Skip to content

Midas Rex® AA10 (UDI 00721902481895)Recalls & reported problems

Medtronic Powered Surgical Solutions · ATT AA10 LEGEND 10CM ANGL (9ST) 2.4MM · Class 2

All models of Midas Rex (154)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00721902481895), received between Jun 15, 2021 and Aug 18, 2026 — classified as 14 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (7 reports).

9 other recalls cover devices cleared under the same FDA submission (K020069); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 3 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device7
  • No Apparent Adverse Event3
  • Difficult to Insert2
  • Loss of Power2
  • Detachment of Device or Device Component1
  • Naturally Worn1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions13
  • Insufficient Information1

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 18, 2026Loss of Power

    It was reported that during intra-op attachment was frozen. It was reported that there was no patient or staff impact.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50298452 ↗
  • MalfunctionJun 30, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT ATTACHMENT WAS NOT WORKING PROPERLY. THERE WAS NO PATIENT INVOLVED. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE BEARINGS ARE BROKEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25663010 ↗
  • MalfunctionJun 11, 2026Overheating of Device

    IT WAS REPORTED THAT THE ATTACHMENT WAS OVERHEATING. NO PATIENT IMPACT REPORTED. ON FOLLOW-UP IT WAS REPORTED THAT THERE IS NO PATIENT OR STAFF IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25468100 ↗
  • MalfunctionJun 7, 2026Difficult to Insert

    IT WAS REPORTED THAT THE DEVICE WITH THE BUR COULD NOT BE INSERTED. NO PATIENT IMPACT REPORTED. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: Insufficient Information

    MAUDE report 25421979 ↗
  • MalfunctionMay 15, 2026Naturally Worn

    IT WAS REPORTED THAT THE DEVICE WAS BROKEN. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25181745 ↗
  • MalfunctionMar 24, 2025Overheating of Device; Overheating of Device

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. NO PATIENT IMPACT REPORTED. REPAIR REQUEST ESCALATED TO PRODUCT EVENT DUE TO CUSTOMER INDICATION THAT THIS REPORT IS A COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21674800 ↗
  • MalfunctionFeb 13, 2025Overheating of Device; Overheating of Device

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21371267 ↗
  • MalfunctionJan 15, 2025Overheating of Device; No Apparent Adverse Event; Overheating of Device; No Apparent Adverse Event

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. NO PATIENT IMPACT REPORTED. REPAIR REQUEST ESCALATED TO PRODUCT EVENT DUE TO CUSTOMER INDICATION THAT THIS REPORT IS A COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21159792 ↗
  • MalfunctionNov 13, 2024Difficult to Insert; Physical Resistance/Sticking; Difficult to Insert; Physical Resistance/Sticking

    IT WAS REPORTED THAT IT IS NOT POSSIBLE TO INSERT DRILL AND LOCK IT. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20671906 ↗
  • MalfunctionOct 2, 2024Overheating of Device; Overheating of Device

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. REPAIR IS BEING ESCALATED TO PRODUCT EVENT DUE TO REASON FOR THE RETURN AND DUE TO RETURN IN LESS THAN 90 DAYS FROM PURCHASE OR REPAIR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20349518 ↗
  • MalfunctionSep 26, 2024Loss of Power; Loss of Power

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF LOSS OF POWER. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. REPAIR IS BEING ESCALATED TO PRODUCT EVENT DUE TO REASON FOR THE RETURN (CE).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20308634 ↗
  • MalfunctionSep 9, 2024Overheating of Device; Overheating of Device

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF OVERHEATING. NO PATIENT IMPACT REPORTED. REPAIR REQUEST ESCALATED TO PRODUCT EVENT DUE TO CUSTOMER INDICATION THAT THIS REPORT IS A COMPLAINT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20171272 ↗
  • MalfunctionSep 3, 2024Detachment of Device or Device Component; No Apparent Adverse Event; Detachment of Device or Device Component; No Apparent Adverse Event

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF PROBLEM NOT REPORTED. NO PATIENT IMPACT REPORTED. REPAIR ESCALATED TO A PRODUCT EVENT DUE TO EVALUATION FINDINGS OF DETACHED BEARINGS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20128228 ↗
  • MalfunctionJun 15, 2021Overheating of Device; Overheating of Device

    REPAIR REQUEST INITIATED FOR DEVICE WITH THE REPORT OF MOTOR, CABLE AND ATTACHMENT WERE TOO HOT HANDLE AND THE MOTOR HAS LOSING POWER. NO PATIENT IMPACT REPORTED. REPAIR REQUEST ESCALATED TO A PRODUCT EVENT BASED ON REASON FOR RETURN AND DUE TO CUSTOMER INDICATION THAT THIS REPORT IS A COMPLAINT. IT WAS ALSO NOTED THAT THE EM200 PRODUCT WAS RETURNED IN LESS THAN 90 DAYS FROM PURCHASE OR REPAIR. ON FOLLOW-UP, IT WAS CONFIRMED THAT THERE WAS NO PATIENT OR STAFF IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12005416 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K020069. One submission often covers many models, so check the product and lot details in each record.

  • Z-1602-2016 · Apr 8, 2016 · Terminated
    Medtronic Midas Rex Dissecting Tool Legend 10 cm 4mm BA DIAM. Intended for use in neurosurgery.
  • Z-0321-2016 · Oct 21, 2015 · Terminated
    Tool Legend 16cm 2.4 mm wire pass for Midas Rex Legend Dissecting Tool. Used by surgeons to drill holes.
  • Z-0546-2010 · Nov 30, 2009 · Terminated
    Medtronic Midas Rex Legend Footed Attachment, AF01. Intended for use with all Legend and Legend EHS motors for general surgery including maxillofacial , craniofacial and sternotom
  • Z-2156-2008 · May 30, 2008 · Terminated
    Medtronic Midas Rex Legend tapered dissecting tool, Model F2/8TA23 Medtronic Powered Surgical Solutions, Forth Worth, TX Intended for use in Spinal, Cranial, Orthopedic, General S
  • Z-1574-05 · Sep 1, 2005 · Terminated
    Medtronic Midas Rex Classic, GS and Legend dissecting tools manufactured by Medtronic Inc. Powered Surgical Solutions, 4620 North Beach Street, Fort Worth, tX 76137 USA, Rx only, S
  • Z-1067-05 · Jun 29, 2005 · Terminated
    Medtronic MIDAS REX Legend High Speed Pneumatic System, Gold Touch Motor. Manufactured by Medtronic, Inc. Powered Surgical Solutions, 4620 North Beach Street, Fort Worth, TX 76137
  • Z-0278-05 · Nov 11, 2004 · Terminated
    Medtronic Midas Rex Legend High Speed Pneumatic System
  • Z-0213-05 · Nov 3, 2004 · Terminated
    Medtronic Midas Rex Legend Lubricant/ Diffuser Cartridge Catalog number PA100-A, manufactured by Medtronic Powered Surgical Solutions, Fort Worth, Texas
  • Z-0419-03 · Nov 11, 2002 · Terminated
    Legend Lubricant System (Legend Gold, Gold Touch and Platinum motors and Legend Lubricant/Diffuser cartridge)

Other MOTOR, DRILL, PNEUMATIC devices

All MOTOR, DRILL, PNEUMATIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page