ORT200 TableRecalls & reported problems
IMRIS Imaging, Inc. · ORT200 50/60Hz Covered Table Assy V2, 114148-000 · Class 1
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00857534006349), received between Feb 3, 2022 and Oct 24, 2025 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Unintended Movement” (11 reports).
2 FDA recall records name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Aug 6, 2026Class IIOpen, Classified
The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the patient if unintended rotational movement occurs during a procedure.
Z-3033-2026 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
- Aug 25, 2021Class IITerminated
ORT200/300 rotational lock may fail to function such that it cannot change states between locked and unlocked, due to a mechanical hardware issue.
Z-2584-2021 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Unintended Movement11
- Difficult or Delayed Positioning3
- Failure to Align1
Patient problems
- No Clinical Signs, Symptoms or Conditions12
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 24, 2025Difficult or Delayed Positioning; Failure to Align; Difficult or Delayed Positioning; Failure to Align
DURING THE ONE(1) YEAR PREVENTATIVE MAINTENANCE (PM), THE MANUFACTURER'S FIELD SERVICE ENGINEER REPORTED THE ROTATIONAL LOCK ON THE ORT200 OPERATING ROOM TABLE WAS STRUGGLING TO OBTAIN THE LOCK STATUS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23383734 ↗MalfunctionSep 5, 2024Difficult or Delayed Positioning; Unintended Movement; Difficult or Delayed Positioning; Unintended Movement
THE CUSTOMER REPORTED THE ORT200 OPERATING ROOM TABLE'S ROTATIONAL LOCKING MECHANISM WAS INTERMITTENTLY FAILING TO LOCK TABLE ROTATION IN PLACE OR TO UNLOCK TO ALLOW TABLE ROTATION. NO PATIENT INVOLVEMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20148313 ↗MalfunctionMay 23, 2024Unintended Movement; Unintended Movement
THE CUSTOMER REPORTED THE ROTATIONAL LOCK OF THE ORT200 FAILED TO LOCK TABLE MOVEMENT IN PLACE AFTER BEING UNLOCKED AND ROTATED. THIS OCCURRED AFTER TERMINAL CLEANING AFTER A CASE AND IN PREPARATION FOR REMOVING THE TABLE WITH THE TABLE CART. NO PATIENT WAS INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19388844 ↗MalfunctionAug 18, 2023Difficult or Delayed Positioning; Unintended Movement; Difficult or Delayed Positioning; Unintended Movement
THE CUSTOMER REPORTED THAT DURING SET-UP OF A NON-IMAGING CASE, THE OPERATING ROOM TABLE'S ROTATIONAL LOCKING MECHANISM FAILED TO UNLOCK TABLE MOVEMENT UNTIL SEVERAL ATTEMPTS WERE MADE. AFTER THE TABLE UNLOCKED ROTATIONAL POSITION AND THE TABLE WAS ROTATED TO THE DESIRED POSITION, THE MECHANISM SUBSEQUENTLY FAILED TO LOCK TABLE MOVEMENT UNTIL AFTER SEVERAL ATTEMPTS WERE MADE. THE CUSTOMER REPORTED THERE WAS A PATIENT ON THE TABLE DURING THE EVENT, AND THAT NO HARM OR IMPACT TO THE SURGICAL PROCEDURE OCCURRED AS A RESULT OF THE INTERMITTENT MECHANISM FUNCTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17577709 ↗MalfunctionApr 20, 2023Unintended Movement; Unintended Movement
THE CUSTOMER REPORTED THAT DURING A MORNING QUALITY ASSURANCE / FUNCTIONAL CHECK, THE OPERATING ROOM TABLE'S ROTATIONAL LOCKING MECHANISM DID FUNCTION PROPERLY TO LOCK TABLE POSITION IN PLACE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16778977 ↗MalfunctionDec 21, 2022Unintended Movement; Unintended Movement
IT WAS REPORTED THAT DURING A QUALITY ASSURANCE/FUNCTIONAL CHECK, THE ORT200 ROTATIONAL LOCKING MECHANISM WOULD NOT LOCK THE TABLE POSITION INTO PLACE. NO PATIENT WAS INVOLVED AT THE TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16031045 ↗MalfunctionFeb 7, 2022Unintended Movement; Unintended Movement
DURING A SURGICAL PROCEDURE ON (B)(6) 2019, THE ORT200 TABLE DRIFTED DOWNWARD IN THE TRENDELENBURG FUNCTION WHILE THE PATIENT WAS ON THE TABLE. THERE WERE NO INJURIES REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13476415 ↗MalfunctionFeb 7, 2022Unintended Movement; Unintended Movement
