AVANTI 504656XRecalls & reported problems
Cordis Corporation · CSI AVANTI+ CARDIOLOGY MS .038 · Class 2
All models of AVANTI (19)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705032010382), received between Sep 28, 2023 and Oct 17, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
1 FDA recall record name this device's identifier.
1 other recall cover devices cleared under the same FDA submission (K970392); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Dec 31, 2015Class IITerminated
Incorrect cannula of the sheath introducer (smaller than intended).
Z-1174-2016 · Root cause: Mixed-up of materials/components · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- Pseudoaneurysm1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 17, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, THE DISTAL TIP OF A .038 AVANTI+ CARDIOLOGY MS CATHETER SHEATH INTRODUCER (CSI) SHEARED WHILE BE REMOVED FROM THE PATIENT AFTER A CARDIAC ABLATION PROCEDURE WAS COMPLETED. THE FRACTURED PIECE MIGRATED TO THE RIGHT ATRIUM. ADDITIONAL INTERVENTION WAS REQUIRED TO REMOVE THE SHEARED PIECE. THE FRACTURED PIECE WAS REMOVED USING ULTRASOUND GUIDANCE, BEGINNING WITH A 21-GAUGE MICRO-PUNCTURE NEEDLE TO ACCESS THE RIGHT INTERNAL JUGULAR VEIN, FOLLOWED BY ADVANCING A MICRO-PUNCTURE WIRE INTO THE SUPERIOR VENA CAVA. A MICRO-PUNCTURE TRANSITION SET AND AN UNKNOWN 8F SHORT-SHEATH WERE THEN PLACED. ONCE THE FRACTURED PORTION WAS IDENTIFIED IN THE RIGHT ATRIUM, A RIGHT GROIN ACCESS SHEATH WAS UTILIZED TO ADVANCE A WIRE, REMOVE THE ORIGINAL SHEATH, AND PLACE AN UNKNOWN 12F SHEATH. AN ENSNARE DEVICE WAS USED THROUGH THIS SHEATH TO GRASP AND REMOVE THE FRACTURED COMPONENT COMPLETELY. A CONTRALATERAL APPROACH WAS ALSO USED, PLACING THREE UNKNOWN 6F SHEATHS IN THE RIGHT FEMORAL VEIN WITHOUT DIFFICULTY, AIDED BY INTRAVASCULAR ULTRASOUND. A NON-CORDIS CATHETER WAS PLACED IN THE CORONARY SINUS, AND AN 8F NON-CORDIS SHEATH WAS ADVANCED INTO THE RIGHT ATRIUM. USING A NON-CORDIS SHEATH INTRODUC
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20474690 ↗InjurySep 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AS REPORTED, A 6F/7F MYNXGRIP VASCULAR CLOSURE DEVICE (VCD) WAS USED WITH 6F CORDIS AVANTI CATHETER SHEATH INTRODUCER (CSI) AND THE PATIENT DEVELOPED A PSEUDOANEURYSM. A THROMBIN INJECTION WAS REQUIRED. THERE WAS NO OTHER REPORTED PATIENT INJURY. THE PROCEDURE WAS A DIAGNOSTIC HEART CATHETERIZATION. THERE WAS NO ANTICOAGULATION. THE USER IS TRAINED TO THE USE OF THE MYNX DEVICE. THE FEMORAL ARTERY¿S SUITABILITY WAS VERIFIED ON ANGIOGRAPHY, INCLUDING THE INSERTION ANGLE (30-45 DEGREES) OF THE VASCULAR SHEATH INTRODUCER. THE VESSEL DIAMETER WAS VERIFIED TO BE GREATER THAN OR EQUAL TO 5 MM IN DIAMETER. THERE WAS NO VESSEL TORTUOSITY AND NO PRESENCE OF PVD / CALCIUM IN THE VICINITY OF THE PUNCTURE SITE. THE TARGET FEMORAL SITE HAD NOT BEEN PREVIOUSLY CLOSED WITH ANY CLOSURE PRIOR TO THIS PROCEDURE. THERE WERE NO MULTIPLE SHEATH EXCHANGES. A SINGLE STICK WAS USED FOR ACCESS. THE DEVICES WILL NOT BE RETURNED FOR EVALUATION.
Patient problems: Pseudoaneurysm
MAUDE report 17834230 ↗
FDA clearance and classification
Cleared through 510(k) K970392 on Apr 24, 1997 (decision: Substantially Equivalent). Submitted by Cordis Corp. as “Cordis Avanti+Catheter Sheath Introducer System”. The FDA review took 80 days. FDA record ↗
The FDA regulates product code DRE (dilator, vessel, for percutaneous catheterization) under 21 CFR 870.1310 as a Class II device. 21 CFR 870.1310 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K970392. One submission often covers many models, so check the product and lot details in each record.
- Z-2079-2021 · Jun 9, 2021 · TerminatedAvanti+ Catheter Sheath Introducer - Product Usage: Is indicated for use in arterial and venous procedures requiring percutaneous introduction of intravascular devices.
Common questions about the AVANTI 504656X
Has the AVANTI 504656X been recalled?
Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-1174-2016 (Class II), began on Dec 31, 2015: Incorrect cannula of the sheath introducer (smaller than intended). If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the AVANTI 504656X?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Sep 28, 2023 and Oct 17, 2024. In those reports, the most common specific device problem is “Detachment of Device or Device Component” (1); the most common patient problem is “Pseudoaneurysm” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the AVANTI 504656X?
None of the reports are classified as deaths, 1 as an injury and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the AVANTI 504656X FDA approved?
It was cleared through 510(k) K970392 on Apr 24, 1997. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the AVANTI 504656X, and is it still sold?
The FDA's device database lists Cordis Corporation as the company responsible for it (brand AVANTI, model 504656X). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION devices
- VersaCross Connect LAAC Access Solution VXAK0007Baylis Médicale Cie Inc · Dilator, Vessel, For Percutaneous Catheterization3 deaths108 reports
- TandemLife 5740-2417Cardiacassist, Inc. · Dilator, Vessel, For Percutaneous Catheterization2 recalls0 reports
- TightRail Rotating Dilator Sheath 545-513Spectranetics Corporation · Dilator, Vessel, For Percutaneous Catheterization20 deaths74 reports
- VersaCross Connect LAAC Access Solution VXAK0003Baylis Médicale Cie Inc · Dilator, Vessel, For Percutaneous Catheterization2 deaths70 reports
- VersaCross Connect LAAC Access Solution VXAK0007Baylis Médicale Cie Inc · Dilator, Vessel, For Percutaneous Catheterization2 deaths54 reports
- Abiomed® 00884450732311Merit Medical Systems, Inc. · Dilator, Vessel, For Percutaneous Catheterization1 recall2 reports
- Mini Stick MAX H965457490Angiodynamics, Inc. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 recall1 report
- Mini Stick MAX H965457510Angiodynamics, Inc. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 recall1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page