DeRoyal 81-080416Recalls & reported problems
Deroyal Industries, Inc. · Foley Catheter with Temperature Sensor · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00749756045766), received between Oct 17, 2019 and Sep 21, 2026 — classified as 1 injury, 10 malfunctions.
The most frequently reported specific device problem is “Infusion or Flow Problem” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Infusion or Flow Problem5
- Suction Problem5
- Device Dislodged or Dislocated2
- Display or Visual Feedback Problem2
- Blocked Connection1
- Burst Container or Vessel1
- Compatibility Problem1
- Decrease in Suction1
Patient problems
- Unspecified Kidney or Urinary Problem5
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4
- No Clinical Signs, Symptoms or Conditions3
- Urinary Incontinence2
- Abdominal Pain1
- Insufficient Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjurySep 21, 2026Entrapment of Device; Difficult to Remove
A user facility reported, \"Foley got stick in the patient's urethra and had to be removed with lubricant, lidocaine and hydromorphone. Foley balloon was appropriately deflated prior to removal. Post removal it was seen to have a ridge around the balloon area.\"
Patient problems: Insufficient Information
MAUDE report 50846103 ↗MalfunctionApr 27, 2026Blocked Connection
A DISTRIBUTING FACILITY REPORTED, \"ALL THREE PATIENTS WERE HAVING DECREASE IN URINE OUTPUT, DURING ASSESSMENT THE UCT TEAM REVIEWED THE CASES AND FOUND THAT ALL THREE FOLEY CATHETERS WERE BLOCKED. \"
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25003051 ↗MalfunctionJul 7, 2025Device Dislodged or Dislocated; Unintended Deflation; Device Dislodged or Dislocated; Unintended Deflation
A USER FACILITY REPORTED, "THE BALLOON DEFLATED ON ITS OWN AND CAME OUT OF THE PATIENT. THE RN TESTED THE BALLOON, AND IT LOOKED LIKE IT HAD A HOLE IN IT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22411051 ↗MalfunctionAug 7, 2024Burst Container or Vessel; Device Dislodged or Dislocated; Gas/Air Leak; Burst Container or Vessel; Device Dislodged or Dislocated; Gas/Air Leak
CATHETER DISLODGED UPON SURGICAL DRAPE REMOVAL. BALLOON INFLATION PERFORMED WHICH DEMONSTRATED PINHOLE LEAKAGE IN BALLOON, WHICH SUBSEQUENTLY BURST.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19928476 ↗MalfunctionSep 22, 2021Suction Problem; Material Perforation; Infusion or Flow Problem; Suction Problem; Material Perforation; Infusion or Flow Problem
THE PATIENT CAME TO THE RECOVERY ROOM WITH A FOLEY CATHETER DRAINING CLEAR AMBER URINE. WHEN THE BLANKETS WERE BEING STRAIGHTENED BY A HEALTH PROFESSIONAL, CATHETER TUBING WAS LEFT ON THE BED. WHEN THE CATHETER TUBING WAS BEING STRAIGHTENED, THE FOLEY BULB FELL OUT OF THE URETHRA. THE BULB WAS NOT INFLATED AND THE NURSE PROCEEDED TO TRY TO INFLATE. THE BULB INFLATED, HOWEVER, THE NEXT MINUTE IT DID NOT STAY INFLATED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Unspecified Kidney or Urinary Problem
MAUDE report 12514412 ↗MalfunctionJul 2, 2021Suction Problem; Compatibility Problem; Infusion or Flow Problem; Suction Problem; Compatibility Problem; Infusion or Flow Problem
THE PATIENT RETURNED FROM THE OPERATION ROOM WITH ZERO URINE IN THE FOLEY CATHETER BAG AND THE UROMETER WAS DRY AND COLLAPSED WITHOUT SIGNS OF ANY COLLECTION INSIDE. THE OPERATION ROOM REPORT SHOWED ULTRAFILTRATION OF 2 L FLUID REMOVAL. THE NURSE RECEIVING THE CASE DID TROUBLESHOOTING OF THE CATHETER AND IRRIGATED WITH NO RETURN, ALSO A BLADDER SCAN DISPLAYED FOR 350 ML. BALLOON DEFLATED AND CATHETER CLEANED AND ADVANCED FURTHER INTO BLADDER, BALLOON RE-INFLATED. URINE OUTPUT CONTINUED TO BE LOW OVER THE NEXT 12 HOURS, HOWEVER WAS CLEAR WITH NO CLOTS AND NO FURTHER MANAGEMENT UNTIL PATIENT COMPLAINED OF SEVERE ABDOMINAL PAIN THE MORNING AFTER SURGERY. BLADDER SCANNED AGAIN DISPLAYED TO BE RETAINING URINE. FOLEY AGAIN IRRIGATED BUT UNABLE TO FLUSH OR WITHDRAW IRRIGATION FLUID, FOLLOWING WITH FOLEY CATHETER REMOVED. PATIENT INCONTINENT OF URINE. FOLEY CATHETER REMAINED OUT AND PATIENT ABLE TO VOID WITHOUT ISSUE.
