Skip to content

Welch Allyn, Inc. 901144 (REF QS-6EC-ACAAB)Recalls & reported problems

Welch Allyn, Inc. · Q-Stress Cardiac Stress- version 6 with adult and pediatric interpretation; includes current software CD and user manual CD; AM12Q Wired Acquisition Module with Trigger Module for TTL/analog signal output; Software kit; US/Canada power cord; Z200+ thermal printer, (includes 2 starter packs of full- grid paper - 8 in.), Q-Stress premium transport cart with base, keyboard tray, storage shelf, large storage drawer, and side storage pocket; AHA 10 wire, 12-lead lead set with 43 in. leadwires with p · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Oct 11, 2024Class IIOpen, Classified

    There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.

    Z-0321-2025 · Root cause: Software design · Recall details · FDA record ↗

  2. Dec 8, 2023Class IIOpen, Classified

    Potential distortion identified in electrocardiogram (ECG) readings when the Source Consistency Filter (SCF) is enabled.

    Z-0705-2024 · Root cause: Software design · Recall details · FDA record ↗

Other Electrocardiograph devices

All Electrocardiograph devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page