FreeStyle Lite 70827Recalls & reported problems

Abbott Diabetes Care Inc · FreeStyle Lite Blood Glucose Test Strips · Class 2

Quick summary

This Abbott FreeStyle Lite test strip is used with a meter to measure blood glucose. The 234 reports mostly describe incorrect or erratic readings, some far from lab results, and one describes a low reading followed by a seizure. The reports include 67 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
234
Deaths reported
0
Injuries reported
67
Recalls naming it
0

Summary

The FDA's MAUDE database contains 234 adverse event reports that list this exact device (DI 00699073708274), received between Mar 6, 2017 and Mar 11, 2024 — classified as 67 injuries, 167 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (178 reports).

2 other recalls cover devices cleared under the same FDA submission (K092602); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings178
  • High Readings16
  • Device Displays Incorrect Message12
  • Low Readings11
  • No Device Output9
  • Unable to Obtain Readings6
  • Packaging Problem3
  • Failure to Power Up2

Patient problems

  • No Known Impact Or Consequence To Patient108
  • No Clinical Signs, Symptoms or Conditions36
  • Hypoglycemia26
  • Hyperglycemia24
  • Loss of consciousness23
  • Insufficient Information15
  • Dizziness10
  • Blurred Vision7

Recent adverse event reports

The 25 most recent of 234 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 11, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED RECEIVING A HIGH READING FROM THEIR ADC DEVICE. CUSTOMER REPORTED A HIGH READING OF 219 MG/DL WHICH WAS HIGHER THAN A LAB READING OF 95 MG/DL. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18874950 ↗
  • MalfunctionNov 3, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18065385 ↗
  • MalfunctionNov 21, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC DEVICE. CUSTOMER REPORTED RECEIVING READINGS OF 88 MG/DL, 110 MG/DL, 398 MG/DL AND 127 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15834763 ↗
  • InjuryNov 14, 2022Low Readings; Low Readings

    A CALLER REPORTED A LOW READING WITH THE ADC METER, WHICH WAS LOWER THAN CUSTOMER FELT. THE CALLER REPORTED A METER RESULT OF 55 MG/DL AND CUSTOMER REPORTEDLY HAD SEIZURE DUE TO THIS ISSUE. PARAMEDICS WERE CALLED BUT REPORTEDLY DID NOT PROVIDE TREATMENT. NO FURTHER DETAILS WERE PROVIDED. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.

    Patient problems: Convulsion/Seizure

    MAUDE report 15794101 ↗
  • MalfunctionOct 12, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC DEVICE. CUSTOMER REPORTED RECEIVING READINGS OF "LO" (BELOW 20 MG/DL) AND 92 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15587549 ↗
  • MalfunctionAug 18, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15247743 ↗
  • MalfunctionJul 29, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC DEVICE. CUSTOMER REPORTED RECEIVING READINGS OF 382 MG/DL AND 110 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: Insufficient Information

    MAUDE report 15129260 ↗
  • MalfunctionJul 19, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC DEVICE. CUSTOMER REPORTED RECEIVING READINGS OF 316 MG/DL, 29 MG/DL AND 297 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15059254 ↗
  • MalfunctionJun 15, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC DEVICE. CUSTOMER REPORTED RECEIVING READINGS OF 358 MG/DL, 257 MG/DL AND 119 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14698064 ↗
  • MalfunctionApr 6, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC DEVICE. CUSTOMER REPORTED RECEIVING READINGS OF 16 MG/DL, 58 MG/DL AND 125 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14022290 ↗
  • MalfunctionMar 23, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13868265 ↗
  • InjuryMar 11, 2022Low Readings; Low Readings

    A CALLER REPORTED ON BEHALF OF CUSTOMER LOW READINGS WITH THE ADC DEVICE. THE CALLER REPORTED UNSPECIFIED BLOOD GLUCOSE RESULTS WITH THE ADC METER, THAT WERE LOWER THAN CUSTOMER FELT. THE CALLER REPORTED CUSTOMER EXPERIENCED EITHER LOSS OF CONSCIOUSNESS OR SEIZURE DUE TO THIS ISSUE, BUT REFUSED TO PROVIDE ANY FURTHER INFORMATION. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.

