Skip to content

CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS - MITRAL 6900P (REF 6900P33MM)Recalls & reported problems

Edwards Lifesciences LLC · Class 3 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00690103152678), received between Feb 4, 2021 and May 26, 2023 — classified as 6 injuries.

The most frequently reported specific device problem is “Gradient Increase” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Gradient Increase4
  • Degraded2
  • Fluid/Blood Leak2
  • Insufficient Device Problem Information2
  • Calcified1
  • Difficult to Open or Close1
  • Patient Device Interaction Problem1

Patient problems

  • Heart Failure/Congestive Heart Failure3
  • Insufficient Information3
  • Dyspnea1
  • Endocarditis1
  • Swelling/ Edema1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 26, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    EDWARDS IMPLANT PATIENT REGISTRY RECEIVED INFORMATION A PATIENT WITH A 33MM 6900P MITRAL VALVE IMPLANTED IN THE TRICUSPID POSITION FOR FOUR (4) YEARS, EIGHT (8) MONTHS, UNDERWENT VALVE-IN-VALVE PROCEDURE DUE TO UNKNOWN REASONS. A 29MM 9600TFX TRANSCATHETER VALVE WAS IMPLANTED INSIDE THE 33MM VALVE. EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL CUSTOMER COMPLAINT. THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE.

    Patient problems: Insufficient Information

    MAUDE report 17014447 ↗
  • InjuryMay 5, 2023Degraded; Fluid/Blood Leak; Gradient Increase; Difficult to Open or Close; Degraded; Fluid/Blood Leak; Gradient Increase; Difficult to Open or Close

    IT WAS REPORTED THAT A PATIENT WITH A 33MM VALVE IN TRICUSPID POSITION IS BEING EVALUATED FOR A VALVE-IN-VALVE PROCEDURE AFTER AN IMPLANT DURATION OF 6 YEARS, 8 MONTHS DUE TO LEAFLET THICKENING, MOTION RESTRICTED, HIGH GRADIENT, AND MODERATE REGURGITATION.

    Patient problems: Insufficient Information

    MAUDE report 16882159 ↗
  • InjuryMar 21, 2023Gradient Increase; Gradient Increase

    IT WAS REPORTED THAT A PATIENT WITH A 33MM MITRAL VALVE IS BEING EVALUATED FOR A VALVE-IN-VALVE PROCEDURE AFTER AN IMPLANT DURATION OF 8 YEARS, 1 MONTHS DUE TO STENOSIS. THE PATIENT PRESENTED WITH HF, SOB, AND LOWER EXTREMITY EDEMA.

    Patient problems: Dyspnea; Heart Failure/Congestive Heart Failure; Swelling/ Edema

    MAUDE report 16584719 ↗
  • InjuryOct 1, 2021Insufficient Device Problem Information; Patient Device Interaction Problem; Insufficient Device Problem Information; Patient Device Interaction Problem

    THROUGH THE IMPLANT PATIENT REGISTRY, IT WAS REPORTED THAT A PATIENT WITH A 33MM 6900P MITRAL VALVE IMPLANTED IN THE TRICUSPID POSITION FOR 2 YEARS, 9 MONTHS WAS EXPLANTED DUE TO ENDOCARDITIS. THE DEVICE WAS REPLACED WITH A 31MM 7300TFX VALVE. PATIENT EXPIRED ON POD#4.PER MEDICAL RECORDS, PATIENT ALSO UNDERWENT REPAIR ASCENDING AORTIC PSEUDOANEURYSM WITH HEMASHIELD PATCH, REMOVAL OF WIRELESS PACEMAKER. THE PROSTHETIC VALVE WAS REMOVED AND REPLACED WITH A 31MM 7300TFX VALVE. POST OPERATIVE ECHO CONFIRMED GOOD VALVE FUNCTION. THE PATIENT WAS TRANSFERRED TO THE POSTOPERATIVE CARDIAC SURGICAL INTENSIVE CARE UNIT IN CRITICAL CONDITION. THE PATIENT DEVELOPED DIFFUSE CEREBRAL EDEMA POSTOPERATIVELY AND EXPIRED ON POD#4.

