Amsco Evolution HC-1200 (REF UE05)Recalls & reported problems
Steris Mexico, S. de R.L. de C.V. · The Amsco Evolution - Medium Steam Sterilizer. Amsco Evolution Prevacuum Steam Sterilizers are configured for Prevacuum sterilization of heat and moisture-stable materials used in healthcare facilities. These units can be equipped with Prevacuum, Steam Flush Pressure Pulse (SFPP), gravity, liquid, and daily air removal test cycles. · Class 2
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00724995151508), received between Feb 3, 2022 and Aug 5, 2026 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Leak/Splash” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Leak/Splash3
- Detachment of Device or Device Component2
- Fire1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Bruise/Contusion1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 5, 2026Leak/Splash
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50146981 ↗MalfunctionJul 9, 2025Leak/Splash; Leak/Splash
THE USER FACILITY REPORTED THAT THEIR AMSCO EVOLUTION STERILIZER WAS LEAKING WATER AND STEAM ONTO THE FLOOR. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22458129 ↗MalfunctionJun 28, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE USER FACILITY REPORTED THAT THE TOP PANEL OF THEIR 54" EVOLUTION STERILIZER FELL OFF AND CONTACTED AN EMPLOYEE'S HEAD. NO MEDICAL TREATMENT WAS NOT SOUGHT OR ADMINISTERED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17221613 ↗MalfunctionApr 19, 2023Fire; Fire
A STERIS SERVICE TECHNICIAN WAS PERFORMING A SERVICE ACTIVITY ON THE USER FACILITY'S EVOLUTION STERILIZER WHEN A BURNING ODOR WAS DETECTED AND SMALL FLAME IN THE WIRING OF THE CONTROL BOX WAS OBSERVED. THE TECHNICIAN PUT OUT THE FLAME. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16773156 ↗MalfunctionFeb 14, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE USER FACILITY REPORTED THAT THE TOP PANEL TO THEIR 54" EVOLUTION STERILIZER FELL OFF AFTER AN EMPLOYEE CLOSED THE STERILIZER'S DOOR. AS THE PANEL FELL, THE EMPLOYEE TRIED TO CATCH IT AND IT CONTACTED THEIR NOSE. THE EMPLOYEE ALSO SUSTAINED AN INJURY TO THEIR SHOULDER. MEDICAL TREATMENT WAS SOUGHT AND ADMINISTERED (ICE PACK).
Patient problems: Bruise/Contusion
MAUDE report 16375385 ↗MalfunctionFeb 3, 2022Leak/Splash; Leak/Splash
THE USER FACILITY REPORTED THAT WATER AND STEAM WERE LEAKING FROM THEIR EVOLUTION STERILIZER ONTO THE FLOOR. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13440439 ↗
Other Sterilizer, steam devices
- NA P2131-2Fort Defiance Industries, LLC · Sterilizer, Steam1 recall0 reports
- STATCLAVE G4 Chamber Autoclave G4-621103Scican Ltd · Sterilizer, Steam1 recall0 reports
- STATCLAVE G4 Chamber Autoclave G4-622303Scican Ltd · Sterilizer, Steam1 recall0 reports
- NA AX083020003Steris Mexico, S. de R.L. de C.V. · Sterilizer, steam23 reports
- AMSCO AMSCO 400 Steam SterilizerSteris Mexico, S. de R.L. de C.V. · Sterilizer, steam11 reports
- Atlas AX0660230830000013Steris Mexico, S. de R.L. de C.V. · Sterilizer, steam9 reports
- AMSCO AMSCO 400 Steam SterilizerSteris Mexico, S. de R.L. de C.V. · Sterilizer, steam9 reports
- AMSCO AMSCO 600 Steam SterilizerSteris Mexico, S. de R.L. de C.V. · Sterilizer, steam8 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.