    Patient problems: Convulsion/Seizure; Loss of consciousness

    MAUDE report 13741701 ↗
  • MalfunctionMar 3, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13659381 ↗
  • MalfunctionFeb 16, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 412 MG/DL AND 132 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13542192 ↗
  • MalfunctionFeb 4, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13449790 ↗
  • MalfunctionFeb 2, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 90 MG/DL, 45 MG/DL, 37 MG/DL AND 123 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: Insufficient Information

    MAUDE report 13423125 ↗
  • MalfunctionJan 24, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 450 MG/DL AND 128 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13338412 ↗
  • InjuryJan 10, 2022Low Readings; Low Readings

    A CUSTOMER REPORTED A DELIVERY DELAY WITH A REPLACEMENT ADC BLOOD GLUCOSE METER. THE REPLACEMENT DEVICE WAS ISSUED AS CUSTOMER HAD REPORTED LOW GLUCOSE RESULTS. DUE TO DELIVERY DELAY, CUSTOMER REPORTED BEING HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE AS SHE HAD NO METHOD TO MONITOR GLUCOSE. THE CUSTOMER DID NOT PROVIDE TREATMENT INFORMATION BUT DID STATE SHE WAS PRESCRIBED INSULIN AFTER HOSPITALIZATION. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.

    Patient problems: Hyperglycemia

    MAUDE report 13213604 ↗
  • InjuryJan 7, 2022Unable to Obtain Readings; Unable to Obtain Readings

    A CUSTOMER REPORTED AN "ERROR 3" MESSAGE WITH THE ADC BLOOD GLUCOSE METER, WHEN APPLYING BLOOD SAMPLE TO TEST STRIP. THE CUSTOMER WAS THEREFORE UNABLE TO OBTAIN GLUCOSE RESULT, AND EXPERIENCED A LOSS OF CONSCIOUSNESS. THE CUSTOMER REPORTED REQUIRING TREATMENT FROM A HEALTHCARE PROVIDER, BUT DECLINED TO PROVIDE FURTHER DETAILS. PLEASE NOTE THAT UPON TROUBLESHOOTING IT WAS FOUND THAT CUSTOMER WAS APPLYING BLOOD SAMPLE TO INCORRECT POSITION ON TEST STRIP. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.

    Patient problems: Loss of consciousness

    MAUDE report 13195364 ↗
  • MalfunctionDec 28, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 546 MG/DL AND 121 MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13101614 ↗
  • MalfunctionDec 22, 2021Failure to Power Up; Failure to Power Up

    THE REPORTER CONTACTED ABBOTT DIABETES CARE, ALLEGING THE METER WILL NOT TURN ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13058295 ↗
  • InjuryNov 9, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER REPORTED AN UNSPECIFIED READING ISSUE WITH THE ADC BLOOD GLUCOSE METER. THE CUSTOMER REPORTED THAT DUE TO THIS ISSUE, SHE "BLACKED OUT" AND WAS TAKEN TO HOSPITAL BY AMBULANCE WHERE SHE WAS TREATED WITH INSULIN. THERE WAS NO REPORT OF DEATH OR PERMANENT INJURY ASSOCIATED WITH THIS EVENT.

    Patient problems: Hyperglycemia; Loss of consciousness

    MAUDE report 12778928 ↗
  • MalfunctionNov 3, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12742728 ↗
  • MalfunctionOct 14, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 112 MG/DL AND 19 (LO) MG/DL WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: Insufficient Information

    MAUDE report 12634735 ↗
  • MalfunctionSep 30, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER REPORTED RECEIVING ERRONEOUS GLUCOSE RESULTS FROM AN ABBOTT DIABETES CARE DEVICE. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO EITHER THE C, D, OR E ZONE. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12553397 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K092602. One submission often covers many models, so check the product and lot details in each record.

  • Z-0485-2014 · Nov 18, 2013 · Terminated
    FreeStyle Blood Glucose Test Strips; For in vitro diagnostic testing. 50 count UPC 6 99073 12050 2; Freestyle Blood Glucose Test Strips; for Institutional Use only 50 count: UPC 6
  • Z-0486-2014 · Nov 18, 2013 · Terminated
    FreeStyle Lite Blood Glucose Test Strips; For in vitro diagnostic testing. 50 count; 100 count Product of Ireland; UPC 6 99073 70819 9; - 50 count UPC 6 99073 70822 9 - 50 count

Other Glucose dehydrogenase, glucose devices

All Glucose dehydrogenase, glucose devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.