MOSAIC® 305C23 (UDI 00721902694745)Recalls & reported problems
Medtronic, Inc. · VALVE 305C23 MOSAIC AOR CINCH US NCE · Class 3 · Implant
Quick summary
This Medtronic Mosaic aortic valve is a tissue heart valve. Its 326 reports describe aortic stenosis, leaking or regurgitation, and higher pressure readings across the valve, and some patients later needed another valve placed inside it. The reports include no deaths and 320 injuries. No recalls name this device. Reports fell to 2 in the last 12 months from 19 in the 12 months before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 327
- Deaths reported
- 0
- Injuries reported
- 321
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 327 adverse event reports that list this exact device (DI 00721902694745), received between Jul 7, 2017 and Sep 16, 2026 — classified as 321 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (242 reports).
3 other recalls cover devices cleared under the same FDA submission (P990064); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information242
- Gradient Increase55
- Device Stenosis27
- Intravalvular regurgitation23
- Adverse Event Without Identified Device or Use Problem15
- Calcified14
- Material Split, Cut or Torn4
- Patient Device Interaction Problem4
Patient problems
- Aortic Valve Stenosis178
- Aortic Valve Insufficiency/ Regurgitation87
- Insufficient Information71
- No Information67
- Aortic Insufficiency36
- Aortic Regurgitation28
- Calcium Deposits/Calcification11
- Stenosis7
Recent adverse event reports
The 25 most recent of 327 reports, in the reporter's own words as published by the FDA.
InjurySep 16, 2026Patient Device Interaction Problem
Medtronic received information that 15 years post implant of this 23mm aortic bioprosthetic valve, it was explanted and replaced with an unknown product. The reason for the replacement was reported as infection. No additional adverse patient effects were reported.
Patient problems: Unspecified Infection
MAUDE report 50788657 ↗InjuryMar 31, 2026Gradient Increase; Intravalvular regurgitation
ADDITIONAL INFORMATION WAS RECEIVED THAT SEVERE AORTIC INSUFFICIENCY WAS NOTED PRIOR TO THE AORTIC VALVE-IN-VALVE REPLACEMENT WITH A TRANSCATHETER VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Calcium Deposits/Calcification; Insufficient Information
MAUDE report 24739290 ↗InjuryNov 24, 2025Insufficient Device Problem Information; Patient Device Interaction Problem; Insufficient Device Problem Information; Patient Device Interaction Problem
IT WAS REPORTED THAT 4 YEARS AND 6 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A NEW 23MM AORTIC BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Endocarditis; Insufficient Information
MAUDE report 23632962 ↗InjurySep 4, 2025Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation; Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation
ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED THE REASON FOR REPLACEMENT AS AORTIC STENOSIS AND AORTIC REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Insufficient Information
MAUDE report 22963201 ↗InjuryAug 18, 2025Gradient Increase; Device Stenosis; Intravalvular regurgitation; Gradient Increase; Device Stenosis; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 17 YEARS AND 7 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS MODERATE AORTIC INSUFFICIENCY, AORTIC STENOSIS AND A MEAN GRADIENT OF 44MMHG. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis
MAUDE report 22834522 ↗InjuryJun 25, 2025Gradient Increase; Device Stenosis; Gradient Increase; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 16 YEARS AND 1 MONTH POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS HIGH GRADIENTS AND AORTIC STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 22330773 ↗InjuryMay 26, 2025Inadequacy of Device Shape and/or Size; Insufficient Device Problem Information; Inadequacy of Device Shape and/or Size; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 13 YEARS AND 10 MONTHS POST IMPLANT OF THIS MOSAIC VALVE IN THE PULMONARY POSITION, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 22083281 ↗MalfunctionMay 21, 2025Calcified; Insufficient Device Problem Information; Calcified; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 15 YEARS AND 7 MONTHS POST IMPLANT OF THIS MOSAIC AORTIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. IT WAS REPORTED THAT DURING EVALUATION, PROTRUDING ANNULAR CALCIFICATION COULD BE SEEN IN THE LEFT VENTRICULAR OUTFLOW TRACT (LVOT) AND UNDER THE LEFT CORONARY CUSP (LCC). NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Calcium Deposits/Calcification
