MOSAIC® 305C25 (UDI 00721902694752)Recalls & reported problems

Medtronic, Inc. · VALVE 305C25 MOSAIC AOR CINCH US NCE · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic Mosaic aortic heart valve has 325 reports, with 320 injuries and 2 deaths. Many describe patients needing a second valve placed inside the first, often because the valve narrowed or leaked backward. No recalls name this device. Reports fell to 5 in the last 12 months from 17 in the 12 months before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
324
Deaths reported
2
Injuries reported
320
Recalls naming it
0

Summary

The FDA's MAUDE database contains 324 adverse event reports that list this exact device (DI 00721902694752), received between Jul 11, 2017 and May 27, 2026 — classified as 2 deaths, 320 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (238 reports).

3 other recalls cover devices cleared under the same FDA submission (P990064); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 5 reportsDec 2023: 1 reportsJan 2024: 3 reportsFeb 2024: 2 reportsMar 2024: 5 reportsApr 2024: 3 reportsMay 2024: 1 reportsJun 2024: 2 reportsJul 2024: 4 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 5 reportsNov 2024: 5 reportsDec 2024: 4 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information238
  • Gradient Increase56
  • Device Stenosis27
  • Intravalvular regurgitation26
  • Adverse Event Without Identified Device or Use Problem13
  • Calcified7
  • Incomplete Coaptation7
  • Patient Device Interaction Problem7

Patient problems

  • Aortic Valve Stenosis160
  • Aortic Valve Insufficiency/ Regurgitation90
  • Insufficient Information69
  • No Information48
  • Aortic Insufficiency39
  • Aortic Regurgitation25
  • Calcium Deposits/Calcification11
  • Non specific EKG/ECG Changes9

Recent adverse event reports

The 25 most recent of 324 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 27, 2026Gradient Increase

    IT WAS REPORTED THAT 16 YEARS AND 8 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. IT WAS REPORTED THAT AN INTERVENTION IS REQUIRED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 6 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS HIGH GRADIENTS OF 20MMHG. THE PATIENT ALSO HAD SYMPTOMS OF SHORTNESS OF BREATH. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Dyspnea; Insufficient Information

    MAUDE report 25306919 ↗
  • InjuryMar 30, 2026Intravalvular regurgitation

    IT WAS REPORTED THAT 13 YEARS AND 1 MONTH POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS REGURGITATION. EVIDENCE OF CALCIFICATION WAS SEEN IN THE PROTRUDING RIGHT COMMON ILIAC (RCI) ARTERY.  IT WAS REPORTED THAT AN INTERVENTION WAS REQUIRED AND SCHEDULED FOR (B)(6) 2026. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Calcium Deposits/Calcification

    MAUDE report 24726041 ↗
  • InjuryFeb 5, 2026Patient Device Interaction Problem

    ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED THE REASON FOR PERMANENT PACEMAKER IMPLANT AS COMPLETE HEART BLOCK. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Heart Block; Non specific EKG/ECG Changes

    MAUDE report 24266977 ↗
  • InjuryNov 27, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT 18 YEARS AND 5 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR PERMANENT PACEMAKER WAS REPORTED AS COMPLETE HEART BLOCK (CHB). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Heart Block

    MAUDE report 23665360 ↗
  • InjurySep 3, 2025Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation; Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation

    ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE REASON FOR INTERVENTION WAS MODERATE - SEVERE AORTIC STENOSIS (AS) AND MODERATE - SEVERE AORTIC REGURGITATION (AR). NO ADDITIONAL ADVERSE PATIENT'S EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Calcium Deposits/Calcification

    MAUDE report 22956459 ↗
  • InjuryAug 28, 2025Insufficient Device Problem Information; Device Stenosis; Insufficient Device Problem Information; Device Stenosis

    MEDTRONIC RECEIVED INFORMATION THAT 17 YEARS AND 3 MONTHS POST IMPLANT OF THIS MOSAIC VALVE IN THE PULMONARY POSITION, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information; Pulmonary Valve Stenosis

