Panda i-Res Warmer 1.0Recalls & reported problems
Datex-Ohmeda, Inc. · Panda iRes Warmer V1.0 · Class 2
All models of Panda (2)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00840682152709), received between Apr 21, 2025 and Sep 23, 2026 — classified as 15 malfunctions.
The most frequently reported specific device problem is “Break” (10 reports).
1 FDA recall record name this device's identifier, 1 still open.
17 other recalls cover devices cleared under the same FDA submission (K070377, K090697, K122267, K123309); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 9, 2024Class IIOpen, Classified
GE HealthCare has become aware that complete electrical safety testing was not conducted during manufacturing of certain Giraffe and Panda iRes Warmers. There are multiple electrical safety protections included in the product design. However, because the electrical safety testing in manufacturing was incomplete for these devices, there is the potential for leakage current to exceed IEC 60601 limits.
Z-3081-2024 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Break10
- Insufficient Flow or Under Infusion3
- Detachment of Device or Device Component1
- No Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions15
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2026Break
It was reported that the side panel was damaged. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50868693 ↗MalfunctionSep 18, 2026Insufficient Flow or Under Infusion
It was reported that there was a malfunction during service causing loss of oxygen concentration. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50817298 ↗MalfunctionJul 30, 2026Break
It was reported that the front panel was broken. There was no patient involvement reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50082663 ↗MalfunctionMar 10, 2026Break
IT WAS REPORTED THAT THE SIDE PANEL WAS DAMAGED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24558094 ↗MalfunctionMar 4, 2026Break
IT WAS REPORTED THAT THE SIDEWALL PANEL LATCH WAS DAMAGED DURING SERVICE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24512251 ↗MalfunctionMar 3, 2026Break
IT WAS REPORTED THAT THE SIDE PANEL WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24496213 ↗MalfunctionMar 3, 2026Insufficient Flow or Under Infusion
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN SIGNIFICANT DEGRADATION OR LOSS OF OXYGEN THERAPY DURING SERVICE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24496071 ↗MalfunctionFeb 27, 2026Detachment of Device or Device Component
IT WAS REPORTED THAT THE SIDE PANEL WAS MISSING. THERE WAS NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24469021 ↗MalfunctionFeb 26, 2026Insufficient Flow or Under Infusion
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF OXYGEN THERAPY. THERE WAS NO PATIENT INVOLVEMENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24455506 ↗MalfunctionDec 31, 2025Break; Break
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN A BROKEN SIDE PANEL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23936198 ↗MalfunctionDec 23, 2025Break; Break
IT WAS REPORTED THAT THERE WAS A BROKEN SIDE PANEL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23877048 ↗MalfunctionJul 2, 2025No Flow; No Flow
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF OXYGEN THERAPY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22385458 ↗MalfunctionMay 13, 2025Break; Break
IT WAS REPORTED THAT THE SIDE PANEL WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22009225 ↗MalfunctionMay 6, 2025Break; Break
IT WAS REPORTED THAT THE SIDE PANEL WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21967652 ↗MalfunctionApr 21, 2025Break; Break
IT WAS REPORTED THAT THE FRONT PANEL WALL WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21871752 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K070377, K090697, K122267, K123309. One submission often covers many models, so check the product and lot details in each record.
- Z-2359-2026 · May 8, 2026 · Open, ClassifiedPanda Freestanding Warmer, Infant Radiant Warmer Product Identifiers: 2063568-001 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenanc
- Z-1847-2019 · Mar 15, 2019 · TerminatedPanda Infant Radiant Warmer Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infa
- Z-1846-2019 · Mar 15, 2019 · TerminatedGiraffe Infant Radiant Warmer Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. In
- Z-0484-2019 · Sep 25, 2018 · TerminatedPanda Wall Mount warmer. Infant radiant warmer
- Z-0482-2019 · Sep 25, 2018 · TerminatedPanda iRes Infant Radiant Warmer. Device sold under the following product names: Panda iRes Warmer and Panda iRes Bedded Warmer.
- Z-0481-2019 · Sep 25, 2018 · TerminatedGiraffe Warmer. Device sold under the following product names: Giraffe Warmer and Giraffe Bedded Warmer. infant radiant warmer.
- Z-0483-2019 · Sep 25, 2018 · TerminatedPanda Freestanding Warmer. Device sold under the following product names: Panda Freestanding Warmer and Panda Freestanding Warmer Generic ATO Model. Infant Radiant warmer.
- Z-1508-2017 · Nov 18, 2016 · TerminatedGiraffe Bedded Warmer, Panda Warmer Modifiction to Giraffe and Panda Warmer- Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to therm
- Z-1506-2017 · Nov 18, 2016 · TerminatedGiraffe Bedded Warmer, Panda Warmer Giraffe and Panda Warmer- Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermoregulate based
- Z-1547-2014 · Apr 2, 2014 · TerminatedGiraffe Infant Warmers integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrated monitorin
- Z-1550-2014 · Apr 2, 2014 · TerminatedPanda Wall-Mount Infant Warmers, integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrated
- Z-1549-2014 · Apr 2, 2014 · TerminatedPanda Freestanding Infant Warmers, integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrat
- Z-1548-2014 · Apr 2, 2014 · TerminatedPanda iRes Infant Warmers integrated with Nellcor SpO2. Controlled infrared heat to neonates who are unable to thermo-regulate based on their own physiology with integrated monito
- Z-0912-2014 · Dec 12, 2013 · TerminatedGiraffe Infant Warmers, part number M1118179, labeled in part *** GE Healthcare 8880 Gorman Road Laurel, MD 20723***www.gehealthcare.com* Giraffe Infant Warmers provide infrared h
- Z-0913-2014 · Dec 12, 2013 · TerminatedPanda iRes Infant Warmers, part number M1112198 and Panda Freestanding Warmers, part number 2063568, labeled in part *** GE Healthcare 8880 Gorman Road Laurel, MD 20723***www.gehea
…and 2 more.
Other WARMER, INFANT RADIANT devices
- F&P Neopuff RD900AEUFisher & Paykel Healthcare Limited · Warmer, Infant Radiant433 reports
- Resuscitaire Radiant Warmer (RW82, RW82 VHA) MU20503Draeger Medical Systems, Inc. · Warmer, Infant Radiant1 recall3 deaths15 reports
- Panda iRes WarmerDatex-Ohmeda, Inc. · WARMER, INFANT RADIANT34 reports
- Giraffe WarmerDatex-Ohmeda, Inc. · WARMER, INFANT RADIANT27 reports
- DeRoyal HNICU-37Deroyal Industries, Inc. · WARMER, INFANT RADIANT25 reports
- F&P Neopuff RD900AFUFisher & Paykel Healthcare Limited · Warmer, Infant Radiant18 reports
- DeRoyal HNICU-03Deroyal Industries, Inc. · WARMER, INFANT RADIANT5 reports
- DeRoyal 81-010400Deroyal Industries, Inc. · WARMER, INFANT RADIANT5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page