928(8/6/2026), 65916929(8/7/2026), 65916930(8/8/2026), 65916931(8/9/2026), 65916932(8/22/2026), 65916933(8/23/2026), 65944791(8/20/2026), 65944792(8/21/2026), 65955857(9/9/2026), 65955858(9/17/2026), 65955859(9/26/2026), 65955860(9/27/2026), 65973739(9/5/2026), 65973740(9/6/2026), 65973741(9/7/2026), 65973742(9/10/2026), 65988571(9/28/2026), 65997478(9/16/2026), 65997479(9/17/2026), 65997481(9/19/2026), 65997482(9/20/2026), 66010504(9/10/2026), 66010505(9/11/2026), 66010506(9/12/2026) ,66010507(9/13/2026) ,66010508(9/14/2026) ,66023074(9/30/2026) ,66033088(10/5/2026) ,66048930(10/9/2026), 66057629(10/6/2026), 66057630(10/8/2026), 66057944(10/15/2026), 66057945(10/17/2026), 66061552(10/24/2026), 66074999(10/4/2026), 66075318(11/19/2026), 66104892(10/21/2026), 66104893(10/22/2026), 66104894(10/23/2026), 66104895(11/13/2026), 66108277(11/15/2026), 66118638(11/22/2026), 66120310(10/29/2026), 66120312(10/30/2026), 66120313(10/31/2026), 66127027(11/15/2026), 66127028(11/20/2026),66132570(11/21/2026), 66132571(11/25/2026), 66132572(11/27/2026), 66140447(11/2/2026), 66140448(11/5/2026), 66140449(11/6/2026),66140450(11/7/2026), 66142269(11/28/2026), 66142274(11/29/2026), 66151008(12/20/2026), 66160820(11/30/2026), 66160821(12/3/2026), 66160822(12/4/2026), 66160823(12/5/2026), 66174935(12/6/2026), 66186389(12/9/2026), 66186390(12/10/2026),66186391(12/11/2026), 66186393(12/19/2026), 66214346(12/12/2026), 66214347(12/17/2026), 66214348(12/13/2026),66214349(12/16/2026), 66225263(1/24/2027), 66230833(1/27/2027), 66230834(1/28/2027), 66231174(1/29/2027), 66236300(2/3/2027), 66243301(1/30/2027), 66243302(2/1/2027), 66254697(2/4/2027), 66254698(2/10/2027), 66273573(2/11/2027),66273574(2/12/2027), 66283862(2/8/2027),66296222(3/5/2027), 66315689(3/6/2027), 66351811(3/8/2027), 66351812(3/11/2027), 66351813(3/12/2027), 66378395(3/13/2027), 66378396(3/18/2027), 66384489(3/18/2027), 66414674(3/22/2027),66438179(4/3/2027), 66438197(4/12/2027), 66438198(5/24/2027),66438199(5/27/2027), 66438200(4/10/2027), 66439368(3/26/2027), 66439369(3/31/2027), 66439370(4/2/2027), 66439371(4/7/2027), 66439372(4/9/2027), 66488597(5/28/2027),66521743(4/16/2027), 66529338(4/15/2027), 66529339(4/21/2027), 66529340(4/22/2027), 66529341(4/23/2027),66529345(5/23/2027),66544240(4/24/2027), 66594945(5/14/2027), 66597237(4/29/2027),66597238(4/30/2027), 66597239(5/10/2027), 66601166(5/2/2027), 66601167(5/8/2027), 66601168(5/12/2027), 66601169(5/13/2027), 66601170(5/14/2027), 66601171(5/29/2027), 66642726(5/9/2027), 66692566(6/6/2027), 66704724(6/4/2027), 66710839(6/2/2027), 66710840(6/3/2027), 66739812(6/13/2027),66739813(6/18/2027), 66739814(6/17/2027), 66739910(6/24/2027), 66739911(7/17/2027), 66739912(7/9/2027), 66739913(6/16/2027), 66739916(6/23/2027), 66740114(7/17/2027), 66758828(7/10/2027), 66767282(6/28/2027), 66769651(7/18/2027), 66782067(7/21/2027), 66826029(7/23/2027),66826030(7/24/2027), 66832079(7/25/2027), 66844048(7/31/2027), 66873884(8/2/2027), 66873885(8/5/2027), 66873886(8/18/2027), 66895223(8/20/2027), 66938771(8/22/2027), 66938772(8/26/2027), 66938773(8/27/2027), 66964710(9/1/2027), 66964711(9/4/2027), 66982688(9/26/2027), 67025469(10/20/2027), 67047800(10/17/2027), 67047801(10/21/2027), 67068446(10/28/2027), 67092101(10/28/2027), 67098252(10/30/2027), 67098253(11/3/2027), 67199591(11/12/2027), 