EV1000 EVHRSRecalls & reported problems
Edwards Lifesciences LLC · HEART REFERENCE SENSOR · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00690103193374), received between Jan 16, 2024 and Jan 16, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 16, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THERE WERE INACCURATE VALUES WITH THE HEART REFERENCE SENSOR DURING PATIENT USE. MONITORING WAS INITIATED. THE CS READINGS AND THE NIBP READINGS CORRELATED. THE PATIENT WAS TILTED TO 70 DEGREES AND THE CLEARSIGHT READINGS WERE NOW HIGHER THAN THE NIBP AND THE MAP READINGS. A 40 POINT DIFFERENCE. THE CS MONITORING STOPPED, THE HRS WAS REZEROED WITH PROPER ALIGNMENT. THE HRS WAS AT THE PHLEBOSTATIC ACCESS AND CLIPPED TO THE PATIENT GOWN AND IT DID NOT MOVE. IT WAS FIRST PLACED WHEN THE PATIENT WAS SUPINE. THE HRS WAS THEN REZEROED SUCCESSFULLY WHILE THE PATIENT WAS UPRIGHT. THERE WAS NO PATIENT INJURY. PATIENT DEMOGRAPHICS WAS NOT PROVIDED. THERE WAS NO INAPPROPRIATE PATIENT TREATMENT ADMINISTERED. INFORMATION RECEIVED BY THE CUSTOMER IS THAT THIS HAS OCCURRED PREVIOUSLY OVER THE LAST FEW MONTHS, EXACT NUMBER IS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18522492 ↗MalfunctionJan 16, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THERE WERE INACCURATE VALUES WITH THE HEART REFERENCE SENSOR DURING PATIENT USE. MONITORING WAS INITIATED. THE CS READINGS AND THE NIBP READINGS CORRELATED. THE PATIENT WAS TILTED TO 70 DEGREES AND THE CLEARSIGHT READINGS WERE NOW HIGHER THAN THE NIBP AND THE MAP READINGS. A 40 POINT DIFFERENCE. THE CS MONITORING STOPPED, THE HRS WAS REZEROED WITH PROPER ALIGNMENT. THE HRS WAS AT THE PHLEBOSTATIC ACCESS AND CLIPPED TO THE PATIENT GOWN AND IT DID NOT MOVE. IT WAS FIRST PLACED WHEN THE PATIENT WAS SUPINE. THE HRS WAS THEN REZEROED SUCCESSFULLY WHILE THE PATIENT WAS UPRIGHT. THERE WAS NO PATIENT INJURY. PATIENT DEMOGRAPHICS WAS NOT PROVIDED. THERE WAS NO INAPPROPRIATE PATIENT TREATMENT ADMINISTERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18522432 ↗
Other SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE devices
- Medline MDS5001Medline Industries, Inc. · System, measurement, blood-pressure, non-invasive747 reports
- Medline MDS4003Medline Industries, Inc. · System, measurement, blood-pressure, non-invasive398 reports
- Medline MDS4001Medline Industries, Inc. · System, measurement, blood-pressure, non-invasive267 reports
- Medline MDS3001UMedline Industries, Inc. · System, measurement, blood-pressure, non-invasive170 reports
- Medline MDS1001CSMedline Industries, Inc. · System, measurement, blood-pressure, non-invasive1 recall57 reports
- Ultraview SL Ultraview SL®Spacelabs Healthcare (Washington), Inc · SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE2 recalls2 reports
- DTX(TM) PLUS DT-4812 682000Argon Critical Care Systems Singapore Pte. Ltd. · System, Measurement, Blood-Pressure, Non-Invasive86 reports
- Medline MDS7001BMedline Industries, Inc. · System, measurement, blood-pressure, non-invasive81 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page