Skip to content

Acumen IQ AIQS8Recalls & reported problems

Edwards Lifesciences LLC · Acumen IQ Sensor · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00690103201086), received between Aug 22, 2024 and Aug 22, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Incorrect Measurement” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K152980); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 4, 2023Open, Classified

    Their is the potential for flushing difficulties during setup of the pressure monitoring kit with pressure transducer.

    Z-0409-2024 · Root cause: Device Design · FDA record ↗

Most reported problems

Device problems

  • Incorrect Measurement1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 22, 2024Incorrect Measurement; Incorrect Measurement

    AS REPORTED, DURING USE IN PATIENT, THIS ACUMEN IQ SENSOR WAS WORKING AS EXPECTED PROVIDING CORRECT VITAL PARAMETERS. HOWEVER, AFTER A FEW HOURS, THE SENSOR STARTED PROVIDING INACCURATE PARAMETERS THAT WERE NOT COMPATIBLE WITH LIFE AND DIFFERENT FROM OTHER REFERENCES THAT WERE IN THE OPERATING ROOM (OR). THE DEVICE WAS REPLACED WITH A NEW UNIT AND ISSUE WAS SOLVED. THERE IS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION. PATIENT DEMOGRAPHICS WERE UNABLE TO BE OBTAINED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20026096 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K152980. One submission often covers many models, so check the product and lot details in each record.

  • Z-0412-2024 · Oct 4, 2023 · Open, Classified
    FloTrac Sensor with Vamp System, REF: MHD6AZ, MHD6AZ5 , MHD6C502 STIRLEEO, Rx only. For use in intravascular pressure monitoring.
  • Z-0410-2024 · Oct 4, 2023 · Open, Classified
    Acumen IQ Sensor with VAMP System, REFs: AIQS6AZ & AIQS6AZ5, STERILEEO, Rx only. For use in intravascular pressure monitoring.
  • Z-0411-2024 · Oct 4, 2023 · Open, Classified
    FloTrac Sensor, REF: MHD8 & MHD85,MHD6R5, MHD8R, MHD8R5, MHD8S, STERILEEO, Rx only. For use in intravascular pressure monitoring.

Other Transducer, Blood-Pressure, Extravascular devices

All Transducer, Blood-Pressure, Extravascular devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.