OptiSite Arterial Perfusion Cannula OPTI18Recalls & reported problems
Edwards Lifesciences LLC · OptiSite Arterial Perfusion Cannula · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00690103180565), received between Mar 4, 2022 and Nov 23, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (K073559); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- May 14, 2025Open, Classified
due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.
Z-2081-2025 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Break1
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 23, 2024Break; Break
PER THE REPORT RECEIVED FROM GERMANY AN OPTI18 CANNULA WAS FOUND TO HAVE THE WIRE OUTSIDE FROM THE CANNULA. AS REPORTED IT WAS NOTICED, WHEN THE CANNULA WAS ABOUT TO BE INSERTED INTO THE ARTERY. THERE WAS NO INJURY FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20766834 ↗InjuryMar 4, 2022Material Deformation; Material Deformation
IT WAS REPORTED THAT AN ARTERIAL CANNULA GOT KINKED DURING PROCEDURE. ANOTHER CANNULA WAS USED. THERE WAS NO PATIENT INJURY. AS REPORTED, NOTHING UNUSUAL HAPPENED DURING CANNULATION PROCESS. THE CANNULA WAS SECURED, WHICH PUTS IT AT SOMEWHAT OF A BEND. THE SURGICAL TEAM WENT ON BYPASS AND NOTICED HIGH LINE PRESSURES. CANNULA WAS CHECKED AND KINKING WAS OBSERVED WHERE THE MAIN CANNULA TUBING MEETS THE THICKER TUBING. THE SURGEON WAS NOT ABLE TO PINCH CANNULA TO RELIEVE THE KINKING. A NEW DEVICE HAD TO BE USED. IT WAS BELIEVED THAT FOR SOME REASON, THE CANNULA IS NOT AS ROBUST AT THAT POINT AS IT USED TO BE, AND IT IS KINKED WHEN IT IS SECURED IN A CURVED FASHION. THE SURGEON IS EXTREMELY SKILLED AND HAS BEEN DOING SURGERY FOR 25 YEARS. HIS TECHNIQUE IS IDENTICAL EVERY TIME. HE HAS USED THIS SAME CANNULA FOR AT LEAST 7 YEARS WITH NO ISSUES BEFORE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13668726 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K073559. One submission often covers many models, so check the product and lot details in each record.
- Z-2080-2025 · May 14, 2025 · Open, ClassifiedBrand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI16 Software Version: N/A Product Description: Cardiac cannula. The Edwards O
- Z-3034-2018 · Mar 7, 2018 · TerminatedEdwards Lifesciences Optisite Arterial Perfusion Cannula, 22F X 24CM, resterilized.
Other CANNULA, CATHETER devices
- OptiSite Arterial Perfusion Cannula OPTI16Edwards Lifesciences LLC · CANNULA, CATHETER1 recall3 reports
- AngioDynamics 06597004Angiodynamics, Inc. · CANNULA, CATHETER4 reports
- AngioDynamics 06597015Angiodynamics, Inc. · CANNULA, CATHETER2 reports
- AngioDynamics 06597053Angiodynamics, Inc. · CANNULA, CATHETER2 reports
- Arterial Cannula A212-65CSorin Group Italia Srl · CANNULA, CATHETER1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.