Skip to content

DeRoyal HNICU-03Recalls & reported problems

Deroyal Industries, Inc. · HYDRO-TEMP® Cover without Foam · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00749756255608), received between Jan 9, 2020 and Jan 30, 2026 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings3
  • Output Problem3
  • Device Dislodged or Dislocated1
  • Manufacturing, Packaging or Shipping Problem1
  • Product Quality Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Hyperthermia1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 30, 2026Device Dislodged or Dislocated

    A USER FACILITY REPORTED, "THE PROBE COVERS NOT PROPERLY MOLDING AND ADHERING TO THE PATIENT'S SKIN TO HOLD DOWN THE NICU PROBE TO THE PATIENT'S SKIN. SEVERAL PATIENTS HAVE OVERHEATED DUE TO THE PROBE NOT PROPERLY MOLDING AND ADHERING TO THE PATIENT'S SKIN."

    Patient problems: Hyperthermia

    MAUDE report 24222173 ↗
  • MalfunctionOct 6, 2021Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE PROBE COVER IS NOT STICKING TO THE PATIENT AROUND THE PROBE, CAUSING IT TO LIFT UNDER THE COVER. THERE WAS AN ELEVATED TEMPERATURE OF 99.8°F TO 100.3°F, HOWEVER THE GIRAFFE WAS READING COOLER TEMPERATURES OF 97.8°F-98.0°F.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12582531 ↗
  • MalfunctionOct 6, 2021Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE PROBE COVER IS NOT STICKING TO THE PATIENT AROUND THE PROBE, CAUSING IT TO LIFT UNDER THE COVER. THERE WAS AN ELEVATED TEMPERATURE OF 99.8°F TO 100.3°F, HOWEVER THE GIRAFFE WAS READING COOLER TEMPERATURES OF 97.8°F-98.0°F.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12582530 ↗
  • MalfunctionOct 6, 2021Incorrect, Inadequate or Imprecise Result or Readings; Output Problem; Incorrect, Inadequate or Imprecise Result or Readings; Output Problem

    THE PROBE COVER IS NOT STICKING TO THE PATIENT AROUND THE PROBE, CAUSING IT TO LIFT UNDER THE COVER. THERE WAS AN ELEVATED TEMPERATURE OF 99.8°F TO 100.3°F, HOWEVER THE GIRAFFE WAS READING COOLER TEMPERATURES OF 97.8°F-98.0°F.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12582529 ↗
  • MalfunctionJan 9, 2020Product Quality Problem; Manufacturing, Packaging or Shipping Problem; Product Quality Problem; Manufacturing, Packaging or Shipping Problem

    WHEN THE TEMP SENSOR IS USED ON THE MICRO-PREEMIE POPULATION, THE PAD IS SLIDING. THE GEL ON THE BACK LOSES ITS "STICKY" PROPERTY, WHICH CAUSES THE DEVICE TO NOT STICK TO THE SKIN. THE NURSES ARE REPORTING UNSTABLE TEMPERATURES IN THAT POPULATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9570128 ↗

Other WARMER, INFANT RADIANT devices

All WARMER, INFANT RADIANT devices →