67199595(11/13/2027), 67199596(11/11/2027), 67199617(11/21/2027), 67199618(11/22/2027), 67204649(12/4/2027), 67217934(1/7/2028), 67217935(12/4/2027), 67266297(1/12/2028), 67268374(12/3/2027), 67275624(12/20/2027), 67275626(12/3/2027), 67322713(1/13/2028), 67322714(1/16/2028), 67357488(1/29/2028), 67395549(2/5/2028), 67395555(1/28/2028); 131F7J/67268374(12/3/2027); 141F7/00690103000504/ 65891632(8/5/2026), 65891637(8/20/2026), 65916934(8/19/2026), 65916935(8/26/2026), 65944795(8/27/2026), 65955863(8/27/2026), 65973744(9/9/2026), 65988572(9/18/2026), 65997483(10/1/2026), 66023075(10/4/2026), 66023076(10/28/2026), 66033077(10/25/2026), 66061553(10/26/2026), 66273575(2/13/2027), 66315690(3/8/2027), 66378397(3/20/2027), 66414675(3/28/2027), 66414681(5/30/2027), 66465246(4/23/2027), 66552320(4/29/2027), 66594946(4/28/2027), 66623671(5/1/2027), 66644932(5/16/2027), 66739789(6/26/2027), 66767253(6/30/2027), 66809856(7/4/2027), 66809857(7/7/2027), 66812062(7/8/2027), 66826050(7/30/2027), 66873887(7/29/2027), 66945954(8/26/2027), 67014463(12/19/2026), 67056072(9/22/2027); 151F7/00690103000566/ 66001841(8/29/2026), 66003074(9/14/2026), 66003075(9/7/2026), 66017563(9/21/2026), 66081903(10/3/2026), 66081904(11/1/2026), 66106780(10/25/2026), 66108265(11/7/2026), 66174936(12/17/2026), 66351108(3/6/2027), 66351109(3/12/2027), 66351111(3/21/2027), 66384522(3/29/2027), 66439364(4/4/2027), 66471244(4/11/2027), 66544241(4/24/2027), 66552321(5/2/2027), 66594947(5/9/2027), 66615747(5/14/2027), 66616064(6/27/2027), 66643585(5/21/2027), 66670993(8/7/2027), 66710846(6/6/2027), 66758827(8/1/2027), 66809858(8/6/2027), 66809859(9/11/2027), 66809860(9/20/2027), 66889370(8/26/2027), 66972526(10/3/2027), 67197204(11/20/2027), 67197205(11/28/2027); 834F75/00690103151756/ 66149781(11/20/2026), 66169452(11/21/2026), 66169453(11/22/2026), 66169454(11/22/2026), 66225269(1/17/2027), 66292951(2/6/2027), 66295865(2/25/2027), 66325010(2/21/2027), 66358925(2/25/2027), 66360368(2/24/2027), 66384531(3/21/2027), 66384532(3/21/2027), 66384533(3/31/2027), 66419531(3/15/2027), 66597740(4/25/2027), 66704730(6/6/2027), 66710835(10/16/2027), 66739754(7/2/2027), 66748559(6/11/2027), 66748561(6/10/2027), 66829581(7/5/2027), 66832094(8/1/2027), 66894451(7/31/2027), 66894452(8/1/2027), 66987645(8/30/2027), 67025475(10/13/2027), 67045003(9/17/2027), 67047807(10/14/2027), 67075111(9/27/2027), 67197192(11/11/2027), 67205219(11/18/2027); 096F6, 096F6P/00690103000245, 00690103218602/66687312(6/9/2027), 66687922(6/10/2027), 66687924(6/13/2027), 66780053(6/26/2027), 66822159(9/22/2027); C144F7/00690103005134/66822158(7/15/2027); 782F75M/00690103197303/66725374(6/19/2027); 132F5/00690103000467/66687936(6/9/2027), 66467271(6/9/2027); C146F7/00690103153132; 831F75P/831F75-00690103003031, 00690103218282/66175089(11/26/2026), 66175090(12/2/2026), 66175091(12/2/2026), 66175092(12/3/2026), 66175093(11/26/2026), 66175094(11/27/2026), 66175095(11/28/2026), 66175096(11/29/2026), 66175097(11/29/2026), 66175103(12/4/2026), 66175104(12/5/2026), 66175105(12/5/2026), 66175120(12/9/2026), 66175121(12/10/2026), 66175122(12/11/2026), 66208306(12/12/2026), 66208323(12/13/2026), 66211381(12/16/2026), 66211385(1/8/2027), 66211386(1/9/2027), 66225264(1/13/2027), 66225265(1/14/2027), 66225266(1/15/2027), 66225267(1/16/2027), 66225268(1/22/2027), 