DeRoyal 81-080408Recalls & reported problems
Deroyal Industries, Inc. · Foley Catheter with Temperature Sensor · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00749756045803), received between Nov 29, 2018 and Jul 29, 2026 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Burst Container or Vessel5
- Material Integrity Problem3
- Failure to Sense2
- Material Separation2
- Crack1
- Device Dislodged or Dislocated1
- Material Deformation1
- Material Puncture/Hole1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- No Consequences Or Impact To Patient1
- No Patient Involvement1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 29, 2026Material Separation
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50073941 ↗MalfunctionJun 25, 2026Material Separation
A USER FACILITY REPORTED, \"PATIENT'S TEMP ALARM WAS RINGING OFF. TEMP WAS READING FROM FOLEY CATHETER. UPON INSPECTION, THE DISTAL PORTION OF THE CATHETER REMAINED IN THE PT, HOWEVER THE REST OF THE DEVICE HAD COME APART AT THE JUNCTION WHERE THE BALLOON AND TEMP PROBE ATTACHED. REMAINDER OF FOLEY REMOVED FROM PT WITHOUT ISSUE AND BALLOON WAS FOUND TO BE DEFLATED.\"
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25620321 ↗MalfunctionAug 12, 2025Failure to Sense; Material Integrity Problem; Failure to Sense; Material Integrity Problem
A USER FACILITY REPORTED ON A MEDWATCH, "FOLEY TEMPERATURE PROBE WIRE EXPOSED. TEMPERATURE FUNCTION NO LONGER USED. FOLEY STILL IN PATIENT, BELIEVED TO BE 8FR."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22786248 ↗MalfunctionFeb 20, 2025Failure to Sense; Failure to Sense
A USER FACILITY REPORTED, "THE FOLEY IS TESTED, WORKS, PLACED IN PATIENT AND THEN NO LONGER WORKS".
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21425966 ↗MalfunctionAug 9, 2024Burst Container or Vessel; Burst Container or Vessel
A USER FACILITY REPORTED, "BALLOON BURST IN PATIENT."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19951599 ↗MalfunctionAug 9, 2024Burst Container or Vessel; Burst Container or Vessel
A USER FACILITY REPORTED, "PT IN ICU AND THE BALLOON IN THE CATHETER BURST."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19951328 ↗MalfunctionAug 9, 2024Burst Container or Vessel; Burst Container or Vessel
A USER FACILITY REPORTED, "BEING USED IN A PT IN THE ICU AND THE BALLOON BURST."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19951305 ↗MalfunctionAug 9, 2024Burst Container or Vessel; Burst Container or Vessel
A USER FACILITY REPORTED, "CHECKED BALLOON AND WATER EVERYWHERE AND UPON REMOVING FROM PATIENT DISCOVERED IT HAD BURST."
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19951288 ↗MalfunctionAug 7, 2024Burst Container or Vessel; Material Puncture/Hole; Device Dislodged or Dislocated; Burst Container or Vessel; Material Puncture/Hole; Device Dislodged or Dislocated
BURST FOLEY BALLOON, LED TO DISLODGEMENT OF INDWELLING URINARY CATHETER. LARGE HOLE NOTED IN BALLOON ON INSPECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19927886 ↗MalfunctionSep 9, 2019Crack; Material Integrity Problem; Crack; Material Integrity Problem
WHEN THE BABY'S DIAPER WAS OPENED, THE FOLEY CATHETER WAS FOUND CRACKED WITH THE BALLOON DEFLATED. THE CATHETER HAD FALLEN APART. THE PRODUCT WAS BEING USED TO MEASURE THE PATIENT'S BODY TEMPERATURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8977597 ↗MalfunctionNov 29, 2018Material Deformation; Material Integrity Problem; Material Deformation; Material Integrity Problem
THE TEMPERATURE PROBE WIRES WERE PROTRUDING FROM THE CATHETER, WHICH PLACED THE PATIENT AT RISK FOR SIGNIFICANT URETHRAL INJURY. THE CATHETER WAS REMOVED AND REPLACED, AND THERE WAS NO PATIENT HARM.
Patient problems: No Patient Involvement
MAUDE report 8116868 ↗
FDA clearance and classification
The FDA regulates product code EYC (catheter, upper urinary tract) under 21 CFR 876.5130 as a Class II device. 21 CFR 876.5130 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the DeRoyal 81-080408
Has the DeRoyal 81-080408 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00749756045803), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the DeRoyal 81-080408?
The FDA's MAUDE database lists 11 adverse event reports for this device, received between Nov 29, 2018 and Jul 29, 2026. In those reports, the most common specific device problems are “Burst Container or Vessel” (5), “Material Integrity Problem” (3) and “Failure to Sense” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the DeRoyal 81-080408?
None of the reports are classified as deaths, 0 as injuries and 11 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the DeRoyal 81-080408, and is it still sold?
The FDA's device database lists Deroyal Industries, Inc. as the company responsible for it (brand DeRoyal, model 81-080408). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CATHETER, UPPER URINARY TRACT devices
- DeRoyal 81-080414EUDeroyal Industries, Inc. · CATHETER, UPPER URINARY TRACT13 reports
- DeRoyal 81-080416EUDeroyal Industries, Inc. · CATHETER, UPPER URINARY TRACT12 reports
- DeRoyal 81-080416Deroyal Industries, Inc. · CATHETER, UPPER URINARY TRACT11 reports
- DeRoyal 81-080414Deroyal Industries, Inc. · CATHETER, UPPER URINARY TRACT6 reports
- DeRoyal 81-080410Deroyal Industries, Inc. · CATHETER, UPPER URINARY TRACT3 reports
- DeRoyal 81-080418Deroyal Industries, Inc. · CATHETER, UPPER URINARY TRACT1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.