FreeStyle Lite 70827Recalls & reported problems
Abbott Diabetes Care Inc · FreeStyle Lite Blood Glucose Test Strips · Class 2
Quick summary
This Abbott FreeStyle Lite test strip is used with a meter to measure blood glucose. The 234 reports mostly describe incorrect or erratic readings, some far from lab results, and one describes a low reading followed by a seizure. The reports include 67 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 234
- Deaths reported
- 0
- Injuries reported
- 67
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 234 adverse event reports that list this exact device (DI 00699073708274), received between Mar 6, 2017 and Mar 11, 2024 — classified as 67 injuries, 167 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (178 reports).
2 other recalls cover devices cleared under the same FDA submission (K092602); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings178
- High Readings16
- Device Displays Incorrect Message12
- Low Readings11
- No Device Output9
- Unable to Obtain Readings6
- Packaging Problem3
- Failure to Power Up2
Patient problems
- No Known Impact Or Consequence To Patient108
- No Clinical Signs, Symptoms or Conditions36
- Hypoglycemia26
- Hyperglycemia24
- Loss of consciousness23
- Insufficient Information15
- Dizziness10
- Blurred Vision7
Recent adverse event reports
The 25 most recent of 234 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 11, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED RECEIVING A HIGH READING FROM THEIR ADC DEVICE. CUSTOMER REPORTED A HIGH READING OF 219 MG/DL WHICH WAS HIGHER THAN A LAB READING OF 95 MG/DL. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18874950 ↗MalfunctionNov 3, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18065385 ↗MalfunctionNov 21, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC DEVICE. CUSTOMER REPORTED RECEIVING READINGS OF 88 MG/DL, 110 MG/DL, 398 MG/DL AND 127 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15834763 ↗InjuryNov 14, 2022Low Readings; Low Readings
A CALLER REPORTED A LOW READING WITH THE ADC METER, WHICH WAS LOWER THAN CUSTOMER FELT. THE CALLER REPORTED A METER RESULT OF 55 MG/DL AND CUSTOMER REPORTEDLY HAD SEIZURE DUE TO THIS ISSUE. PARAMEDICS WERE CALLED BUT REPORTEDLY DID NOT PROVIDE TREATMENT. NO FURTHER DETAILS WERE PROVIDED. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: Convulsion/Seizure
MAUDE report 15794101 ↗MalfunctionOct 12, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC DEVICE. CUSTOMER REPORTED RECEIVING READINGS OF "LO" (BELOW 20 MG/DL) AND 92 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15587549 ↗MalfunctionAug 18, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15247743 ↗MalfunctionJul 29, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC DEVICE. CUSTOMER REPORTED RECEIVING READINGS OF 382 MG/DL AND 110 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: Insufficient Information
MAUDE report 15129260 ↗MalfunctionJul 19, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC DEVICE. CUSTOMER REPORTED RECEIVING READINGS OF 316 MG/DL, 29 MG/DL AND 297 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15059254 ↗MalfunctionJun 15, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC DEVICE. CUSTOMER REPORTED RECEIVING READINGS OF 358 MG/DL, 257 MG/DL AND 119 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14698064 ↗MalfunctionApr 6, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC DEVICE. CUSTOMER REPORTED RECEIVING READINGS OF 16 MG/DL, 58 MG/DL AND 125 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14022290 ↗MalfunctionMar 23, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13868265 ↗InjuryMar 11, 2022Low Readings; Low Readings
A CALLER REPORTED ON BEHALF OF CUSTOMER LOW READINGS WITH THE ADC DEVICE. THE CALLER REPORTED UNSPECIFIED BLOOD GLUCOSE RESULTS WITH THE ADC METER, THAT WERE LOWER THAN CUSTOMER FELT. THE CALLER REPORTED CUSTOMER EXPERIENCED EITHER LOSS OF CONSCIOUSNESS OR SEIZURE DUE TO THIS ISSUE, BUT REFUSED TO PROVIDE ANY FURTHER INFORMATION. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: Convulsion/Seizure; Loss of consciousness