Patient problems: Abdominal Pain; Urinary Incontinence
MAUDE report 12106759 ↗MalfunctionJul 2, 2021Decrease in Suction; Suction Problem; Decrease in Suction; Suction Problem
PATIENT WAS STARTED ON A LASIX INFUSION DUE TO LOW URINE OUTPUT OVER A FEW HOURS. IN THE FOLLOWING HOURS HER FOLEY CATHETER WAS BYPASSING LARGE AMOUNTS OF URINE, THE NURSE INVESTIGATED THE SYSTEM INCLUDING A BLADDER SCAN. THE BLADDER SCAN WAS FOUND TO HAVE >999 ML. AS PER UNIT PROTOCOL, THE CATHETER WAS IRRIGATED TO RELIEVE ANY OBSTRUCTION BUT NON WERE FOUND. THE CATHETER BALLOON WAS DEFLATED IN PREPARATION FOR REMOVAL AND THE PATIENT WAS RELEASING LARGE VOLUME OF URINE INTO THE BED. FOLLOWING WHICH, A NEW FOLEY WAS INSERTED AND AN ADDITIONAL 1.5L OF URINE WAS COLLECTED. AN ADDITIONAL >1L WAS COLLECTED IN THE FOLLOWING HOUR AS WELL. THE PATIENT WAS SEDATED AND INTUBATED POST OPERATIVELY, HENCE UNABLE TO COMMUNICATE DISCOMFORT ASSOCIATED WITH SIGNIFICANT URINARY RETENTION.
Patient problems: Unspecified Kidney or Urinary Problem; Urinary Incontinence
MAUDE report 12106752 ↗MalfunctionJun 15, 2021Display or Visual Feedback Problem; Infusion or Flow Problem; Display or Visual Feedback Problem; Infusion or Flow Problem
DURING DAY ZERO OF POST-OPERATION, THE PATIENT WAS SHOWING WITH NO URINE OUTPUT IN OPERATING ROOM. WHEN ASSESSED BY THE NURSE BY A BLADDER SCANNER HE HAD 350 ML IN HIS BLADDER DESPITE HAVING A FOLEY CATHETER INSITU. THIS WAS IRRIGATED AND URINE CLEAR OF CLOTS CAME OUT. LOW VOLUMES OF URINE CONTINUED TO DRAIN IN THE FOLLOWING 12 HOURS, HOWEVER THE PATIENT WAS EXPERIENCING INCREASING ABDOMINAL PAIN AND URGE TO VOID. WHEN ASSESSED THE CATHETER WAS UNABLE TO IRRIGATE, AND THE BLADDER SCAN SHOWED RESIDUAL URINE IN THE BLADDER. THE FOLEY CATHETER WAS REMOVED AND THE PATIENT WAS INCONTINENT OF LARGE AMOUNTS OF URINE. THIS PATIENT EXPERIENCES SIGNIFICANT AND PAINFUL URINARY RETENTION DESPITE THE CATHETER BEING INSITU.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Unspecified Kidney or Urinary Problem
MAUDE report 11998430 ↗MalfunctionJun 11, 2021Display or Visual Feedback Problem; Suction Problem; Infusion or Flow Problem; Display or Visual Feedback Problem; Suction Problem; Infusion or Flow Problem
DURING DAY THREE OF POST-OPERATION, THE PATIENT WAS SHOWING SIGNS OF LOW URINE OUTPUT, SHE WAS STARTED ON A DIURETIC INFUSION TO FACILITATE FLUID REMOVAL AND DIURESIS. THE FOLEY CATHETER WAS BYPASSING URINE ONTO THE BED, SO THE NURSE ASSESSED THE SYSTEM. THIS INCLUDED A BLADDER SCAN WHICH SHOWED A RETAINED VOLUME OF >999ML IN HER BLADDER DESPITE HAVING A CATHETER INSITU. THE NURSE ATTEMPTED TO IRRIGATE THE FOLEY FOR OBSTRUCTION AND WAS UNSUCCESSFUL. IN FACT THE IRRIGATION FLUID BYPASSED AROUND THE CATHETER AN ONTO THE BED AS WELL. THE BALLOON OF THE FOLEY WAS DEFLATED IN PREPARATION FOR REMOVAL AND THE PATIENT WAS INCONTINENT OF LARGE VOLUMES OF URINE. FOLLOWING THIS, A NEW CATHETER WAS PLACED AND DRAINED 1.5 L OF URINE, AND >1 L IN THE HOUR FOLLOWING. THE PATIENT WAS SEDATED DURING THIS EVENT SO WAS UNABLE TO COMMUNICATE DISCOMFORT OR URGE TO VOID. THIS PATIENT WAS UNNECESSARILY EXPOSED TO HIGH DOSE DIURETICS AND HIGH-VOLUME URINE RETENTION WHICH COULD HAVE CAUSED SIGNIFICANTLY RENAL DYSFUNCTION.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Unspecified Kidney or Urinary Problem
