Welch Allyn, Inc. · XScribe Cardiac Stress - version 6 with adult and pediatric interpretation; includes current software CD and User Manual CD; WAM Wireless Acquisition Module with Trigger Module for TTL/analog signal output; Turnkey system with configured CPU and color LCD 24" monitor with 1920 x 1080 resolution; Australia power cord; Z200+ thermal printer, (includes starter pack of full-grid paper - A4), XScribe premium transport cart with base, keyboard tray, storage shelf, large storage drawer, and side stora · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00732094302974), received between May 8, 2023 and May 8, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Patient Data Problem” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Oct 11, 2024Class IIOpen, Classified
There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.
Z-0325-2025 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Patient Data Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 8, 2023Patient Data Problem; Patient Data Problem
IT WAS REPORTED THAT THREE EXAMS HAD THE SAME PATIENT DETAILS EVEN THOUGH THESE EXAMS WERE DONE FOR DIFFERENT PATIENTS. THIS INCIDENT WAS CAPTURED UNDER HILLROM COMPLAINT REF # (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16886662 ↗
Other Electrocardiograph devices
- ELI 380 ELI380-DCS11Mortara Instrument, Inc. · Electrocardiograph1 recall1 death70 reports
- Cardiac Workstation Cardiac Workstation 7000Philips Medizin Systeme Böblingen GmbH · Electrocardiograph2 recalls17 reports
- Q-Stress QSERV-6AA-XXXXBMortara Instrument, Inc. · Electrocardiograph2 recalls1 report
- Cardiac Workstation Cardiac Workstation 5000Philips Medizin Systeme Böblingen GmbH · Electrocardiograph2 recalls1 report
- MAC VU360GE Medical Systems Information Technologies, Inc. · Electrocardiograph2 recalls1 report
- XScribe Cardiopulmonary Kit 41000-030-60Mortara Instrument, Inc. · Electrocardiograph2 recalls0 reports
- Welch Allyn, Inc. 901049Welch Allyn, Inc. · ELECTROCARDIOGRAPH2 recalls0 reports
- Welch Allyn, Inc. 901133Welch Allyn, Inc. · Electrocardiograph2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page