    MAUDE report 22913116 ↗
  • InjuryJun 3, 2025Device Stenosis; Intravalvular regurgitation; Device Stenosis; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 17 YEARS AND 4 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVAL UATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS MODERATE AORTIC REGURGITATION AND AORTIC STENOSIS. IT WAS ALSO REPORTED THAT ON EVALUATION, CALCIUM COULD BE SEEN IN THE PROXIMAL LEFT CORONARY ARTERY AND RIGHT CORONARY ARTERY, AS WELL AS THE SINOTUBULAR JUNCTION. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 3 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis

    MAUDE report 22136427 ↗
  • InjuryMay 6, 2025Device Stenosis; Intravalvular regurgitation; Device Stenosis; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 6 YEARS AND 6 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, ANOTHER MANUFACTURER'S TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC STENOSIS AND MODERATE AORTIC REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis

    MAUDE report 21970129 ↗
  • InjuryApr 3, 2025Intravalvular regurgitation; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 14 YEARS AND 10 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVA LUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS SEVERE AORTIC REGURGITATION. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY 21 DAYS LATER, THE PATIENT SUCCESSFULLY HAD AN AORTIC VALVE-IN-VALVE REPLACEMENT WITH A TRANSCATHETER VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 21761958 ↗
  • MalfunctionMar 21, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT AFTER AN UNKNOWN DURATION POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 21662026 ↗
  • InjuryFeb 14, 2025Intravalvular regurgitation; Intravalvular regurgitation

    IT WAS REPORTED THAT A PATIENT IMPLANTED WITH A 25 MM MOSAIC AORTIC VALVE WAS EXPERIENCING CENTRAL AORTIC REGURGITATION. THE TREATMENT PROVIDED WAS A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 21383981 ↗
  • InjuryJan 7, 2025Device Stenosis; Intravalvular regurgitation; Device Stenosis; Intravalvular regurgitation

    IT WAS REPORTED THAT A PATIENT IMPLANTED WITH A 25MM MOSAIC AORTIC SURGICAL VALVE HAD SEVERE SYMPTOMATIC AORTIC STENOSIS (AS). THE PLANNED TREATMENT FOR THE CONDITION IS A TRANSCATHETER AORTIC VALVE REPLACEMENT. NO PROCEDURAL DATE HAS BEEN SCHEDULED YET. THE PATIENT OUTCOME WAS REPORTED AS ALIVE WITH NO INJURY.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis

    MAUDE report 21090679 ↗
  • InjuryDec 12, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT 1 YEAR AND 12 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A DEFIBRILLATOR WAS IMPLANTED. THE REASON FOR THE DEFIBRILLATOR IMPLANT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 20914660 ↗
  • InjuryDec 12, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 4 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A DEFIBRILLATOR WAS IMPLANTED. THE REASON FOR THE DEFIBRILLATOR IMPLANT WAS REPORTED AS SICK SINUS SYNDROME (SSS)/ SINUS NODE DYSFUNCTION. IT WAS STATED THAT THE DEFIBRILLATOR IMPLANT WAS FOR PRIMARY PREVENTION AND THE PATIENT WAS NEVER SHOCKED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Non specific EKG/ECG Changes

    MAUDE report 20914497 ↗
  • InjuryDec 9, 2024Gradient Increase; Perivalvular Leak; Device Stenosis; Intravalvular regurgitation; Gradient Increase; Perivalvular Leak; Device Stenosis; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT APPROXIMATELY 15 YEARS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS A MEAN GRADIENT OF 38MMHG, MILD TO MODERATE AORTIC REGURGITATION (AR), MODERATE PARAVALVULAR LEAK (PVL) AND AORTIC STENOSIS (AS). IT WAS ALSO REPORTED THAT ANOTHER MANUFACTURERS PLUG PRODUCT WAS USED TO TREAT THE PVL IN ACCORDANCE WITH THE TRANSCATHETER VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Insufficient Information; Valvular Insufficiency/ Regurgitation

    MAUDE report 20881241 ↗
  • MalfunctionDec 8, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT A PATIENT WAS BEING EVALUATED FOR A TRANSCATHETER AORTIC VALVE REPLACEMENT PROCEDURE INVOLVING A VALVE MOSAIC AORTIC CINCH 25MM. THE PATIENT WAS UNDERGOING EVALUATION FOR THIS PROCEDURE, BUT NO SPECIFIC SYMPTOMS OR ADVERSE OUTCOMES WERE DETAILED. THE DEVICE WAS NOTED TO BE IMPLANTED AND REMAINS IN SERVICE. THERE WERE NO REPORTED PRODUCT ISSUES OR PROCEDURAL DELAYS, AND THE CAUSE OR CONTRIBUTING FACTORS TO THE EVENT WERE NOT SPECIFIED. IT WAS STATED THAT A POSSIBLE LEFT ATRIAL APPENDAGE (LAA) THROMBUS WAS PRESENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Thrombosis/Thrombus