66268847(1/23/2027), 66278675(2/5/2027), 66283844(2/14/2027), 66283845(2/26/2027), 66283846(2/20/2027), 66296223(3/3/2027), 66298719(3/1/2027), 66301299(2/17/2027), 66301300(2/19/2027), 66439365(4/3/2027), 66439366(4/5/2027), 66439367(4/11/2027), 66488586(3/28/2027), 66529347(5/31/2027), 66554378(4/12/2027), 66591113(5/2/2027), 66660610(5/20/2027), 66664653(5/21/2027), 66667556(5/17/2027), 66667557(5/19/2027), 66704725(6/5/2027), 66704758(6/27/2027), 66739799(7/15/2027), 66758829(7/11/2027), 66759128(6/12/2027), 66767284(6/25/2027), 66782068(7/28/2027), 66804400(6/30/2027), 66804401(7/2/2027), 66804402(7/3/2027), 66826032(7/26/2027), 66873891(8/8/2027), 66934651(8/23/2027), 66934677(8/29/2027), 66964714(8/25/2027), 66972587(9/18/2027), 66972588(9/22/2027), 67003703(9/30/2027), 67003704(10/1/2027), 67005151(9/9/2027), 67005156(9/3/2027), 67005540(10/7/2027), 67005541(10/9/2027), 67005542(9/10/2027), 67005543(9/11/2027), 67005544(9/23/2027), 67005545(10/10/2027), 67184885(11/1/2027); TS105F5/00690103018295; 774F75/00690103146998/66521746(5/7/2027), 66844440(7/8/2027), 66844441(7/9/2027), 66844442(7/10/2027), 66844443(7/11/2027), 66844444(7/11/2027), 66844445(7/12/2027), 66844446(7/14/2027), 66844447(7/15/2027), 66844448(7/15/2027), 66844449(7/16/2027), 66844450(7/16/2027), 66844451(7/17/2027), 66844452(7/18/2027), 66844453(8/4/2027), 66919298(8/6/2027), 66919299(8/20/2027), 66920022(8/5/2027); 777F8/00690103146554/66417428(4/21/2027), 66417429(4/22/2027), 66417430(4/22/2027), 66419577(4/11/2027), 66419578(4/12/2027), 66419579(4/14/2027), 66419580(4/14/2027), 66419581(4/15/2027), 66419582(4/15/2027), 66419583(4/16/2027), 66419584(4/16/2027), 66419585(4/21/2027), 66438181(4/23/2027), 66465249(4/25/2027), 66523321(4/23/2027), 66523328(4/24/2027), 66523330(4/28/2027), 66523331(4/25/2027), 66544242(4/30/2027), 66568201(5/7/2027), 66590392(5/12/2027), 66600645(4/28/2027), 66600646(4/29/2027), 66600647(4/29/2027), 66600648(4/30/2027), 66600649(4/30/2027), 66615750(5/1/2027), 66615751(5/3/2027), 66615752(5/19/2027), 66615753(5/20/2027), 66619148(5/2/2027), 66643587(5/22/2027), 66643588(5/23/2027), 66643589(5/26/2027), 66643590(5/27/2027), 66643591(5/27/2027), 66643592(5/28/2027), 66643593(5/29/2027), 66643594(5/30/2027), 66643595(6/2/2027), 66643596(6/3/2027), 66646363(5/12/2027), 66646364(5/13/2027), 66646365(5/14/2027), 66646366(5/15/2027), 66646367(5/16/2027), 66646368(5/16/2027), 66660611(6/3/2027), 66660612(6/4/2027), 66660613(6/4/2027), 66660614(6/5/2027), 66660615(6/5/2027), 66660616(6/7/2027), 66687707(6/7/2027), 66687708(6/9/2027), 66687805(6/9/2027), 66694839(7/5/2027), 66704726(6/10/2027), 66704727(6/11/2027), 66704728(6/11/2027), 66704729(6/12/2027), 66764452(6/12/2027), 66764453(6/13/2027), 66764454(6/16/2027), 66764455(6/16/2027), 66764456(6/17/2027), 66764457(6/17/2027), 66769776(6/18/2027), 66769777(6/23/2027), 66782069(7/7/2027), 66800668(6/23/2027), 66804171(6/28/2027), 66928758(8/7/2027), 67142955(10/30/2027), 67142972(11/14/2027), 67143094(11/27/2027); AIQSGF8/00690103217094/66414809(3/16/2027), 66487805(5/11/2027), 66417072(3/22/2027), 66493141(5/18/2027), 66493138(5/12/2027), 66417073(3/23/2027), 66437580(3/28/2027);
Devices named in this recall
17 devices in the FDA device registry whose identifier appears in the recall's code information.
