Welch Allyn, Inc. 901127 (REF RSCRIBE-6AC-DXCBB)Recalls & reported problems
Welch Allyn, Inc. · RScribe 12-Lead Multi-Channel Electrocardiograph - version 6, current software CD; WAM Wireless Acquisition Module; Adult and pediatric interpretation; Software kit; International power cord; No printer; Suction electrodes (6) and clamps (4); IEC 10 wire, 12-lead lead set with banana plugs; Advanced connectivity-includes DICOM order and output and XML order and XML/PDF output · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Oct 11, 2024Class IIOpen, Classified
There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.
Z-0322-2025 · Root cause: Software design · Recall details · FDA record ↗
Other Electrocardiograph devices
- ELI 380 ELI380-DCS11Mortara Instrument, Inc. · Electrocardiograph1 recall1 death70 reports
- Cardiac Workstation Cardiac Workstation 7000Philips Medizin Systeme Böblingen GmbH · Electrocardiograph2 recalls17 reports
- Q-Stress QSERV-6AA-XXXXBMortara Instrument, Inc. · Electrocardiograph2 recalls1 report
- Cardiac Workstation Cardiac Workstation 5000Philips Medizin Systeme Böblingen GmbH · Electrocardiograph2 recalls1 report
- MAC VU360GE Medical Systems Information Technologies, Inc. · Electrocardiograph2 recalls1 report
- XScribe Cardiopulmonary Kit 41000-030-60Mortara Instrument, Inc. · Electrocardiograph2 recalls0 reports
- Welch Allyn, Inc. 901049Welch Allyn, Inc. · ELECTROCARDIOGRAPH2 recalls0 reports
- Welch Allyn, Inc. 901133Welch Allyn, Inc. · Electrocardiograph2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.