MAUDE report 11983416 ↗MalfunctionMay 20, 2021Suction Problem; Suction Problem
POST-OP PATIENTS HAVING MULTIPLE ISSUES WITH SIGNIFICANT HEMATURIA, CATHETER OBSTRUCTION AND FOLEY CATHETER DISPLACEMENT CAUSING URINARY RETENTION. UPON REVIEW, THERE ARE CONCERNS RELATED TO MIGRATION OUT OF THE BLADDER.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Unspecified Kidney or Urinary Problem
MAUDE report 11855876 ↗MalfunctionOct 17, 2019Obstruction of Flow; Infusion or Flow Problem; Obstruction of Flow; Infusion or Flow Problem
UNABLE TO DRAW URINE FROM THE CATHETER. THE CATHETER LOOKS LIKE IT IS BLOCKED. THIS OCCURRED DURING USE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9202211 ↗
FDA clearance and classification
The FDA regulates product code EYC (catheter, upper urinary tract) under 21 CFR 876.5130 as a Class II device. 21 CFR 876.5130 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the DeRoyal 81-080416
Has the DeRoyal 81-080416 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00749756045766), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the DeRoyal 81-080416?
The FDA's MAUDE database lists 11 adverse event reports for this device, received between Oct 17, 2019 and Sep 21, 2026. In those reports, the most common specific device problems are “Infusion or Flow Problem” (5), “Suction Problem” (5) and “Device Dislodged or Dislocated” (2); the most common patient problems are “Unspecified Kidney or Urinary Problem” (5), “Urinary Incontinence” (2) and “Abdominal Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the DeRoyal 81-080416?
None of the reports are classified as deaths, 1 as an injury and 10 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the DeRoyal 81-080416, and is it still sold?
The FDA's device database lists Deroyal Industries, Inc. as the company responsible for it (brand DeRoyal, model 81-080416). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, UPPER URINARY TRACT devices
- DeRoyal 81-080414EUDeroyal Industries, Inc. · CATHETER, UPPER URINARY TRACT13 reports
- DeRoyal 81-080416EUDeroyal Industries, Inc. · CATHETER, UPPER URINARY TRACT12 reports
- DeRoyal 81-080408Deroyal Industries, Inc. · CATHETER, UPPER URINARY TRACT11 reports
- DeRoyal 81-080414Deroyal Industries, Inc. · CATHETER, UPPER URINARY TRACT6 reports
- DeRoyal 81-080410Deroyal Industries, Inc. · CATHETER, UPPER URINARY TRACT3 reports
- DeRoyal 81-080418Deroyal Industries, Inc. · CATHETER, UPPER URINARY TRACT1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.