Skip to content

EV1000 EVHRSRecalls & reported problems

Edwards Lifesciences LLC · HEART REFERENCE SENSOR · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00690103193374), received between Jan 16, 2024 and Jan 16, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 16, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THERE WERE INACCURATE VALUES WITH THE HEART REFERENCE SENSOR DURING PATIENT USE. MONITORING WAS INITIATED. THE CS READINGS AND THE NIBP READINGS CORRELATED. THE PATIENT WAS TILTED TO 70 DEGREES AND THE CLEARSIGHT READINGS WERE NOW HIGHER THAN THE NIBP AND THE MAP READINGS. A 40 POINT DIFFERENCE. THE CS MONITORING STOPPED, THE HRS WAS REZEROED WITH PROPER ALIGNMENT. THE HRS WAS AT THE PHLEBOSTATIC ACCESS AND CLIPPED TO THE PATIENT GOWN AND IT DID NOT MOVE. IT WAS FIRST PLACED WHEN THE PATIENT WAS SUPINE. THE HRS WAS THEN REZEROED SUCCESSFULLY WHILE THE PATIENT WAS UPRIGHT. THERE WAS NO PATIENT INJURY. PATIENT DEMOGRAPHICS WAS NOT PROVIDED. THERE WAS NO INAPPROPRIATE PATIENT TREATMENT ADMINISTERED. INFORMATION RECEIVED BY THE CUSTOMER IS THAT THIS HAS OCCURRED PREVIOUSLY OVER THE LAST FEW MONTHS, EXACT NUMBER IS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18522492 ↗
  • MalfunctionJan 16, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THERE WERE INACCURATE VALUES WITH THE HEART REFERENCE SENSOR DURING PATIENT USE. MONITORING WAS INITIATED. THE CS READINGS AND THE NIBP READINGS CORRELATED. THE PATIENT WAS TILTED TO 70 DEGREES AND THE CLEARSIGHT READINGS WERE NOW HIGHER THAN THE NIBP AND THE MAP READINGS. A 40 POINT DIFFERENCE. THE CS MONITORING STOPPED, THE HRS WAS REZEROED WITH PROPER ALIGNMENT. THE HRS WAS AT THE PHLEBOSTATIC ACCESS AND CLIPPED TO THE PATIENT GOWN AND IT DID NOT MOVE. IT WAS FIRST PLACED WHEN THE PATIENT WAS SUPINE. THE HRS WAS THEN REZEROED SUCCESSFULLY WHILE THE PATIENT WAS UPRIGHT. THERE WAS NO PATIENT INJURY. PATIENT DEMOGRAPHICS WAS NOT PROVIDED. THERE WAS NO INAPPROPRIATE PATIENT TREATMENT ADMINISTERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18522432 ↗

Other SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE devices

All SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page