MAUDE report 22063523 ↗InjuryMay 1, 2025Calcified; Device Stenosis; Calcified; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 21 YEARS AND 6 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC STENOSIS. IT WAS ALSO REPORTED THAT CALCIUM COULD BE SEEN IN THE LEFT VENTRICULAR OUTFLOW TRACT (LVOT) UNDER THE LEFT CORONARY CUSP (LCC). AN INTERVENTION IS PLANNED BUT NOT SCHEDULED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis; Calcium Deposits/Calcification
MAUDE report 21941940 ↗InjuryApr 30, 2025Material Split, Cut or Torn; Device Stenosis; Intravalvular regurgitation; Material Split, Cut or Torn; Device Stenosis; Intravalvular regurgitation
IT WAS REPORTED THAT THE PATIENT EXPERIENCED SEVERE AORTIC INSUFFICIENCY AND SEVERE AORTIC STENOSIS DUE TO THE FAILING 23MM MOSAIC AORTIC VALVE. THE VALVE EXHIBITED PERFORMANCE ISSUES, INCLUDING STENOSIS AND A TORN LEAFLET, ATTRIBUTED TO LONG-TERM WEAR AND TEAR AFTER BEING IMPLANTED 16 YEARS AGO. THE PATIENT UNDERWENT A SUCCESSFUL VALVE-IN-VALVE REPLACEMENT WITH A 23FX+ TRANSCATHETER VALVE. THE AORTIC INSUFFICIENCY AND STENOSIS WERE RESOLVED FOLLOWING THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Cusp Tear
MAUDE report 21935138 ↗InjuryApr 17, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT POST IMPLANT OF THIS 23MM MOSAIC TISSUE VALVE, IT WAS EXPLANTED AND REPLACED WITH A NON MEDTRONIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Insufficient Information
MAUDE report 21861549 ↗InjuryMar 6, 2025Gradient Increase; Device Stenosis; Intravalvular regurgitation; Gradient Increase; Device Stenosis; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 16 YEARS AND 5 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AORTIC STENOSIS, MEAN GRADIENT OF 39MMHG AND MILD AORTIC REGURGITATION. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY 1 MONTH LATER, THE PATIENT SUCCESSFULLY HAD AN AORTIC VALVE-IN-VALVE REPLACEMENT WITH A TRANSCATHETER VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis
MAUDE report 21534857 ↗MalfunctionJan 31, 2025Gradient Increase; Gradient Increase
MEDTRONIC RECEIVED INFORMATION THAT 15 YEARS AND 7 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS HIGH GRADIENT. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 21287461 ↗InjuryJan 20, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT POST IMPLANT OF THIS 23MM MOSAIC AORTIC VALVE IMPLANTED IN THE PULMONARY POSITION, A TRANSCATHETER PULMONARY BIOPROSTHETIC VALVE (TPBV) WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 21191234 ↗MalfunctionJan 9, 2025Calcified; Thickening of Material; Device Stenosis; Calcified; Thickening of Material; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 14 YEARS AND 7 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC STENOSIS, CALCIFICATION AND LEAFLET THICKENING. IT WAS ALSO REPORTED THAT THE PATIENT MAY BE TRANSFERRED TO A DIFFERENT FACILITY TO CONDUCT A BASILICA PROCEDURE CONCOMITANTLY WITH THE TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 21121029 ↗MalfunctionJan 8, 2025Device Stenosis; Intravalvular regurgitation; Device Stenosis; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION 13 YEARS AND 10 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE THAT THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS SEVERE AORTIC STENOSIS AND TRACE CENTRAL AORTIC REGURGITATION. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis
MAUDE report 21103672 ↗MalfunctionJan 8, 2025Device Stenosis; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 14 YEARS AND 5 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC STENOSIS (AS). NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 21102902 ↗InjuryDec 9, 2024Incomplete Coaptation; Mechanics Altered; Intravalvular regurgitation; Incomplete Coaptation; Mechanics Altered; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 16 YEARS AND 1 MONTH POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS VALVE DYSFUNCTION WITH POOR COAPTATION DUE TO MODERATE AORTIC REGURGITATION WITH REDUCED LEAFLET MOTION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Insufficient Information
MAUDE report 20885875 ↗InjuryNov 27, 2024Insufficient Device Problem Information; Device Stenosis; Insufficient Device Problem Information; Device Stenosis
ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED SUBSEQUENTLY, 5 MONTHS AND 8 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis; Dyspnea; Fatigue
