Padlock Clip C910001 (UDI 00724995184247)Recalls & reported problems

US Endoscopy · The Padlock Clip defect closure system is used in flexible endoscopy for the compression of tissue in the gastrointestinal tract, for hemostasis or for treating lesions of the wall of the gastrointestinal organs. · Class 2 · Implant

All models of Padlock Clip (3)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This US Endoscopy Padlock Clip is a metal clip used in flexible endoscopy to close tissue in the digestive tract, stop bleeding, or treat lesions. The 46 reports mostly describe the clip failing to deploy or fire when the trigger was pulled, with a few describing clips found already deployed or removed while partly released, sometimes delaying the procedure. No deaths and no injuries are reported. There are no recalls naming this device. Reports fell from 28 in the prior 12 months to 4 in the most recent 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
46
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 46 adverse event reports that list this exact device (DI 00724995184247), received between Mar 26, 2020 and Apr 23, 2026 — classified as 46 malfunctions.

The most frequently reported specific device problem is “Failure to Fire” (36 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 1 reportsMay 2025: 3 reportsJun 2025: 18 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202618
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Fire36
  • Activation, Positioning or Separation Problem3
  • Failure to Advance3
  • Failure to Eject2
  • Insufficient Device Problem Information2
  • Loss of or Failure to Bond2
  • Material Protrusion/Extrusion2
  • Positioning Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions42
  • Insufficient Information2
  • Blood Loss1
  • No Known Impact Or Consequence To Patient1
  • Obstruction/Occlusion1
  • Ulcer1

Recent adverse event reports

The 25 most recent of 46 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 23, 2026Failure to Fire

    THE USER FACILITY REPORTED DURING A PATIENT PROCEDURE THE PADLOCK CLIP STANDARD DID NOT DEPLOY THE CLIP. A SECOND PADLOCK CLIP STANDARD WAS USED TO COMPLETE THE PROCEDURE RESULTING IN A PROCEDURE DELAY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24965016 ↗
  • MalfunctionFeb 26, 2026Premature Activation

    THE USER FACILITY REPORTED VIA VIGILANCE REPORT THAT THEIR PADLOCK CLIP STANDARD WAS ALREADY DEPLOYED INTO THE SAFETY CAP BEFORE USE. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24454649 ↗
  • MalfunctionFeb 20, 2026Failure to Fire

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THEIR PADLOCK CLIP DEFECT CLOSURE SYSTEM WAS REMOVED WITH THE CLIP PARTIALLY DEPLOYED. ANOTHER DEVICE WAS UTILIZED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24400829 ↗
  • MalfunctionNov 18, 2025Failure to Advance; Failure to Advance

    THE USER FACILITY REPORTED VIA VIGILANCE REPORT THAT DURING A PATIENT PROCEDURE THEIR PADLOCK CLIP STANDARD FAILED TO DEPLOY AFTER THE CLIP WAS USED TO TREAT A BLEEDING ULCER. THE CLIP WAS REMOVED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: Ulcer

    MAUDE report 23585889 ↗
  • MalfunctionJul 24, 2025Failure to Fire; Failure to Fire

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THE PADLOCK CLIP DEFECT CLOSURE SYSTEM DID NOT DEPLOY THE CLIP. A SECOND PADLOCK CLIP WAS USED TO COMPLETE THE PROCEDURE RESULTING IN A PROCEDURE DELAY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22591360 ↗
  • MalfunctionJun 27, 2025Failure to Fire; Failure to Fire

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THE PADLOCK CLIP DEFECT CLOSURE SYSTEM DID NOT DEPLOY THE CLIP. THE PROCEDURE WAS COMPLETED WITH VACUUM THERAPY. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22344879 ↗
  • MalfunctionJun 27, 2025Failure to Fire; Failure to Fire

    THE USER FACILITY REPORTED THAT TWO PADLOCK CLIP DEFECT CLOSURE SYSTEMS DID NOT DEPLOY THE CLIP AND A CRACKING SOUND WAS NOTICED WITHIN THE HANDLE. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22344568 ↗
  • MalfunctionJun 27, 2025Failure to Fire; Failure to Fire

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THE PADLOCK CLIP DEFECT CLOSURE SYSTEM DID NOT DEPLOY THE CLIP. A SECOND PADLOCK CLIP WAS USED TO COMPLETE THE PROCEDURE. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22344559 ↗
  • MalfunctionJun 26, 2025Failure to Fire; Failure to Fire

    THE DISTRIBUTOR REPORTED THAT DURING A PATIENT PROCEDURE THE PADLOCK CLIP DEFECT CLOSURE SYSTEM DID NOT DEPLOY THE CLIP. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22342900 ↗
  • MalfunctionJun 26, 2025Failure to Fire; Failure to Fire

    THE DISTRIBUTOR REPORTED THAT DURING A PATIENT PROCEDURE THE PADLOCK CLIP DEFECT CLOSURE SYSTEM DID NOT DEPLOY THE CLIP. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22342801 ↗
  • MalfunctionJun 25, 2025Failure to Fire; Failure to Fire