    Patient problems: Endocarditis; Insufficient Information

    MAUDE report 12566054 ↗
  • InjuryMay 10, 2021Gradient Increase; Gradient Increase

    IT WAS REPORTED THAT A PATIENT WITH A 33MM 6900P MITRAL VALVE UNDERWENT A VALVE-IN-VALVE PROCEDURE AFTER AN IMPLANT DURATION OF EIGHT (8) YEARS, TWO (2) MONTHS DUE TO SEVERE MITRAL VALVE STENOSIS. THE PATIENT PRESENTED NYHA CLASS IV. THE TMVR WAS PERFORMED WITH A 29MM 9600TFX VALVE. ON POD #2, THE PATIENT WAS DISCHARGED IN STABLE CONDITION.

    Patient problems: Heart Failure/Congestive Heart Failure

    MAUDE report 11798717 ↗
  • InjuryFeb 4, 2021Calcified; Degraded; Fluid/Blood Leak; Gradient Increase; Calcified; Degraded; Fluid/Blood Leak; Gradient Increase

    THROUGH CLINICAL SAFETY, IT WAS REPORTED THAT A PATIENT WITH A 33MM 6900P MITRAL VALVE UNDERWENT A VALVE-IN-VALVE PROCEDURE AFTER AN IMPLANT DURATION OF 14 YEARS DUE TO SEVERE MITRAL REGURGITATION, MILD MITRAL STENOSIS, THICKENING, AND TORN LEAFLET SECONDARY TO CALCIFICATION. THE TMVR WAS PERFORMED WITH A 29MM 9600TFX TRANSCATHETER VALVE. THERE WERE NO COMPLICATIONS NOTED. THE PATIENT WAS TRANSFERRED TO THE ICU IN STABLE CONDITION POST PROCEDURE. THE PATIENT WAS DISCHARGED ON POD #4 IN STABLE CONDITION.

    Patient problems: Heart Failure/Congestive Heart Failure

    MAUDE report 11283004 ↗

FDA clearance and classification

Approved through premarket approval (PMA) P860057 on Sep 26, 1991. Submitted by Edwards Lifesciences, LLC as “Edwards Lifesciences Perimount Aortic And Mitral Bioprostheses”. The FDA review took 1,773 days. The FDA has since approved 211 supplements (changes to the device or its labeling), most recently on Oct 28, 2026. FDA record ↗

The FDA regulates product code DYE (replacement heart-valve) under 21 CFR 870.3925 as a Class III device. 21 CFR 870.3925 ↗

Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS - MITRAL 6900P

Has the CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS - MITRAL 6900P been recalled?

No FDA recall record in our data names this exact device identifier (DI 00690103152678), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS - MITRAL 6900P?

The FDA's MAUDE database lists 6 adverse event reports for this device, received between Feb 4, 2021 and May 26, 2023. In those reports, the most common specific device problems are “Gradient Increase” (4), “Degraded” (2) and “Fluid/Blood Leak” (2); the most common patient problems are “Heart Failure/Congestive Heart Failure” (3), “Dyspnea” (1) and “Endocarditis” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS - MITRAL 6900P?

None of the reports are classified as deaths, 6 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS - MITRAL 6900P FDA approved?

It was approved through premarket approval (PMA) P860057 on Sep 26, 1991.

Is the CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS - MITRAL 6900P MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS - MITRAL 6900P, and is it still sold?

The FDA's device database lists Edwards Lifesciences LLC as the company responsible for it (brand CARPENTIER-EDWARDS PERIMOUNT PLUS PERICARDIAL BIOPROSTHESIS - MITRAL, model 6900P). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other replacement Heart-valve devices

All replacement Heart-valve devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page