MAUDE report 20797792 ↗InjuryNov 22, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 13 YEARS AND 9 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY 1 MONTH LATER, THE PATIENT SUCCESSFULLY HAD AN AORTIC VALVE-IN-VALVE REPLACEMENT WITH A TRANSCATHETER VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 20759330 ↗InjuryNov 5, 2024Device Stenosis; Intravalvular regurgitation; Device Stenosis; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 13 YEARS AND 6 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVAL UATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS MILD AORTIC REGURGITATION (AR). NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 2 MONTHS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS MILD AORTIC REGURGITATION (AR) AND AORTIC STENOSIS (AS). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis
MAUDE report 20610250 ↗InjuryNov 4, 2024Device Stenosis; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 1 MONTH POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE OF A DIFFERENT MANUFACTURER WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC STENOSIS (AS). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 20597288 ↗InjuryOct 11, 2024Calcified; Gradient Increase; Device Stenosis; Calcified; Gradient Increase; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 19 YEARS AND 8 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC STENOSIS, A MEAN GRADIENT OF 34MMHG AND A PEAK GRADIENT OF 68MMHG. IT WAS ALSO REPORTED THAT CALCIUM COULD BE SEE NEAR THE RIGHT CORONARY CUSP AND THE LEFT CORONARY CUSP COMMISSURE EXTENDING 23 TO 35MM ABOVE THE BASAL PLANE. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 6 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis; Calcium Deposits/Calcification; Insufficient Information
MAUDE report 20432243 ↗MalfunctionSep 17, 2024Calcified; Insufficient Device Problem Information; Calcified; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 20 YEARS AND 6 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. IT WAS STATED THAT CALCIUM WAS SEEN IN THE SINOTUBULAR JUNCTION (STJ) AND PROXIMAL LEFT CORONARY ARTERY (LCA). IT WAS REPORTED THAT POSSIBLE SMALL SHORT SEGMENT DISSECTIONS WERE SEEN IN THE AORTIC ARCH, ABDOMINAL AORTA AND DESCENDING THORACIC AORTA. IT WAS ALSO STATED THAT PLAQUE/THROMBUS AND CONCENTRIC CALCIUM WERE SEEN IN THE ABDOMINAL AORTA. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Calcium Deposits/Calcification
MAUDE report 20234676 ↗InjurySep 11, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 12 YEARS AND 7 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. IT WAS STATED THAT CALCIUM WAS NOTED BELOW THE LEFT CORONARY CUSP AND NON-CORONARY CUSP EXTENDING INTO THE LEFT VENTRICULAR OUTFLOW TRACT (LVOT). IT WAS REPORTED THAT CALCIUM WAS NOTED IN THE PROXIMAL LEFT CORONARY ARTERY, "CONCENTRIC CALCIFICATION IN LCI AND ABDOMINAL AORTA". NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Calcium Deposits/Calcification
MAUDE report 20190778 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P990064. One submission often covers many models, so check the product and lot details in each record.
- Z-1880-2019 · May 21, 2019 · TerminatedMosaic 310 Bioprosthesis, Mitral, REF 310C29 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
- Z-1881-2019 · May 21, 2019 · TerminatedMosaic 310 Bioprosthesis, Mitral, REF 310C31 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
- Z-1225-2009 · Jan 26, 2009 · TerminatedMedtronic Mosaic Porcine Bioprostheses with Cinch, Heart Valve; Model 305; Catalog #s 305C21AA, 305C2701, 305C2301, 305C2501, and 305C2901; Sterile LC; Rx only; Medtronic Inc 710 M
Other heart-valve, non-allograft tissue devices
- INSPIRIS RESILIA AORTIC VALVE 11500AEdwards Lifesciences LLC · Heart-Valve, Non-Allograft Tissue2 deaths404 reports
- INSPIRIS RESILIA AORTIC VALVE 11500AEdwards Lifesciences LLC · Heart-Valve, Non-Allograft Tissue385 reports
- PERCEVAL PVS23Corcym Canada Corp · heart-valve, non-allograft tissue26 deaths337 reports
- MOSAIC® 305C25Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths324 reports
- FREESTYLE® FR995-27Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths318 reports
- FREESTYLE® FR995-29Medtronic, Inc. · heart-valve, non-allograft tissue288 reports
- FREESTYLE® FR99525Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths280 reports
- MOSAIC ULTRA® 305U23Medtronic, Inc. · heart-valve, non-allograft tissue239 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.