Skip to content

Medline MDS1001CSRecalls & reported problems

Medline Industries, Inc. · DBD-MONITOR, BP, AUTO, DIGITAL, ADULT · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
57
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 57 adverse event reports that list this exact device (DI 00888277629656), received between Sep 8, 2026 and Sep 30, 2026 — classified as 56 malfunctions.

The most frequently reported specific device problem is “Computer Software Problem” (22 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 57 reportsOct 2026: 0 reports20242025202657
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2024Open, Classified

    It has been determined that certain Blood Pressure Monitors containing a faulty microchip have failed to power on or have shown display screen issues after the initial few uses, as well as overheating at times making the unit warm to the touch.

    Z-0032-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Computer Software Problem22
  • Incorrect, Inadequate or Imprecise Result or Readings12
  • Unable to Obtain Readings12
  • Gas/Air Leak6
  • High Readings5
  • Inflation Problem5
  • Electrical /Electronic Property Problem3
  • Low Readings3

Patient problems

  • No Clinical Signs, Symptoms or Conditions55
  • Discomfort1
  • Stroke/CVA1

Recent adverse event reports

The 25 most recent of 57 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Low Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50955047 ↗
  • MalfunctionSep 30, 2026Computer Software Problem; Power Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949038 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948366 ↗
  • MalfunctionSep 30, 2026Unable to Obtain Readings; Failure to Deflate

    It was reported that "CUFF PUMPS UP AND THEN CUFF DOES NOT DEFLATE. AFTER IT PUMPS UP IT DOES NOT DO ANYTHING."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947037 ↗
  • MalfunctionSep 30, 2026Computer Software Problem; Display or Visual Feedback Problem

    It was reported that the device was "returned by the customer as not functioning"; "the motherboard is not working just as the one before"; and "We saw an error message on the screen which is a \" 3+ \"."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946713 ↗
  • MalfunctionSep 30, 2026Computer Software Problem; Electrical /Electronic Property Problem; Unable to Obtain Readings

    It was reported that "Blood Pressure machine # MDS1001 that was returned by the customer as not functioning. Our technician checked and advised that the motherboard is not working just as the one before. We saw an error message on the screen which is a " 3+ "."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946390 ↗
  • MalfunctionSep 30, 2026Incorrect, Inadequate or Imprecise Result or Readings

    It was reported that the blood pressure monitor is giving them inaccurate readings. customer states that the readings are fluctuating.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945476 ↗
  • MalfunctionSep 29, 2026Computer Software Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50939782 ↗
  • MalfunctionSep 29, 2026Computer Software Problem; Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50938328 ↗
  • MalfunctionSep 29, 2026Inflation Problem; Gas/Air Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934173 ↗
  • MalfunctionSep 29, 2026Unable to Obtain Readings; High Readings; Low Readings

    It was reported that when using the cuff the reading will be really high then ER2 will show right after then she will try again it and it'll show a very low reading then ER2 error will show again. This has been happening on a regular basis at one point it showed ERR and wouldn't read.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50934034 ↗
  • MalfunctionSep 29, 2026High Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50932311 ↗
  • MalfunctionSep 28, 2026Computer Software Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923196 ↗
  • MalfunctionSep 28, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50920797 ↗
  • MalfunctionSep 28, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918100 ↗
  • MalfunctionSep 28, 2026Incorrect, Inadequate or Imprecise Result or Readings

    It was reported that 2 of the cuffs we ordered have stopped working properly and are not giving accurate readings.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918091 ↗
  • MalfunctionSep 28, 2026Unable to Obtain Readings; Fitting Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50916920 ↗
  • MalfunctionSep 28, 2026Incorrect, Inadequate or Imprecise Result or Readings

    It was reported that "The monitor is providing inconsistent and inaccurate readings."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50915832 ↗
  • MalfunctionSep 28, 2026Computer Software Problem

    It was reported that "Numbers go up and down. It over inflates - extreme pressure on my arm where I have to press the stop button."

    Patient problems: Discomfort

    MAUDE report 50913444 ↗
  • MalfunctionSep 24, 2026High Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50889730 ↗
  • MalfunctionSep 24, 2026Computer Software Problem; Unable to Obtain Readings

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50885631 ↗
  • MalfunctionSep 24, 2026Incorrect, Inadequate or Imprecise Result or Readings

    It was reported that "2 of the cuffs we ordered have stopped working properly and are not giving accurate readings."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50885335 ↗
  • OtherSep 24, 2026Unable to Obtain Readings; Incorrect, Inadequate or Imprecise Result or Readings

    It was reported that "DIGITAL BLOOD PRESSURE IS READING UP AND DOWN AND NOT GIVING A READING ACCURACY. TRIED TO LEAVE ON LONGER AND ADJUSTED THE CUFF BUT STILL NOT WORKING. DOES NOT SHOW NO ERROR CODE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884940 ↗
  • MalfunctionSep 24, 2026Overheating of Device; Battery Problem

    It was reported that "SHE CHANGED BP BATTERIES AND NOW IT DOESNT WORK AND ITS JUST GETTING HOT."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884383 ↗
  • MalfunctionSep 24, 2026Gas/Air Leak

    It was reported that "HE CAN HEAR A HISSING SOUND COMING FROM THE CUFF AND NOT GETTING ACCURATE READING. THERE IS AN ERROR MESSAGE COMING FROM THE MACHINE."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884040 ↗

Other System, measurement, blood-pressure, non-invasive devices

All System, measurement, blood-pressure, non-invasive devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page