DURING A CLEAR POINT PROCEDURE ON (B)(6), 2018, THE CUSTOMER REPORTED THAT THE ORT200 TABLE DRIFTED IN THE TRENDELENBURG POSITION. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13475732 ↗MalfunctionFeb 4, 2022Unintended Movement; Unintended Movement
DURING A 14-HOUR ABLATION PROCEDURE ON (B)(6) 2021, THE CUSTOMER OBSERVED THAT THE ORT200 TABLE EXHIBITED SOME DRIFT.THERE WAS NO PATIENT HARM OR USER INJURY REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13451188 ↗MalfunctionFeb 3, 2022Unintended Movement; Unintended Movement
DURING A MONTERIS PROCEDURE ON (B)(6) 2020, THE ORT200 TABLE DRIFTED APPROXIMATELY 5MM IN THE TILT FUNCTION. THERE WAS NO PATIENT HARM REPORTED DUE TO THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13446251 ↗MalfunctionFeb 3, 2022Unintended Movement; Unintended Movement
ON (B)(6) 2019, THE CUSTOMER REPORTED THAT THE ORT200 TABLE ROTATION WAS NOT LOCKING OR UNLOCKING.THE EVENT WAS OBSERVED DURING DAILY QA TESTING AND THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13441687 ↗MalfunctionFeb 3, 2022Unintended Movement; Unintended Movement
ON (B)(6) 2019, THE CUSTOMER REPORTED THAT THE ROTATIONAL LOCK ON THE ORT200 FAILED TO LOCK.THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13441384 ↗
FDA clearance and classification
The FDA regulates product code FQO (table, operating-room, ac-powered) under 21 CFR 878.4960 as a Class I device. 21 CFR 878.4960 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the ORT200 Table
Has the ORT200 Table been recalled?
Yes. 2 FDA recall records name this device's identifier: 1 still open and 1 closed (terminated) by the FDA. The most recent, Z-3033-2026 (Class II), began on Aug 6, 2026: The linear actuator and rotational lock mechanism components may experience a mechanical failure that prevents the system from transitioning between locked and unlocked state. This may pose a risk of injury to the… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the ORT200 Table?
The FDA's MAUDE database lists 12 adverse event reports for this device, received between Feb 3, 2022 and Oct 24, 2025. In those reports, the most common specific device problems are “Unintended Movement” (11), “Difficult or Delayed Positioning” (3) and “Failure to Align” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ORT200 Table?
None of the reports are classified as deaths, 0 as injuries and 12 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ORT200 Table FDA approved?
Devices of this type (product code FQO) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by IMRIS Imaging, Inc.
Is the ORT200 Table MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the ORT200 Table, and is it still sold?
The FDA's device database lists IMRIS Imaging, Inc. as the company responsible for it (brand ORT200 Table, model ORT200). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Table, Operating-Room, Ac-Powered devices
- ORT300 TableIMRIS Imaging, Inc. · Table, Operating-Room, Ac-Powered2 recalls21 reports
- MEERA mobile operating table 720001F2MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED2 recalls0 reports
- HarmonyAIR A-Series Suspension SystemSteris Corporation · TABLE, OPERATING-ROOM, AC-POWERED2 recalls0 reports
- HarmonyAIR A-Series Suspension SystemSteris Corporation · TABLE, OPERATING-ROOM, AC-POWERED2 recalls0 reports
- HarmonyAIR A-Series Suspension SystemSteris Corporation · TABLE, OPERATING-ROOM, AC-POWERED2 recalls0 reports
- MEERA mobile operating table 720001B2MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED1 recall19 reports
- MAGNUS operating table system 118001B2MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED1 recall17 reports
- MEERA mobile operating table 720001B0MAQUET GmbH · TABLE, OPERATING-ROOM, AC-POWERED1 recall11 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page