Welch Allyn, Inc. 901049 (REF CP150A-1ENB)Recalls & reported problems

Welch Allyn, Inc. · Welch Allyn CP 150 12-Lead Resting Electrocardiograph with Automatic ECG Interpretation, AHA ECG Cable with Disposable Electrodes, CardioPerfect Workstation Connectivity via USB and Ethernet, Color Touchscreen Display, 100-Test Memory; 100–240 V, 50/60 Hz AC, IEC Plug Type-B · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00732094177633), received between Oct 11, 2024 and Oct 11, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 24, 2025Class IIOpen, Classified

    There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.

    Z-1560-2025 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • Electric Shock1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 11, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT A CAREGIVER RECEIVED A SHOCK DUE TO A DAMAGED POWER CORD ON THE CP150 ELECTROCARDIOGRAPH DEVICE. NO MEDICAL INTERVENTION WAS REQUIRED, THE CAREGIVER WAS PLACED UNDER OBSERVATION AND WAS NOTED TO BE WELL AND RETURNED TO WORK. THIS INCIDENT WAS CAPTURED UNDER (B)(4).

    Patient problems: Electric Shock

    MAUDE report 20426082 ↗