Padlock Clip Pro-Select C913131 (UDI 00724995184254)Recalls & reported problems

US Endoscopy · The Padlock Clip Pro-select is used in flexible endoscopy for the compression of tissue in the gastrointestinal tract, for hemostasis or for treating lesions of the wall of the gastrointestinal organs. · Class 2 · Implant

All models of Padlock Clip Pro-Select (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
14
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00724995184254), received between Oct 16, 2020 and Jul 1, 2025 — classified as 1 injury, 13 malfunctions.

The most frequently reported specific device problem is “Failure to Fire” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 4 reportsJul 2025: 4 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Fire11
  • Detachment of Device or Device Component3
  • Difficult to Remove2
  • Activation, Positioning or Separation Problem1
  • Failure to Advance1
  • Positioning Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions12
  • Perforation2

Recent adverse event reports

The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 1, 2025Failure to Fire; Failure to Fire

    THE DISTRIBUTOR REPORTED THAT DURING A PATIENT PROCEDURE TWO PADLOCK CLIP PRO-SELECT DEFECT CLOSURE SYSTEMS DID NOT DEPLOY THE CLIP. THE PROCEDURE WAS COMPLETED WITH TTS CLIPS. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22375142 ↗
  • MalfunctionJul 1, 2025Failure to Fire; Failure to Fire

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THE PADLOCK CLIP PRO-SELECT DEFECT CLOSURE SYSTEM DID NOT DEPLOY THE CLIP. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22375122 ↗
  • MalfunctionJul 1, 2025Failure to Fire; Failure to Fire

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE TWO PADLOCK CLIP PRO-SELECT DEFECT CLOSURE SYSTEM DID NOT DEPLOY. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22375121 ↗
  • MalfunctionJul 1, 2025Failure to Fire; Failure to Fire

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THE PADLOCK CLIP PRO-SELECT DEFECT CLOSURE SYSTEM DID NOT DEPLOY THE CLIP. A SECOND PADLOCK CLIP WAS USED TO COMPLETE THE PROCEDURE. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22375119 ↗
  • MalfunctionJun 26, 2025Failure to Fire; Failure to Fire

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THE PADLOCK CLIP PRO-SELECT DEFECT CLOSURE SYSTEM DID NOT DEPLOY THE CLIP. A SECOND PADLOCK CLIP WAS USED TO COMPLETE THE PROCEDURE. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22342868 ↗
  • MalfunctionJun 25, 2025Failure to Fire; Failure to Fire

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THE PADLOCK CLIP PRO-SELECT DEFECT CLOSURE SYSTEM DID NOT DEPLOY THE CLIP. A SECOND PADLOCK CLIP PRO-SELECT WAS USED TO COMPLETE THE PROCEDURE. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22328822 ↗
  • MalfunctionJun 24, 2025Failure to Fire; Failure to Fire

    THE USER FACILITY REPORTED THAT THE PADLOCK CLIP PRO-SELECT DEFECT CLOSURE SYSTEM WAS UNABLE TO DEPLOY THE CLIP. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A DIFFERENT DEVICE. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22321245 ↗
  • MalfunctionJun 24, 2025Failure to Fire; Failure to Fire

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THE PADLOCK CLIP PRO-SELECT DEFECT CLOSURE SYSTEM DID NOT DEPLOY THE CLIP. A SECOND PADLOCK CLIP WAS USED TO COMPLETE THE PROCEDURE. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22321233 ↗
  • MalfunctionApr 30, 2025Failure to Fire; Failure to Fire

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THE PADLOCK CLIP WOULD NOT DEPLOY. ADDITIONALLY, USER FACILITY PERSONNEL USED A PADLOCK CIP PRO SELECT WHICH WOULD NOT DEPLOY RESULTING IN THE CANCELLATION OF THE PROCEDURE. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21937979 ↗
  • InjuryFeb 4, 2025Difficult to Remove; Failure to Advance; Detachment of Device or Device Component; Difficult to Remove; Failure to Advance; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THE HOUSING OF THEIR PADLOCK CLIP PRO-SELECT CAME OFF THE SCOPE. THE PROCEDURE WAS DELAYED AS THE DOCTOR ATTEMPTED TO CUT THE CATHETER OF THE PADLOCK AND REMOVE THE PADLOCK CHAMBER FROM THE PATIENT WITH AN OVERTUBE. THE OUTER DIAMETER OF THE PADLOCK WAS TOO LARGE TO GO THROUGH THE OVERTUBE. THE DOCTOR REMOVED THE OVERTUBE AND THEN ATTEMPTED TO REMOVE THE PADLOCK FROM THE PATIENT AGAIN, WHICH CAUSED A PERFORATION. THE PROCEDURE WAS DELAYED AS THE PATIENT WAS TRANSFERRED TO A HIGHER LEVEL OF CARE FOR THE REMOVAL OF THE DEVICE AND TREATMENT OF THE PERFORATION.

    Patient problems: Perforation

    MAUDE report 21306805 ↗
  • MalfunctionJun 26, 2024Failure to Fire; Activation, Positioning or Separation Problem; Positioning Problem; Failure to Fire; Activation, Positioning or Separation Problem; Positioning Problem

    THE USER FACILITY REPORTED THAT DURING A PROCEDURE THE PADLOCK CLIP PRO-SELECT DID NOT FULLY DEPLOY INTO THE TISSUE. THE CLIP WAS RELEASED INTO THE LUMEN OF THE PATIENT AND WAS THEN RETRIEVED WITHOUT ISSUE. THERE WAS NO REPORT OF INJURY, AND THE PROCEDURE WAS COMPLETED WITH A SECOND PADLOCK CLIP PRO-SELECT DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19617712 ↗
  • MalfunctionJul 3, 2023Failure to Fire; Failure to Fire

    THE DISTRIBUTOR REPORTED THAT DURING A PATIENT PROCEDURE THE PADLOCK CLIP PRO-SELECT DEFECT CLOSURE SYSTEM DID NOT DEPLOY. FACILITY PERSONNEL UTILIZED A DIFFERENT CLIP TO COMPLETE THE PROCEDURE RESULTING IN AN APPROXIMATELY 20 MINUTE DELAY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY, NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17246448 ↗
  • MalfunctionJun 15, 2023Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component

    THE USER FACILITY REPORTED VIA MEDWATCH REPORT (B)(4) THAT DURING A PATIENT PROCEDURE WHILE WITHDRAWING THEIR SCOPE AFTER PADLOCK CLIP DEPLOYMENT, THE CAP DETACHED. THE CAP WAS REMOVED FROM THE PATIENT. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17138031 ↗
  • MalfunctionOct 16, 2020Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT THE HOUSING COMPONENT OF A PADLOCK CLIP PRO-SELECT DEFECT CLOSURE DEVICE BECAME DETACHED FROM THE ENDOSCOPE FOLLOWING USE TO CLOSE A DEFECT IN THE CECUM. THE ENDOSCOPE WAS RE-INSERTED TO ATTEMPT RECOVERY OF THE COMPONENT, AND A PERFORATION WAS OBSERVED IN THE SIGMOID. THE PATIENT WAS SENT TO SURGERY TO REPAIR THE PERFORATION AND RETRIEVE THE DETACHED COMPONENT. THE PATIENT IS REPORTED TO BE STABLE AND RECOVERING FROM SURGERY.

    Patient problems: Perforation

    MAUDE report 10692699 ↗

Other Hemostatic metal clip for the GI tract devices

All Hemostatic metal clip for the GI tract devices →