    MAUDE report 20875379 ↗
  • InjuryNov 26, 2024Patient Device Interaction Problem; Patient Device Interaction Problem

    MEDTRONIC RECEIVED INFORMATION THAT 15 YEARS AND 3 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. IT WAS STATED THAT THERE WAS A POSSIBLE PSEUDOANEURYSM OR ANEURYSM OBSERVED IN THE RIGHT CORONARY CUSP (RCC) AND RCC/LEFT CORONARY CUSP (LCC). IT WAS ALSO STATED THAT THERE WAS CALCIFICATION SEEN IN THE PROXIMAL LEFT CORONARY ARTERY AND RIGHT CORONARY ARTERY. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aneurysm; Calcium Deposits/Calcification

    MAUDE report 20777713 ↗
  • InjuryNov 18, 2024Device Stenosis; Intravalvular regurgitation; Device Stenosis; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 14 YEARS AND 11 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC REGURGITATION AND SEVERE AORTIC STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis

    MAUDE report 20710261 ↗
  • InjuryNov 14, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 14 YEARS AND 8 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 27 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 20685373 ↗
  • InjuryNov 7, 2024Calcified; Device Stenosis; Intravalvular regurgitation; Calcified; Device Stenosis; Intravalvular regurgitation

    ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED SUBSEQUENTLY, 1 MONTH AND 9 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Calcium Deposits/Calcification

    MAUDE report 20630756 ↗
  • InjuryNov 1, 2024Gradient Increase; Device Stenosis; Gradient Increase; Device Stenosis

    MEDTRONIC RECEIVED INFORMATION THAT 16 YEARS AND 12 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC STENOSIS AND A GRADIENT OF 42MMHG. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis

    MAUDE report 20587571 ↗
  • InjuryOct 31, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 3 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 25MM AORTIC BIOPROSTHETIC VALVE OF THE SAME MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 20578820 ↗
  • InjuryOct 30, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT 6 WEEKS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INT ERVENTION WAS REPORTED AS "FAILED SURGICAL AORTIC VALVE". NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Calcium Deposits/Calcification; Thrombosis/Thrombus

    MAUDE report 20565396 ↗
  • InjuryOct 15, 2024Gradient Increase; Intravalvular regurgitation; Gradient Increase; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 20 YEARS AND 6 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS HIGH GRADIENTS. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 19 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC INSUFFICIENCY. IT WAS ALSO REPORTED THAT DUE TO THE AORTIC INSUFFICIENCY, IT WAS DIFFICULT FOR THE SURGEON TO IMPLANT THE TRANSCATHETER VALVE WITHOUT MOVEMENT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 20457462 ↗
  • InjuryOct 15, 2024Calcified; Intravalvular regurgitation; Calcified; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 16 YEARS AND 6 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. IT WAS STATED THAT TORTUOSITY WAS SEEN IN ILIAC ARTERIES. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 6 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC REGURGITATION. IT WAS ALSO REPORTED THAT DEGENERATIVE CALCIFICATION WAS PRESENT WHICH RESTRICTED THE MOVEMENT OF 2 OF THE LEAFLETS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation

    MAUDE report 20457199 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P990064. One submission often covers many models, so check the product and lot details in each record.

  • Z-1881-2019 · May 21, 2019 · Terminated
    Mosaic 310 Bioprosthesis, Mitral, REF 310C31 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1880-2019 · May 21, 2019 · Terminated
    Mosaic 310 Bioprosthesis, Mitral, REF 310C29 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1225-2009 · Jan 26, 2009 · Terminated
    Medtronic Mosaic Porcine Bioprostheses with Cinch, Heart Valve; Model 305; Catalog #s 305C21AA, 305C2701, 305C2301, 305C2501, and 305C2901; Sterile LC; Rx only; Medtronic Inc 710 M

Other heart-valve, non-allograft tissue devices

All heart-valve, non-allograft tissue devices →