STERIS 5095 Surigical Table (REF ST015211002)Recalls & reported problems

Steris Corporation · The STERIS 5095 General Surgical Table, SS Featherweight Leg Section Velcro 2-inch TLT Pads is a flexible design gives OR staff the versatility they need to perform a wide range of surgeries using a variety of patient positions. · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
17
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 00724995195649), received between Jun 2, 2022 and Jul 31, 2026 — classified as 1 injury, 16 malfunctions.

The most frequently reported specific device problem is “Unintended Movement” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 2 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unintended Movement6
  • Adverse Event Without Identified Device or Use Problem2
  • Leak/Splash2
  • Detachment of Device or Device Component1
  • Device Handling Problem1
  • Failure to Align1
  • Improper or Incorrect Procedure or Method1
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions12
  • Abrasion2
  • Discomfort2
  • Pain1

Recent adverse event reports

The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 31, 2026Unintended Movement

    The user facility reported during a patient procedure while their STERIS 5095 Surgical Table was in Reverse Trendelenburg position that the seat section began to move "downward" without being commanded to do so. User facility personnel utilized the hand control to raise the seat section, and the procedure was completed successfully without delay. No report of injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50098267 ↗
  • MalfunctionJul 9, 2026Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT AN EMPLOYEE OBTAINED A CUT TO THEIR HAND WHILE CLEANING THE BASE OF THEIR 5095 SURGICAL TABLE. NO MEDICAL TREATMENT WAS SOUGHT OR ADMINISTERED.

    Patient problems: Abrasion

    MAUDE report 25771030 ↗
  • MalfunctionApr 17, 2026Unintended Movement

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THEIR 5095 SURGICAL TABLE BEGAN TO TIP AND A PATIENT BEGAN TO SLIDE WHILE THE TABLE WAS BEING PUT IN TRENDELENBURG POSITION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND THE PATIENT DID NOT FALL OFF THE TABLE. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24914496 ↗
  • MalfunctionFeb 19, 2026Leak/Splash

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE, THEIR 5095 SURGICAL TABLE WOULD NOT HOLD THE COMMANDED POSITION. USER FACILITY PERSONNEL STATED THAT FLUID WAS ALSO LEAKING FROM THE BASE OF THE TABLE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY AND NO INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24390979 ↗
  • MalfunctionDec 22, 2025Unintended Movement; Unintended Movement

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE, THEIR 5095 SURGICAL TABLE WAS IN REVERSE TRENDELENBURG POSITION, WHEN THE FOOT END OF THE TABLE "ABRUPTLY DROPPED DOWN" CAUSING THE HEAD SECTION TO RISE UPWARDS. USER FACILITY PERSONNEL STATED A "LOUD NOISE" WAS HEARD DURING THE MOVEMENT. USER FACILITY PERSONNEL ELECTED TO CANCEL THE PROCEDURE.

    Patient problems: Pain

    MAUDE report 23863112 ↗
  • MalfunctionDec 11, 2025Unintended Movement; Unintended Movement

    THE USER FACILITY REPORTED THAT WHILE TRANSFERRING A PATIENT TO A BED, THEIR 5095 SURGICAL TABLE BEGAN TO SLIDE. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23783892 ↗
  • MalfunctionJul 16, 2025Unintended Movement; Unintended Movement

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE A PATIENT SLID TOWARDS THE HEAD SECTION TO THEIR 5095 SURGICAL TABLE WHILE THE TABLE WAS IN THE TRENDELENBURG POSITION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22525024 ↗
  • MalfunctionMay 21, 2025Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    THE USER FACILITY REPORTED THAT WHILE A PATIENT WAS POSITIONED ON THEIR 5095 SURGICAL TABLE, THE DUAL ARTICULATING HEADREST "COLLAPSED" SUBSEQUENTLY CAUSING THE PATIENT TO SUSTAIN AN INJURY. USER FACILITY PERSONNEL ELECTED TO CANCEL THE PROCEDURE.

