SWAN-GANZ VIP 831F75 (REF 831F75P)Recalls & reported problems
Edwards Lifesciences LLC · SWAN-GANZ THERMODILUTION VEINOUS FUSION PORT CATHETER · Class 2
All models of SWAN-GANZ VIP (4)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00690103218282), received between Mar 4, 2025 and Sep 23, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).
1 FDA recall record name this device's identifier, 1 still open.
5 other recalls cover devices cleared under the same FDA submission (K160084, K231248); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Apr 8, 2026Class IIOpen, Classified
Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.
Z-2202-2026 · Root cause: Process change control · Recall details · FDA record ↗
Most reported problems
Device problems
- Fluid/Blood Leak3
- Contamination1
- Detachment of Device or Device Component1
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2026Contamination
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50872949 ↗MalfunctionAug 20, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT, DURING USE, A SWAN GANZ CATHETER HAD SALINE AND MEDICATION LEAK AT THE CVP PORT, WHERE THE WHITE FACTORY INSTALLED CAP JOINS THE BLUE TUBING. LOT NUMBER MAY BE 66225267 AS THIS WAS THE REMAINING LOT ON THE SHELF. THERE WERE NO PATIENT INJURIES. THIS EVENT OCCURRED IN THE PAST SO SPECIFIC INFORMATION IS NOT AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22857180 ↗MalfunctionAug 20, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT, DURING USE, A SWAN GANZ CATHETER HAD SALINE AND MEDICATION LEAK AT THE CVP PORT, WHERE THE WHITE CAP JOINS THE BLUE TUBING. LEAKAGE OCCURRED 5 DAYS AFTER USE. CATHETER WAS REMOVED AS THE STAFF WAS UNABLE TO USE THE PORT. LOT NUMBER MAY BE 66225267 AS THIS WAS THE REMAINING LOT ON THE SHELF. THERE WERE NO PATIENT INJURIES, BUT PATIENT MAY HAVE HAD A NEW SWAN INSERTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22857166 ↗MalfunctionAug 20, 2025Fluid/Blood Leak; Fluid/Blood Leak
IT WAS REPORTED THAT, DURING USE, A SWAN GANZ CATHETER HAD SALINE AND MEDICATION LEAK AT THE CVP PORT, WHERE THE WHITE FACTORY INSTALLED CAP JOINS THE BLUE TUBING. LEAKAGE OCCURRED 3 DAYS AFTER USE. CATHETER WAS REMOVED AS THE STAFF WAS UNABLE TO USE THE PORT. LOT NUMBER MAY BE 66225267 AS THIS WAS THE REMAINING LOT ON THE SHELF. THERE WERE NO PATIENT INJURIES, BUT PATIENT MAY HAVE HAD A NEW SWAN INSERTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22857113 ↗MalfunctionApr 23, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, UPON OPENING AND INSPECTION OF THIS SWAN-GANZ CATHETER, MATERIAL PROTRUDING FROM THE CATHETER WAS FOUND CLOSE TO THE DISTAL TIP. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION. AS PER EVALUATION RESULTS, APPROXIMATELY 4MM IN LENGTH OF THE CATHETER BODY APPEARED TO BE MISSING AT 2CM FROM THE DISTAL TIP. PATIENT DEMOGRAPHIC DATA UNABLE TO BE OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21890438 ↗MalfunctionMar 4, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT, DURING USE, THIS SWAN GANZ CATHETER PROVIDED INACCURATE CARDIAC OUTPUT READINGS. THE PATIENT RECEIVED MEDICATION MANAGEMENT, ADDITIONAL TESTING (ECHOCARDIOGRAM), AND EXTUBATION WAS DELAYED AS A RESULT OF THE INACCURATE VALUES. NO PATIENT INJURY WAS REPORTED. THE DEVICE IS NOT AVAILABLE FOR RETURN AS IT WAS DISCARDED BY THE FACILITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21520796 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K160084, K231248. One submission often covers many models, so check the product and lot details in each record.
- Z-2203-2026 · Apr 8, 2026 · Open, ClassifiedSwan-Ganz Jr Catheter, Models: SGPT54, SGPT64P, SGPT755P
- Z-0762-2019 · Dec 21, 2018 · TerminatedEdwards Lifesciences Swan Ganz Thermodilution Catheter, REF131F 7J Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly det
- Z-0763-2019 · Dec 21, 2018 · TerminatedEdwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131F7P (5 pack of 131F7) Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians
- Z-0764-2019 · Dec 21, 2018 · TerminatedEdwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131VF7P Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly de
- Z-0765-2019 · Dec 21, 2018 · TerminatedEdwards Lifesciences Swan Ganz Thermodilution Catheter, REF 151F7 Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly dete
Other CATHETER, FLOW DIRECTED devices
- Graftmaster 1012580-19Abbott Vascular Inc. · CATHETER, FLOW DIRECTED30 deaths225 reports
- SWAN-GANZ CCOMBO V 777F8Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED2 recalls92 reports
- Graftmaster 1012581-16Abbott Vascular Inc. · CATHETER, FLOW DIRECTED21 deaths147 reports
- SWAN-GANZ CCOMBO V 774F75Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED2 recalls2 deaths41 reports
- Graftmaster 1012581-19Abbott Vascular Inc. · CATHETER, FLOW DIRECTED12 deaths135 reports
- Graftmaster 1012582-19Abbott Vascular Inc. · CATHETER, FLOW DIRECTED13 deaths77 reports
- SWAN-GANZ VIP 831F75Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED1 recall23 reports
- Graftmaster 1012580-26Abbott Vascular Inc. · CATHETER, FLOW DIRECTED13 deaths72 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page