- SWAN-GANZ CCOMBO V 777F8Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED2 recalls92 reports
- SWAN-GANZ CCOMBO V 774F75Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED2 recalls2 deaths41 reports
- SWAN-GANZ VIP 831F75Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED1 recall23 reports
- SWAN-GANZ CONTROLCATH C146F7Edwards Lifesciences LLC · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall16 reports
- SWAN-GANZ 131F7Edwards Lifesciences LLC · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall13 reports
- SWAN-GANZ VIP+ 834F75Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED1 recall12 reports
- SWAN-GANZ TRUE SIZE 096F6Edwards Lifesciences LLC · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall9 reports
- SWAN-GANZ VIP 831F75Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED1 recall6 reports
- SWAN-GANZ VIP 782F75MEdwards Lifesciences LLC · Catheter, continuous flush1 recall4 reports
- SWAN-GANZ HI-SHORE TRUE SIZE 141F7Edwards Lifesciences LLC · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall3 reports
- SWAN-GANZ 131F7Edwards Lifesciences LLC · Catheter, Intravascular, Diagnostic1 recall2 reports
- SWAN-GANZ TRUE SIZE 151F7Edwards Lifesciences LLC · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall2 reports
- SWAN-GANZ TRUE SIZE TS105F5Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED1 recall2 reports
- SWAN-GANZ CONTROLCATH TRUE SIZE C144F7Edwards Lifesciences LLC · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall1 report
- Swan-Ganz AIQSGF8Edwards Lifesciences LLC · Catheter, Flow Directed1 recall1 report
- SWAN-GANZ TRUE SIZE 132F5Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED1 recall0 reports
- SWAN-GANZ TRUE SIZE 096F6Edwards Lifesciences LLC · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall0 reports
Product lines
- SWAN-GANZ TRUE SIZEEdwards Lifesciences LLC · 8 models · 26 reports · 1 recall
- SWAN-GANZ VIPEdwards Lifesciences LLC · 4 models · 45 reports · 1 recall
- SWAN-GANZEdwards Lifesciences LLC · 13 models · 108 reports · 3 recalls
- SWAN-GANZ CCOMBO VEdwards Lifesciences LLC · 2 models · 133 reports · 3 recalls
Other devices under the same FDA submission
4 devices cleared under K231248, K233824 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- SWAN GANZ SGPT54Edwards Lifesciences LLC · Catheter, Oximeter, Fiber-Optic1 recall0 reports
- SWAN GANZ SGPT64PEdwards Lifesciences LLC · Catheter, Oximeter, Fiber-Optic1 recall0 reports
- SWAN-GANZ SGPT755PEdwards Lifesciences LLC · Catheter, Oximeter, Fiber-Optic1 recall0 reports
- SWAN-GANZ CCOMBO V 777F8Becton, Dickinson And Company · Catheter, continuous flush1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 98603).
- Z-2203-2026 · Open, ClassifiedSwan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755P
- Z-2204-2026 · Open, ClassifiedSwan-Ganz Pacing Catheter, Models: D200F7;
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page