Skip to content

HemoSphere Alta ALTASR1Recalls & reported problems

Edwards Lifesciences LLC · HemoSphere Alta advanced monitor (Smart Recovery) · Class 2

All models of HemoSphere Alta (3)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00690103217964), received between Nov 26, 2025 and Feb 12, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Incorrect Measurement” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect Measurement2
  • Incorrect, Inadequate or Imprecise Result or Readings2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 12, 2026Incorrect Measurement; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT, DURING USE, THE HEMOSPHERE ALTA MONITOR EXPERIENCED SIGNIFICANT MEAN ARTERIAL PRESSURE (MAP) VARIABILITY WHEN USING CLEARSIGHT. SPECIFICALLY, ARTERIAL LINE CORRELATION DRIFTING BY 20 TO 40 MMHG. IT WAS ALSO NOTED THAT THE SUSPECT DEVICE EXPERIENCED FAULT: CLEARSIGHT SYSTEM AIR SUPPLY ERROR AND INSUFFICIENT PRESSURE BUILD UP. THERE ARE NO ALLEGATIONS OF PATIENT INJURY OR HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24328911 ↗
  • MalfunctionNov 26, 2025Incorrect Measurement; Incorrect, Inadequate or Imprecise Result or Readings; Incorrect Measurement; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT PRIOR TO USE, THE HEMOSPHERE ALTA SYSTEM WITH CLEARSIGHT DISPLAYED A SUBSYSTEM FAILURE ALERT. THE END USER RESTARTED THE DEVICE AND CLEARED THE ALERT; HOWEVER, SUBSEQUENT MEASUREMENTS SHOWED VARIATIONS IN BLOOD PRESSURE (BP) AND PULSE RATE (PR). THE ISSUE COULD NOT BE RESOLVED EVEN AFTER REPLACING THE PC2K MODULE. ADDITIONAL FOLLOW-UP EFFORTS HAVE FOCUSED ON DETERMINING PATIENT INVOLVEMENT AND WHETHER ANY INJURY OR HARM OCCURRED AS A RESULT OF THE REPORTED EVENT. AT THIS TIME, NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23659549 ↗

Other Catheter, Oximeter, Fiber-Optic devices

All Catheter, Oximeter, Fiber-Optic devices →

Sources and updates