    Patient problems: Abrasion

    MAUDE report 22062143 ↗
  • MalfunctionApr 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT THE FLOOR LOCKS TO THEIR 5095 SURGICAL TABLE WERE NOT OPERATING PROPERLY. THE PATIENT WAS TRANSFERRED TO A DIFFERENT SURGICAL TABLE AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY FOLLOWING A SHORT DELAY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21778670 ↗
  • MalfunctionFeb 7, 2025Failure to Align; Failure to Align

    THE USER FACILITY REPORTED THE FLOOR LOCKS TO THEIR 5095 SURGICAL TABLE WOULD NOT ENGAGE PROPERLY. USER FACILITY PERSONNEL HAD TO MANUALLY LOCK THE TABLE, RESULTING IN A SHORT DELAY IN A PATIENT PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21336784 ↗
  • MalfunctionJan 8, 2025Unintended Movement; Unintended Movement

    THE USER FACILITY REPORTED DURING A PATIENT PROCEDURE THAT THEIR 5095 SURGICAL TABLE SLID DOWN WITHOUT BEING COMMANDED TO DO SO. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21103928 ↗
  • MalfunctionJul 25, 2024Leak/Splash; Leak/Splash

    THE USER FACILITY REPORTED THAT AT THE CONCLUSION OF A PATIENT PROCEDURE IT WAS OBSERVED THAT HYDRAULIC FLUID WAS LEAKING FROM THEIR 5095 SURGICAL TABLE. THE PATIENT PROCEDURE WAS COMPLETED SUCCESSFULLY HOWEVER, THE FOLLOWING PROCEDURE WAS DELAYED. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19829375 ↗
  • MalfunctionMay 23, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THE DOCTOR RAISED THE LEG SECTION OF THEIR 5095 SURGICAL TABLE CAUSING THE E-Z LIFT BEACH CHAIR TO FALL OFF THE TABLE AND ONTO THE FLOOR. THE PATIENT WAS NOT INJURED. A PROCEDURE DELAY OCCURRED AS THE PATIENT WAS REPOSITIONED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19386168 ↗
  • MalfunctionFeb 23, 2024Use of Device Problem; Use of Device Problem

    THE USER FACILITY REPORTED THAT PRIOR TO THE START OF A PATIENT PROCEDURE THE FLOOR LOCKS ON (B)(4) OF THEIR (B)(4) SURGICAL TABLES WOULD NOT DISENGAGE FROM THE FLOOR AND IN ADDITION (B)(4) OF THE TABLES WAS LEAKING OIL NEAR THE LEG OF THE TABLE. A 15-MINUTE DELAY OCCURRED AS A RESULT OF THE EVENT. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18768382 ↗
  • MalfunctionSep 18, 2023Mechanical Problem; Insufficient Device Problem Information; Mechanical Problem; Insufficient Device Problem Information

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE, THEIR 5095 SURGICAL TABLE "STOPPED WORKING". NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17769492 ↗
  • MalfunctionAug 18, 2022Device Handling Problem; Device Handling Problem

    THE USER FACILITY REPORTED THAT AN EMPLOYEE HAD GRABBED A HEADREST FOR THE 5095 SURGICAL TABLE FROM THEIR STORAGE AREA AND WHILE TRANSPORTING THE HEADREST THE EMPLOYEE'S FINGERS WERE PINCHED. THE EMPLOYEE RECEIVED X-RAYS AND ICE WAS APPLIED.

    Patient problems: Discomfort

    MAUDE report 15254624 ↗
  • MalfunctionJun 2, 2022Physical Resistance/Sticking; Physical Resistance/Sticking

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE, AN EMPLOYEE OBTAINED AN INJURY TO THEIR BACK WHILE ATTEMPTING TO RE-POSITION ONE OF THEIR 5095 SURGICAL TABLES THAT WAS STICKING TO THE FLOOR. NO REPORT OF PROCEDURE DELAY. THE USER FACILITY DID NOT DISCLOSE IF MEDICAL TREATMENT WAS SOUGHT OR ADMINISTERED.

    Patient problems: Discomfort

    MAUDE report 14573588 ↗

FDA clearance and classification

The FDA regulates product code FQO (table, operating-room, ac-powered) under 21 CFR 878.4960 as a Class I device. 21 CFR 878.4960 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the STERIS 5095 Surigical Table

Has the STERIS 5095 Surigical Table been recalled?

No FDA recall record in our data names this exact device identifier (DI 00724995195649), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the STERIS 5095 Surigical Table?

The FDA's MAUDE database lists 17 adverse event reports for this device, received between Jun 2, 2022 and Jul 31, 2026. In those reports, the most common specific device problems are “Unintended Movement” (6), “Leak/Splash” (2) and “Detachment of Device or Device Component” (1); the most common patient problems are “Abrasion” (2), “Discomfort” (2) and “Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the STERIS 5095 Surigical Table?

None of the reports are classified as deaths, 1 as an injury and 16 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the STERIS 5095 Surigical Table FDA approved?

Devices of this type (product code FQO) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Steris Corporation.

Who makes the STERIS 5095 Surigical Table, and is it still sold?

The FDA's device database lists Steris Corporation as the company responsible for it (brand STERIS, model 5095 Surigical Table). Its record shows it as in commercial distribution.