MAUDE report 13741701 ↗MalfunctionMar 3, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13659381 ↗MalfunctionFeb 16, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 412 MG/DL AND 132 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13542192 ↗MalfunctionFeb 4, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13449790 ↗MalfunctionFeb 2, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 90 MG/DL, 45 MG/DL, 37 MG/DL AND 123 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: Insufficient Information
MAUDE report 13423125 ↗MalfunctionJan 24, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 450 MG/DL AND 128 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13338412 ↗InjuryJan 10, 2022Low Readings; Low Readings
A CUSTOMER REPORTED A DELIVERY DELAY WITH A REPLACEMENT ADC BLOOD GLUCOSE METER. THE REPLACEMENT DEVICE WAS ISSUED AS CUSTOMER HAD REPORTED LOW GLUCOSE RESULTS. DUE TO DELIVERY DELAY, CUSTOMER REPORTED BEING HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE AS SHE HAD NO METHOD TO MONITOR GLUCOSE. THE CUSTOMER DID NOT PROVIDE TREATMENT INFORMATION BUT DID STATE SHE WAS PRESCRIBED INSULIN AFTER HOSPITALIZATION. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: Hyperglycemia
MAUDE report 13213604 ↗InjuryJan 7, 2022Unable to Obtain Readings; Unable to Obtain Readings
A CUSTOMER REPORTED AN "ERROR 3" MESSAGE WITH THE ADC BLOOD GLUCOSE METER, WHEN APPLYING BLOOD SAMPLE TO TEST STRIP. THE CUSTOMER WAS THEREFORE UNABLE TO OBTAIN GLUCOSE RESULT, AND EXPERIENCED A LOSS OF CONSCIOUSNESS. THE CUSTOMER REPORTED REQUIRING TREATMENT FROM A HEALTHCARE PROVIDER, BUT DECLINED TO PROVIDE FURTHER DETAILS. PLEASE NOTE THAT UPON TROUBLESHOOTING IT WAS FOUND THAT CUSTOMER WAS APPLYING BLOOD SAMPLE TO INCORRECT POSITION ON TEST STRIP. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: Loss of consciousness
MAUDE report 13195364 ↗MalfunctionDec 28, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 546 MG/DL AND 121 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13101614 ↗MalfunctionDec 22, 2021Failure to Power Up; Failure to Power Up
THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13058295 ↗InjuryNov 9, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED AN UNSPECIFIED READING ISSUE WITH THE ADC BLOOD GLUCOSE METER. THE CUSTOMER REPORTED THAT DUE TO THIS ISSUE, SHE "BLACKED OUT" AND WAS TAKEN TO HOSPITAL BY AMBULANCE WHERE SHE WAS TREATED WITH INSULIN. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.
Patient problems: Hyperglycemia; Loss of consciousness
MAUDE report 12778928 ↗MalfunctionNov 3, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12742728 ↗MalfunctionOct 14, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 112 MG/DL AND 19 (LO) MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: Insufficient Information
MAUDE report 12634735 ↗MalfunctionSep 30, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12553397 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K092602. One submission often covers many models, so check the product and lot details in each record.
- Z-0485-2014 · Nov 18, 2013 · TerminatedFreeStyle Blood Glucose Test Strips; For in vitro diagnostic testing. 50 count UPC 6 99073 12050 2; Freestyle Blood Glucose Test Strips; for Institutional Use only 50 count: UPC 6
- Z-0486-2014 · Nov 18, 2013 · TerminatedFreeStyle Lite Blood Glucose Test Strips; For in vitro diagnostic testing. 50 count; 100 count Product of Ireland; UPC 6 99073 70819 9; - 50 count UPC 6 99073 70822 9 - 50 count
Other Glucose dehydrogenase, glucose devices
- OneTouch Verio IQ 022-275LifeScan Europe GmbH · Glucose dehydrogenase, glucose6,035 reports
- OneTouch Verio 022-659LifeScan Europe GmbH · Glucose dehydrogenase, glucose4,219 reports
- TRUE METRIX TRUE METRIX SELF MONITORING BLOOD GLUCOSE TEST STRIPS - 50 Test StripsTrividia Health, Inc. · Glucose Dehydrogenase, Glucose2 deaths2,548 reports
- FreeStyle Lite 70805Abbott Diabetes Care Inc · Glucose dehydrogenase, glucose2,366 reports
- FreeStyle Freedom Lite 70920Abbott Diabetes Care Inc · Glucose dehydrogenase, glucose1,037 reports
- FreeStyle Freedom Lite 70914Abbott Diabetes Care Inc · Glucose dehydrogenase, glucose760 reports
- Accu-Chek Guide 08256292001Roche Diagnostics Operations, Inc. · Glucose Dehydrogenase, Glucose476 reports
- ACCU-CHEK AVIVA PLUS 06908268001Roche Diagnostics Operations, Inc. · Glucose Dehydrogenase, Glucose472 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.