- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00690103193213), received between Oct 5, 2021 and Oct 5, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K140312, K160552); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 5, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE PUMP UNIT FAILED FUNCTIONAL TEST FOR PRESSURE AT VARIOUS FLOWS WITH THE DIFFERENCE PRESSURE OUTLET AD AT 150 NORMAL LITERS PER HOUR. THE MEASUREMENT IS 136.3 MMHG AND THE LIMITS ARE FROM PLUS OR MINUS 10 MMHG. THE TEST TECHNICIAN FOUND THAT THE PUMPS MANIFOLD BOTTOM HAD CRACKS AROUND THE PEMS FOR THE MANIFOLD LID. ALL TESTS PASSED AFTER THE MANIFOLD BOTTOM WAS REPLACED. THERE WERE NO ERROR MESSAGES OR ALARMS THAT DISPLAYED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12579949 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K140312, K160552. One submission often covers many models, so check the product and lot details in each record.
- Z-1193-2019 · Mar 21, 2019 · TerminatedEV1000 Clinical Platform, Models: EV1000A, EV1000NI, EV1000CS Product Usage: EV1000A Monitor is used to optimize fluid management through the use of minimally invasive monitori
- Z-1732-2015 · May 18, 2015 · TerminatedClearSight, Model No. EVHRS Heart Reference Sensor. The EV1000 Clinical Platform Nl and ClearSight finger cuffs noninvasively measures blood pressure and associated hemodynamic pa
Other SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE devices
- Medline MDS5001Medline Industries, Inc. · System, measurement, blood-pressure, non-invasive747 reports
- Medline MDS4003Medline Industries, Inc. · System, measurement, blood-pressure, non-invasive398 reports
- Medline MDS4001Medline Industries, Inc. · System, measurement, blood-pressure, non-invasive267 reports
- Medline MDS3001UMedline Industries, Inc. · System, measurement, blood-pressure, non-invasive170 reports
- Medline MDS1001CSMedline Industries, Inc. · System, measurement, blood-pressure, non-invasive1 recall57 reports
- Ultraview SL Ultraview SL®Spacelabs Healthcare (Washington), Inc · SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE2 recalls2 reports
- DTX(TM) PLUS DT-4812 682000Argon Critical Care Systems Singapore Pte. Ltd. · System, Measurement, Blood-Pressure, Non-Invasive86 reports
- Medline MDS7001BMedline Industries, Inc. · System, measurement, blood-pressure, non-invasive81 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page