Acumen IQ recallZ-0409-2024 · Edwards Lifesciences, LLC · initiated Oct 4, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Edwards Lifesciences, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Their is the potential for flushing difficulties during setup of the pressure monitoring kit with pressure transducer.
FDA root cause category: Device Design
What the company is doing
On 10/18/2023, Edwards Lifesciences issued a "Urgent: Medical device Correction Action via FedEx mail. Edwards ask consignees to take the following actions: 1. Follow the IFU for proper setup and flushing of your device. 2. Follow the instructions included in the enclosed customer acknowledgement form to complete the acknowledgement process. 3. Verify your inventory on the attached customer acknowledgement form. 3a.E-mail the completed form to Edwards Customer Service at [email protected], within 15 days from receipt of this notification. 4. Distribute this notice within your organization or to any organization where the potentially impacted product has been transferred. 5. Please post a copy of this notice near and/or with affected devices. 6. Contact Edwards Customer Service at 1-888-352-0904 if you would like to return any of your devices or have further questions. 7. Action Requested for Distributors: In addition to the actions above, please forward this customer communication to any of your customers who have purchased the impacted product.
Recall details
- Recall number
- Z-0409-2024
- Status
- Open, Classified
- Initiated by the company
- Oct 4, 2023
- Posted by the FDA
- Nov 29, 2023
- Recalling firm
- Edwards Lifesciences, LLC
- Quantity
- 33,327 units
- Product code
- DRS
- FDA submissions
- K152980
- Distribution
- US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA WA WI WV WY OUS: Argentina Australia Austria Bahrain Belgium Brazil Canada CANARY ISLANDS Chile Colombia Costa Rica Czech Republic Denmark Ecuador Faroe Islands France Germany Greece Guatemala Hong Kong Indonesia Ireland Israel Italy Japan Malaysia Netherlands New Zealand Northern Ireland Norway Panama Poland Portugal Saudi Arabia Singapore Slovakia Slovenia South Korea Spain Sweden Switzerland United Arab Emirates (UAE) United Kingdom Vietnam
Product description
Acumen IQ Sensor, REFs: AIQS6, AIQS65, AIQS6R, AIQS6R5, AIQS8, AIQS85, AIQS8R5, Rx only, Sterile EO. For use in intravascular pressure monitoring.
Code information (models, lots, serial numbers)
AIQS6: UDI:00690103201079; Lot #s: 64858340 64933520 65027617; AIQS65: UDI:00690103200935; Lot # 64858715; AIQS6R: UDI:00690103201093/Lot #: 64858050 AIQS6R5: UDI:00690103200959/Lot #: 64835570 AIQS8: 00690103201086 UDI:/Lot # 64858150 64858716 64891667 64939370 65179001; AIQS85: UDI:00690103200942/Lot# 64858717 64892324 64892325 64892326 64892327 64904497 64904498 64918260 64918261 64918262 64918263 64918264 64918646 64918647 64918648 64918649 64918650 64918651 64918652 64988038 64988040 65067612; AIQS8R5: 00690103201178/UDI:; Lot # 64817843 64858645 64891410
Devices named in this recall
7 devices in the FDA device registry whose identifier appears in the recall's code information.
- Acumen IQ AIQS85Edwards Lifesciences LLC · Transducer, Blood-Pressure, Extravascular1 recall7 reports
- Acumen IQ AIQS8Edwards Lifesciences LLC · Transducer, Blood-Pressure, Extravascular1 recall1 report
- Acumen IQ AIQS8R5Edwards Lifesciences LLC · Transducer, Blood-Pressure, Extravascular1 recall1 report
- Acumen IQ AIQS65Edwards Lifesciences LLC · Transducer, Blood-Pressure, Extravascular1 recall0 reports
- Acumen IQ AIQS6R5Edwards Lifesciences LLC · Transducer, Blood-Pressure, Extravascular1 recall0 reports
- Acumen IQ AIQS6Edwards Lifesciences LLC · Transducer, Blood-Pressure, Extravascular1 recall0 reports
- Acumen IQ AIQS6REdwards Lifesciences LLC · Transducer, Blood-Pressure, Extravascular1 recall0 reports
Product lines
Other devices under the same FDA submission
10 devices cleared under K152980 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- FLOTRAC MHD8Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall5 reports
- FLOTRAC, VAMP MHD6AZEdwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall4 reports
- FLOTRAC, VAMP, TRUWAVE MHD6C502Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall2 reports
- FLOTRAC, VAMP MHD6AZ5Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 recall1 report
- ACUMEN IQ SENSOR AIQS6AZ5Edwards Lifesciences LLC · Transducer, Blood-Pressure, Extravascular1 recall1 report
- FLOTRAC MHD85Edwards Lifesciences LLC · CATHETER, OXIMETER, FIBEROPTIC1 recall0 reports
- ACUMEN IQ SENSOR AIQS6AZEdwards Lifesciences LLC · Transducer, Blood-Pressure, Extravascular1 recall0 reports
- FLOTRAC MHD6Edwards Lifesciences LLC · CATHETER, OXIMETER, FIBEROPTIC1 report
- FLOTRAC MHD65Edwards Lifesciences LLC · COMPUTER, DIAGNOSTIC, PRE-PROGRAMMED, SINGLE-FUNCTION1 report
- Acumen IQ AIQS6C502Edwards Lifesciences LLC · Transducer, Blood-Pressure, Extravascular1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 93279).
- Z-0410-2024 · Open, ClassifiedAcumen IQ Sensor with VAMP System, REFs: AIQS6AZ & AIQS6AZ5, STERILEEO, Rx only. For use in intravascular pressure monitoring.
- Z-0411-2024 · Open, ClassifiedFloTrac Sensor, REF: MHD8 & MHD85,MHD6R5, MHD8R, MHD8R5, MHD8S, STERILEEO, Rx only. For use in intravascular pressure monitoring.
- Z-0412-2024 · Open, ClassifiedFloTrac Sensor with Vamp System, REF: MHD6AZ, MHD6AZ5 , MHD6C502 STIRLEEO, Rx only. For use in intravascular pressure monitoring.
- Z-0413-2024 · Open, ClassifiedTruWave Pressure monitoring set, REF: PX212: cc/12 in (30cm), PX260: (3 cc/60in (150cm), PX272: 3 cc/72in (180cm), PX284: 3 cc/84 in (210 cm), PX2X2: x2 (3cc), PX3X3284C: X3 (3cc),…
- Z-0414-2024 · Open, ClassifiedTruWave with Vamp System Pressure monitoring set, REF: PXAVMP(3cc)/VAMP 68 in (173cm) PXAVMP3: x3 (3cc)/VAMP PXVK0364:(3cc)/VAMP 96in (241cm); PXVK0853: x3 (3cc)/VAMP 76in (193cm)…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page