Skip to content

STERIS 4095 Surigical Table (REF ST013911002)Recalls & reported problems

Steris Corporation · The STERIS 4095 General Surgical Table with SS Featherweight Leg Section Velcro and 2-inch TLT PADS is a hydraulic and electromechanically driven surgical table intended to position and support the full patient body during diagnostic examinations and or surgical procedures. · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
19
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 00724995195236), received between May 24, 2023 and Aug 13, 2026 — classified as 1 injury, 18 malfunctions.

The most frequently reported specific device problem is “Unintended Movement” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 3 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unintended Movement8
  • Use of Device Problem4
  • Device Tipped Over2
  • Connection Problem1
  • Detachment of Device or Device Component1
  • Key or Button Unresponsive/not Working1
  • Leak/Splash1
  • Positioning Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions15
  • Pain3
  • Fall1

Recent adverse event reports

The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 13, 2026Connection Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50236872 ↗
  • MalfunctionAug 7, 2026Use of Device Problem

    Patient problems: Pain

    MAUDE report 50177378 ↗
  • InjuryAug 5, 2026Unintended Movement

    The user facility reported during a patient procedure with their STERIS 4095 Surgical Table that the patient slid towards the head section and the table ¿tilted slightly towards Trendelenburg position¿. The patient was repositioned on the table, and the procedure was completed successfully without delay. No report of injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50147483 ↗
  • MalfunctionApr 29, 2026Device Tipped Over

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THEIR 4095 SURGICAL TABLE TIPPED OVER. THE PATIENT WAS TRANSFERRED TO A DIFFERENT TABLE, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY FOLLOWING A SHORT DELAY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25034664 ↗
  • MalfunctionApr 21, 2026Unintended Movement

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THE FOOT END TO THEIR 4095 SURGICAL TABLE HAD LIFTED OFF THE FLOOR. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24945623 ↗
  • MalfunctionMar 18, 2026Unintended Movement

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE, THE HEAD SECTION OF THEIR STERIS 4095 SURGICAL TABLE BEGAN TO TILT DOWN WITHOUT BEING COMMANDED TO DO SO. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24632259 ↗
  • MalfunctionJan 8, 2026Unintended Movement

    THE USER FACILITY REPORTED VIA MEDWATCH REPORT #(B)(4) THAT A LOUD NOISE WAS HEARD DURING PATIENT TRANSFER FROM AN ICU BED. THE BOTTOM TWO "PIECES" OF THE STERIS 4095 SURGICAL TABLE FELL TO THE FLOOR CAUSING THE PATIENT'S LEGS TO DROP. AN EMPLOYEE HELD THE PATIENT'S LEGS AND TRANSFERRED THEM BACK TO THE ICU BED. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24004104 ↗
  • MalfunctionDec 18, 2025Leak/Splash; Leak/Splash

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THEIR STERIS 4095 SURGICAL TABLE WAS LEAKING OIL. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT DELAY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23838003 ↗
  • MalfunctionOct 31, 2025Unintended Movement; Unintended Movement

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE, THEIR 4095 SURGICAL TABLE UNLOCKED AND LOWERED IN HEIGHT WITHOUT BEING COMMANDED TO DO SO. THE PATIENT WAS TRANSFERRED TO ANOTHER TABLE, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: Pain

    MAUDE report 23438807 ↗
  • MalfunctionJul 25, 2025Unintended Movement; Unintended Movement

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THEIR 4095 SURGICAL TABLE WENT INTO TRENDELENBURG POSITION AND TILTED TO THE LEFT WITHOUT BEING COMMANDED TO DO SO. A "CRACKING NOISE" WAS HEARD AND USER FACILITY PERSONNEL STATED THE PATIENT "FELL" FROM THE TABLE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY, AND NO PROCEDURE DELAY OR INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22612380 ↗
  • MalfunctionJul 17, 2025Use of Device Problem; Use of Device Problem

