Skip to content

Amsco Evolution HC-1500 (REF UE06)Recalls & reported problems

Steris Mexico, S. de R.L. de C.V. · The Amsco Evolution - Medium Steam Sterilizer. Amsco Evolution Prevacuum Steam Sterilizers are configured for Prevacuum sterilization of heat and moisture-stable materials used in healthcare facilities. These units can be equipped with Prevacuum, Steam Flush Pressure Pulse (SFPP), gravity, liquid, and daily air removal test cycles. · Class 2

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00724995151522), received between Feb 17, 2022 and Feb 25, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component5
  • Leak/Splash1

Patient problems

  • Insufficient Information2
  • Abrasion1
  • Bruise/Contusion1
  • Burn(s)1
  • Pain1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 25, 2026Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT THE UPPER PANEL OF THEIR AMSCO EVOLUTION STERILIZER DETACHED AND CONTACTED THE EMPLOYEE. THE USER FACILITY DID NOT DISCLOSE FURTHER DETAILS REGARDING THE INJURY OR IF MEDICAL TREATMENT WAS SOUGHT OR ADMINISTERED.

    Patient problems: Insufficient Information

    MAUDE report 24446872 ↗
  • MalfunctionJul 29, 2025Leak/Splash; Leak/Splash

    THE USER FACILITY REPORTED THAT FOLLOWING A DART TEST CYCLE AN EMPLOYEE OPENED THE DOOR TO THEIR AMSCO EVOLUTION STERILIZER AND OBTAINED A BURN. IT IS UNKNOWN IF MEDICAL TREATMENT WAS SOUGHT OR ADMINISTERED.

    Patient problems: Burn(s)

    MAUDE report 22643188 ↗
  • MalfunctionFeb 22, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT WHILE OPENING THE DOOR ON THEIR MEDIUM EVOLUTION STERILIZER, THE TOP PANEL ON THE STERILIZER FELL AND CONTACTED THE EMPLOYEE. THE EMPLOYEE RECEIVED MEDICAL TREATMENT, BUT IT WAS NOT DISCLOSED WHAT TYPE OF MEDICAL TREATMENT WAS ADMINISTERED.

    Patient problems: Bruise/Contusion

    MAUDE report 21441120 ↗
  • MalfunctionNov 13, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT THE TOP PANEL ON THEIR 60" EVOLUTION STERILIZER FELL OFF THE UNIT CONTACTING AN EMPLOYEE'S HEAD. MEDICAL TREATMENT WAS SOUGHT AND ADMINISTERED. THE USER FACILITY DID NOT DISCLOSE WHAT MEDICAL TREATMENT WAS ADMINISTERED.

    Patient problems: Insufficient Information

    MAUDE report 18125399 ↗
  • MalfunctionDec 21, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT THE TOP PANEL OF THEIR 66" EVOLUTION STERILIZER FELL OFF AND CONTACTED AN EMPLOYEE'S HEAD. IT IS UNKNOWN IF THE EMPLOYEE SOUGHT OR RECEIVED MEDICAL TREATMENT.

    Patient problems: Pain

    MAUDE report 16028605 ↗
  • MalfunctionFeb 17, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT THE TOP PANEL ON THEIR EVOLUTION STERILIZER FELL OFF OF THE UNIT CONTACTING AN EMPLOYEE'S HEAD. MEDICAL TREATMENT WAS ADMINISTERED (SKIN ADHESIVE).

    Patient problems: Abrasion

    MAUDE report 13552322 ↗

Other Sterilizer, steam devices

All Sterilizer, steam devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.