Skip to content

CARPENTIER-EDWARDS PERIMOUNT MAGNA PERICARDIAL BIOPROSTHESIS - AORTIC 3000TFX (REF 3000TFX19MM)Recalls & reported problems

Edwards Lifesciences LLC · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00690103163636), received between Apr 12, 2022 and Aug 8, 2024 — classified as 5 injuries.

The most frequently reported specific device problem is “Gradient Increase” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gradient Increase4
  • Fluid/Blood Leak3
  • Difficult to Open or Close1
  • Inadequacy of Device Shape and/or Size1
  • Insufficient Device Problem Information1

Patient problems

  • Insufficient Information4
  • Dyspnea2
  • Heart Failure/Congestive Heart Failure2
  • Cardiogenic Shock1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 8, 2024Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED AND LEARNED THROUGH INVESTIGATION THAT A PATIENT WITH A 19MM 3000TFX AORTIC VALVE IN AORTIC POSITION UNDERWENT A VALVE REPLACEMENT PROCEDURE AFTER AN IMPLANT DURATION OF SIX (6) YEARS, 10 MONTHS DUE TO REGURGITATION. THE PATIENT PRESENTED WITH HEART FAILURE AND SHORTNESS OF BREATH. AVR WAS COMPLETED WITH A 21MM 11500A AORTIC VALVE.

    Patient problems: Dyspnea; Heart Failure/Congestive Heart Failure; Insufficient Information

    MAUDE report 19937295 ↗
  • InjuryJul 26, 2022Fluid/Blood Leak; Gradient Increase; Fluid/Blood Leak; Gradient Increase

    IT WAS REPORTED THAT A PATIENT WITH A 19MM AORTIC VALVE IS BEING EVALUATED FOR A VALVE-IN-VALVE PROCEDURE AFTER AN IMPLANT DURATION OF 4 YEARS, 7 MONTHS DUE TO SEVERE STENOSIS AND MILD REGURGITATION. THE PATIENT PRESENTED WITH SEVERE SOB AND ACUTE DIASTOLIC CHF. THE PATIENT HAS A HISTORY OF AS, CHF, PULMONARY HTN, HTN, HLD, DM, BREAST CARCINOMA, AND FORMER SMOKER.

    Patient problems: Heart Failure/Congestive Heart Failure

    MAUDE report 15103662 ↗
  • InjuryJul 12, 2022Gradient Increase; Gradient Increase

    IT WAS REPORTED THAT THIS PATIENT WITH A 3000TFX 19MM VALVE, IMPLANTED FOR FOUR (4) YEARS AND 10 MONTHS, IS BEING EVALUATED FOR A VALVE-IN-VALVE PROCEDURE DUE TO AORTIC STENOSIS.

    Patient problems: Insufficient Information

    MAUDE report 14988102 ↗
  • InjuryMay 21, 2022Fluid/Blood Leak; Gradient Increase; Difficult to Open or Close; Fluid/Blood Leak; Gradient Increase; Difficult to Open or Close

    IT WAS REPORTED THAT A PATIENT WITH A 19MM 3000TFX AORTIC VALVE UNDERWENT A VALVE-IN-VALVE PROCEDURE AFTER AN IMPLANT DURATION OF 4 YEARS, 3 MONTHS DUE TO AORTIC STENOSIS. THE TAVR WAS PERFORMED WITH A 23MM 9750TFX TRANSCATHETER VALVE SUCCESSFULLY. THE PATIENT DID WELL.

    Patient problems: Cardiogenic Shock; Insufficient Information

    MAUDE report 14466622 ↗
  • InjuryApr 12, 2022Gradient Increase; Inadequacy of Device Shape and/or Size; Insufficient Device Problem Information; Gradient Increase; Inadequacy of Device Shape and/or Size; Insufficient Device Problem Information

    IT WAS LEARNED THROUGH THE IMPLANT PATIENT REGISTRY THAT A 3000TFX 19MM BIOPROSTHETIC AORTIC VALVE, IMPLANTED FOR THREE (3) YEARS AND SEVEN (7) MONTHS, WAS EXPLANTED DUE TO UNKNOWN REASONS. THE EXPLANTED DEVICE WAS REPLACED WITH A 3300TFX 21MM BIOPROSTHETIC VALVE. PATIENT POST PROCEDURE IN RECOVERY. EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL CUSTOMER COMPLAINT. THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE.

    Patient problems: Dyspnea; Insufficient Information

    MAUDE report 14090961 ↗

Other replacement Heart-valve devices

All replacement Heart-valve devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page