Skip to content

PEDIASAT XT245KTPRecalls & reported problems

Edwards Lifesciences LLC · PEDIASAT OXIMETRY CATHETER KIT · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00690103197570), received between Dec 10, 2025 and Dec 10, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Air/Gas in Device” (1 report).

6 other recalls cover devices cleared under the same FDA submission (K053609, K110167); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Air/Gas in Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 10, 2025Air/Gas in Device; Air/Gas in Device

    AS REPORTED, AFTER CATHETER PLACEMENT DURING USE IN PATIENT WHEN BLOOD BACKFLOW WAS CHECKED, A PEDIASAT OXIMETRY CATHETER ASPIRATED AIR. IT IS UNCLEAR WHICH LUMEN (DISTAL, MEDIAL, PROXIMAL) ASPIRATED AIR. NO MEDICATION LEAKAGE OCCURRED. AFTER THE CATHETER REMOVAL, THE CUSTOMER PERFORMED A TEST IN THE WATER AND FOUND A HOLE ON THE PATIENT (DISTAL) SIDE OF BACKFORM, AND LEAKAGE WAS OBSERVED FROM THE SAME LOCATION. THERE WAS NO ALLEGATION OF PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23770701 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K053609, K110167. One submission often covers many models, so check the product and lot details in each record.

  • Z-2280-2021 · Jul 9, 2021 · Terminated
    Edwards PediaSat Oximetry Catheter Kit, REF: XT358SJ, Usable Length 8 cm, Number of Lumens 3, Exterior Diameter 5.5F, Rx Only, Sterile EO, UDI: (01)00690103197495 *NOT DISTRIBUT
  • Z-2283-2021 · Jul 9, 2021 · Terminated
    Edwards PediaSat Oximetry Catheter Set, REF: XT3515SJ, Exterior Diameter 5.5F, Usable Length 15 cm, Number of Lumens: 3, Rx Only, Sterile EO, UDI: (01)00690103197501 *NOT DISTR
  • Z-2281-2021 · Jul 9, 2021 · Terminated
    Edwards PediaSat Oximetry Catheter Set, REF: XT358SP, Exterior Diameter 5.5F (1.83 mm), Usable Length 9 cm, Number of Lumens: 3, Recommended Guidewire Size: GW: 0.025" (0.64 mm),
  • Z-2279-2021 · Jul 9, 2021 · Terminated
    Edwards PediaSat Oximetry Catheter Kit, REF: XT3515SP, Exterior Diameter 5.5F (1.83 mm), Usable Length 15 cm, Number of Lumens 3, Recommended Guidewire Size 0.025" (0.64 mm), Rx On
  • Z-2278-2021 · Jul 9, 2021 · Terminated
    Edwards PediaSat Oximetry Catheter Kit, REF: XT3515KTP, Exterior Diameter 5.5F (1.83 mm), Usable Length 15 cm, Number of Lumens 3, Recommended Guidewire Size 0.025" (0.64 mm), Rx O
  • Z-2282-2021 · Jul 9, 2021 · Terminated
    Edwards PediaSat Oximetry Catheter Set, REF: XT358KTP, Exterior Diameter 5.5F, Usable Length 8 cm, Number of Lumens: 3, Recommended Guidewire Size: GW: 0.025" (0.64 mm), Rx Only,

Other DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION devices

All DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page