CARPENTIER-EDWARDS PERIMOUNT RSRPERICARDIAL BIOPROSTHESIS - AORTIC 2800TFX (REF 2800TFX23MM)Recalls & reported problems
Edwards Lifesciences LLC · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00690103177244), received between Sep 20, 2024 and Jan 3, 2025 — classified as 2 injuries.
The most frequently reported specific device problem is “Degraded” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Degraded1
- Device Stenosis1
- Insufficient Device Problem Information1
Patient problems
- Chest Pain1
- Cough1
- Insufficient Information1
- Syncope/Fainting1
- Wheezing1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJan 3, 2025Insufficient Device Problem Information; Device Stenosis; Insufficient Device Problem Information; Device Stenosis
IT WAS LEARNED THROUGH IMPLANT PATIENT REGISTRY THAT A PATIENT WITH A 23MM 2800TFX AORTIC VALVE UNDERWENT A VALVE-IN-VALVE PROCEDURE AFTER AN IMPLANT DURATION OF 9 YEARS, 3 MONTHS DUE TO UNKNOWN REASON. THE PROCEDURE WAS PERFORMED WITH A 23MM 9755RSL TRANSCATHETER VALVE. THE PATIENT WAS IN RECOVERY POST PROCEDURE. EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL CUSTOMER COMPLAINT. THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE.
Patient problems: Chest Pain; Insufficient Information; Syncope/Fainting
MAUDE report 21074442 ↗InjurySep 20, 2024Degraded; Degraded
EDWARDS RECEIVED NOTIFICATION FROM PE RETROSPECTIVE STUDY (CHINA) THAT THIS 61-YEAR-OLD PATIENT WITH A 2800TFX23MM VALVE MODEL IMPLANTED IN AORTIC POSITION UNDERWENT REINTERVENTION FOR A VALVE-IN-VALVE PROCEDURE AFTER AN IMPLANT DURATION OF EIGHT (8) YEARS AND SIX (6) MONTHS DUE TO STRUCTURAL DETERIORATION WHICH LED TO SIGNIFICANT LEAFLET OPENING RESTRICTION, SEVERE STENOSIS AND MILD REGURGITATION. VALVE WAS PERFORMING WITH A MEAN GRADIENT OF 63MMHG AND A VELOCITY OF 5.21M/S. THESE FINDINGS WERE DETECTED ON AN ULTRASOUND PERFORMED EIGHTEEN (18) DAYS BEFORE THE REINTERVENTION. AS REPORTED, PATIENT PRESENTED MAINLY WITH COUGH AND WHEEZING. A 24MM NON-EDWARDS TRANSCATHETER VALVE WAS IMPLANTED WITHIN THE EDWARDS SURGICAL VALVE. PATIENT WAS DISCHARGED ON POSTOPERATIVE DAY TEN AND WAS NOTED TO BE CURRENTLY IN A STABLE CONDITION.
Patient problems: Cough; Wheezing
MAUDE report 20264859 ↗
Other replacement Heart-valve devices
- CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL BIOPROSTHESIS - AORTIC 3300TFXEdwards Lifesciences LLC · replacement Heart-valve4 deaths671 reports
- CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL BIOPROSTHESIS - AORTIC 3300TFXEdwards Lifesciences LLC · replacement Heart-valve1 death656 reports
- CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL BIOPROSTHESIS - AORTIC 3300TFXEdwards Lifesciences LLC · replacement Heart-valve5 deaths533 reports
- CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL EASE PERICARDIAL BIOPROSTHESIS 7300TFXEdwards Lifesciences LLC · replacement Heart-valve1 death425 reports
- CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL EASE PERICARDIAL BIOPROSTHESIS 7300TFXEdwards Lifesciences LLC · replacement Heart-valve1 death391 reports
- CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL EASE PERICARDIAL BIOPROSTHESIS 7300TFXEdwards Lifesciences LLC · replacement Heart-valve1 death290 reports
- CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL BIOPROSTHESIS - AORTIC 3300TFXEdwards Lifesciences LLC · replacement Heart-valve260 reports
- CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL EASE PERICARDIAL BIOPROSTHESIS 7300TFXEdwards Lifesciences LLC · replacement Heart-valve2 deaths229 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page