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THE PADLOCK CLIP DEFECT CLOSURE SYSTEM DID NOT DEPLOY THE CLIP. A SECOND PADLOCK CLIP WAS USED TO COMPLETE THE PROCEDURE. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22333401 ↗
  • MalfunctionJun 25, 2025Failure to Fire; Failure to Fire

    THE DISTRIBUTOR REPORTED THAT DURING A PATIENT PROCEDURE THE PADLOCK CLIP DEFECT CLOSURE SYSTEM DID NOT DEPLOY THE CLIP. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22333400 ↗
  • MalfunctionJun 24, 2025Failure to Fire; Failure to Fire

    THE DISTRIBUTOR REPORTED THAT DURING A PATIENT PROCEDURE THE PADLOCK CLIP DEFECT CLOSURE SYSTEM DID NOT DEPLOY THE CLIP. THE PROCEDURE WAS COMPLETED WITH THE PLACEMENT OF CLIPS. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22321149 ↗
  • MalfunctionJun 24, 2025Failure to Fire; Failure to Fire

    THE DISTRIBUTOR REPORTED THAT DURING A PATIENT PROCEDURE THE PADLOCK CLIP DEFECT CLOSURE SYSTEM DID NOT DEPLOY THE CLIP. THE PROCEDURE WAS COMPLETED WITH THE PLACEMENT OF CLIPS. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22321032 ↗
  • MalfunctionJun 24, 2025Failure to Fire; Failure to Fire

    THE DISTRIBUTOR REPORTED THAT THE PADLOCK CLIP DEFECT CLOSURE SYSTEM DID NOT DEPLOY THE CLIP. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22320954 ↗
  • MalfunctionJun 24, 2025Failure to Fire; Failure to Fire

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THE PADLOCK CLIP DEFECT CLOSURE SYSTEM DID NOT DEPLOY THE CLIP. A SECOND PADLOCK CLIP WAS USED TO COMPLETE THE PROCEDURE. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22320798 ↗
  • MalfunctionJun 24, 2025Failure to Fire; Failure to Fire

    THE USER FACILITY REPORTED THAT THEIR PADLOCK CLIP DEFECT CLOSURE SYSTEM DID NOT DEPLOY THE CLIP. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22320712 ↗
  • MalfunctionJun 24, 2025Failure to Fire; Failure to Fire

    THE DISTRIBUTOR REPORTED THAT THEIR PADLOCK CLIP DEFECT CLOSURE SYSTEM WAS UNABLE TO DEPLOY THE CLIP DUE TO A DAMAGED HANDLE. NO REPORT OF INJURY OF PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22320578 ↗
  • MalfunctionJun 24, 2025Failure to Fire; Failure to Fire

    THE DISTRIBUTOR REPORTED THAT THEIR PADLOCK CLIP DEFECT CLOSURE SYSTEM WAS UNABLE TO DEPLOY THE CLIP DUE TO A DAMAGED HANDLE. NO REPORT OF INJURY OF PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22317795 ↗
  • MalfunctionJun 24, 2025Failure to Fire; Failure to Fire

    THE DISTRIBUTOR REPORTED THAT THEIR PADLOCK CLIP DEFECT CLOSURE SYSTEM WAS UNABLE TO DEPLOY THE CLIP DUE TO A DAMAGED HANDLE. NO REPORT OF INJURY OF PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22317787 ↗
  • MalfunctionJun 20, 2025Failure to Fire; Failure to Fire

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THE PADLOCK CLIP DEFECT CLOSURE SYSTEM DID NOT DEPLOY THE CLIP. A SECOND PADLOCK CLIP WAS USED TO COMPLETE THE PROCEDURE. NO REPORT OF INJURY OR PROCEDURE DELAY

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22297667 ↗
  • MalfunctionJun 19, 2025Failure to Fire; Failure to Fire

    THE USER FACILITY REPORTED THAT THE PADLOCK CLIP DEFECT CLOSURE SYSTEM WAS UNABLE TO DEPLOY THE CLIP DUE TO A BROKEN HANDLE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A NEW DEVICE. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22293594 ↗
  • MalfunctionJun 18, 2025Failure to Fire; Failure to Fire

    THE USER FACILITY REPORTED THAT THE PADLOCK CLIP DEFECT CLOSURE SYSTEM WAS HARD TO PUSH AND DID NOT DEPLOY THE CLIP. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22286318 ↗
  • MalfunctionMay 16, 2025Failure to Fire; Failure to Fire

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THEIR PADLOCK CLIP WOULD NOT DEPLOY RESULTING IN A PROCEDURE DELAY. NO INJURY WAS REPORTED; HOWEVER, THE OUTCOME OF THE PROCEDURE WAS NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22035094 ↗
  • MalfunctionMay 16, 2025Failure to Fire; Failure to Fire

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THEIR PADLOCK CLIP WOULD NOT DEPLOY RESULTING IN A PROCEDURE DELAY. NO INJURY WAS REPORTED; HOWEVER, THE OUTCOME OF THE PROCEDURE WAS NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22035011 ↗

Other Hemostatic metal clip for the GI tract devices

All Hemostatic metal clip for the GI tract devices →