    THE USER FACILITY REPORTED THAT FOLLOWING A PATIENT PROCEDURE, THEY WERE TRANSFERRING A PATIENT FROM THEIR 4095 SURGICAL TABLE ONTO A STRETCHER AND WHEN WEIGHT WAS APPLIED TO THE TABLE'S HEAD SECTION IT "GAVE WAY" AND FELL. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22529507 ↗
  • MalfunctionNov 15, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT THE LEG SECTION DETACHED FROM THEIR 4095 SURGICAL TABLE DURING A PATIENT PROCEDURE. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20695065 ↗
  • MalfunctionNov 15, 2024Positioning Problem; Positioning Problem

    THE USER FACILITY REPORTED DURING A PATIENT PROCEDURE, THE LEG SECTION OF THE 4095 SURGICAL TABLE MOVED DOWNWARDS WITHOUT BEING COMMANDED TO DO SO. THERE WAS NO REPORT OF INJURY OR PROCEDURE DELAY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20694502 ↗
  • MalfunctionAug 29, 2024Use of Device Problem; Use of Device Problem

    THE USER FACILITY REPORTED THAT DURING AN "ABDOMINAL SURGERY" WITHOUT ANY CHEST RESTRAINT TO THE PATIENT, THEIR 4095 SURGICAL TABLE PAD BEGAN TO SLIDE TOWARD THE LOWERED HEAD SECTION WHILE THE TABLE WAS IN TRENDELENBURG POSITION. USER FACILITY PERSONNEL LEVELED THE TABLE AND MOVED THE PATIENT TO THEIR ORIGINAL POSITION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20098210 ↗
  • MalfunctionAug 9, 2024Unintended Movement; Unintended Movement

    THE USER FACILITY REPORTED DURING A PATIENT PROCEDURE, THE LEG SECTION OF THE 4095 SURGICAL TABLE MOVED DOWNWARDS WITHOUT BEING COMMANDED TO DO SO. THERE WAS NO REPORT OF INJURY OR PROCEDURE DELAY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19949519 ↗
  • MalfunctionMar 8, 2024Use of Device Problem; Use of Device Problem

    THE USER FACILITY REPORTED THAT AT THE CONCLUSION OF A PATIENT PROCEDURE, THE PATIENT WAS REPOSITIONING THEMSELVES TO GET OFF THE 4095 SURGICAL TABLE AND THE LEG SECTION DETACHED, SUBSEQUENTLY CAUSING THE PATIENT TO OBTAIN AN INJURY TO THE BACK OF THEIR LEG.

    Patient problems: Pain

    MAUDE report 18866934 ↗
  • MalfunctionAug 16, 2023Unintended Movement; Unintended Movement

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE, USER FACILITY PERSONNEL COMMANDED THE 4095 SURGICAL TABLE TO LOWER IN HEIGHT VIA THE HAND CONTROL. WHEN THE "HEIGHT DOWN" BUTTON ON THE HAND CONTROL WAS PRESSED, THE SURGICAL TABLE APPEARED TO GO INTO REVERSE TRENDELENBURG POSITION. USER FACILITY PERSONNEL PROCEEDED TO PRESS THE "LEVEL" BUTTON AND THE TABLE DID NOT RESPOND TO COMMANDS AND PROCEEDED TO LOWER IN HEIGHT FOLLOWED BY A LOUD "BANG NOISE". THE PATIENT PROCEDURE WAS COMPLETED SUCCESSFULLY, AND NO INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17561528 ↗
  • MalfunctionAug 2, 2023Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working

    THE USER FACILITY REPORTED THAT THEIR 4095 SURGICAL TABLE WOULD NOT RESPOND TO HAND CONTROL COMMANDS DURING A PATIENT PROCEDURE. THE PROCEDURE WAS COMPLETED WITH THE TABLE AS POSITIONED. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17445406 ↗
  • MalfunctionMay 24, 2023Device Tipped Over; Device Tipped Over

    THE USER FACILITY REPORTED THAT PRIOR TO THE START OF THE PROCEDURE A PATIENT TRANSFERRED THEMSELVES TO THE TABLE CAUSING THEIR 4095 SURGICAL TABLE TO TIP DOWNWARD HITTING THE FLOOR. THE PATIENT FELL FROM THE TABLE TO THE FLOOR RESULTING IN BACK PAIN. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: Fall

    MAUDE report 16993949 ↗

Other TABLE, OPERATING-ROOM, AC-POWERED devices

All TABLE, OPERATING-ROOM, AC-POWERED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page